High failure rate of the interspinous distraction device (X-Stop) for the treatment of lumbar spinal stenosis caused by degenerative spondylolisthesis

被引:141
作者
Verhoof, Olaf J. [1 ]
Bron, Johannes L. [1 ]
Wapstra, Frits H. [1 ]
van Royen, Barend J. [1 ]
机构
[1] Vrije Univ Amsterdam, Med Ctr, Dept Orthopaed Surg, Amsterdam, Netherlands
关键词
lumbar spinal stenosis; X-Stop; degenerative spondylolisthesis;
D O I
10.1007/s00586-007-0492-x
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The X-Stop interspinous distraction device has shown to be an attractive alternative to conventional surgical procedures in the treatment of symptomatic degenerative lumbar spinal stenosis. However, the effectiveness of the X-Stop in symptomatic degenerative lumbar spinal stenosis caused by degenerative spondylolisthesis is not known. A cohort of 12 consecutive patients with symptomatic lumbar spinal stenosis caused by degenerative spondylolisthesis were treated with the X-Stop interspinous distraction device. All patients had low back pain, neurogenic claudication and radiculopathy. Pre-operative radiographs revealed an average slip of 19.6%. MRI of the lumbosacral spine showed a severe stenosis. In ten patients, the X-Stop was placed at the L4 - 5 level, whereas two patients were treated at both, L3 - 4 and L4 - 5 level. The mean follow-up was 30.3 months. In eight patients a complete relief of symptoms was observed post-operatively, whereas the remaining 4 patients experienced no relief of symptoms. Recurrence of pain, neurogenic claudication, and worsening of neurological symptoms was observed in three patients within 24 months. Post-operative radiographs and MRI did not show any changes in the percentage of slip or spinal dimensions. Finally, secondary surgical treatment by decompression with posterolateral fusion was performed in seven patients (58%) within 24 months. In conclusion, the X-Stop interspinous distraction device showed an extremely high failure rate, defined as surgical re-intervention, after short term follow-up in patients with spinal stenosis caused by degenerative spondylolisthesis. We do not recommend the X-Stop for the treatment of spinal stenosis complicating degenerative spondylolisthesis.
引用
收藏
页码:188 / 192
页数:5
相关论文
共 48 条
[21]   Minimally Invasive Decompression Versus X-Stop in Lumbar Spinal Stenosis A Randomized Controlled Multicenter Study [J].
Lonne, Greger ;
Johnsen, Lars Gunnar ;
Rossvoll, Ivar ;
Andresen, Hege ;
Storheim, Kjersti ;
Zwart, John Anker ;
Nygaard, Oystein P. .
SPINE, 2015, 40 (02) :77-85
[22]   Two-year results of interspinous spacer (X-Stop) implantation in 175 patients with neurologic intermittent claudication due to lumbar spinal stenosis [J].
Kuchta, Johannes ;
Sobottke, Rolf ;
Eysel, Peer ;
Simons, Patrick .
EUROPEAN SPINE JOURNAL, 2009, 18 (06) :823-829
[23]   Two-Year Evaluation of the X-STOP Interspinous Spacer in Different Primary Patient Populations With Neurogenic Intermittent Claudication Because of Lumbar Spinal Stenosis [J].
Hartjen, Charles A. ;
Resnick, Daniel K. ;
Hsu, Ken Y. ;
Zucherman, James F. ;
Hsu, Emily H. ;
Skidmore, Grant A. .
CLINICAL SPINE SURGERY, 2016, 29 (07) :305-311
[24]   Endoscopic lumbar spinal hybrid surgery in the treatment of multiple-level lumbar degenerative spondylolisthesis and spinal stenosis [J].
Chien, Kai -Ting ;
Peng, Pao-Sheng .
INTERDISCIPLINARY NEUROSURGERY-ADVANCED TECHNIQUES AND CASE MANAGEMENT, 2022, 29
[25]   A randomized controlled trial of the X-Stop interspinous distractor device versus laminectomy for lumbar spinal stenosis with 2-year quality-of-life and cost-effectiveness outcomes [J].
Borg, Anouk ;
Hill, Ciaran Scott ;
Nurboja, Besnik ;
Critchley, Giles ;
Choi, David .
JOURNAL OF NEUROSURGERY-SPINE, 2021, 34 (04) :544-552
[26]   Efficacy of Surgical Treatment for Lumbar Stenosis with the X-STOP: An Issue in Need of Closer Inspection [J].
Hsieh, Patrick C. .
WORLD NEUROSURGERY, 2013, 80 (1-2) :74-77
[27]   Comparing Cost-effectiveness of X-Stop With Minimally Invasive Decompression in Lumbar Spinal Stenosis A Randomized Controlled Trial [J].
Lonne, Greger ;
Johnsen, Lars Gunnar ;
Aas, Eline ;
Lydersen, Stian ;
Andresen, Hege ;
Ronning, Roar ;
Nygaard, Oystein P. .
SPINE, 2015, 40 (08) :514-520
[28]   Biomechanical characteristics of a novel interspinous distraction fusion device in the treatment of lumbar degenerative diseases: a finite element analysis [J].
Chen, Mengmeng ;
Deng, Jiechao ;
Bao, Li ;
Jia, Pu ;
Feng, Fei ;
Shi, Guan ;
Tang, Hai ;
Chen, Hao .
BMC MUSCULOSKELETAL DISORDERS, 2023, 24 (01)
[29]   Surgical Treatment of the Spinal Stenosis with an Interspinous Distraction Device: Do We Really Restore the Foraminal Height? [J].
Celik, Haydar ;
Derincek, Alihan ;
Koksal, Ismet .
TURKISH NEUROSURGERY, 2012, 22 (01) :50-54
[30]   Lumbar Interspinous Process Fixation and Fusion with Stand-Alone Interlaminar Lumbar Instrumented Fusion Implant in Patients with Degenerative Spondylolisthesis Undergoing Decompression for Spinal Stenosis [J].
Postacchini, Franco ;
Postacchini, Roberto ;
Menchetti, Pier Paolo Maria ;
Sessa, Pasquale ;
Paolino, Michela ;
Cinotti, Gianluca .
ASIAN SPINE JOURNAL, 2016, 10 (01) :27-37