Design and operation of a low cost, reliable millimeter-wave interferometer

被引:10
作者
Porte, L [1 ]
Rettig, CL [1 ]
Peebles, WA [1 ]
Ngyuen, X [1 ]
机构
[1] Univ Calif Los Angeles, Dept Elect Engn, Los Angeles, CA 90024 USA
关键词
D O I
10.1063/1.1149372
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Knowledge of basic plasma parameters, such as line integrated density and turbulence spectra, is essential for physics understanding and plasma control. Integration of plasma diagnostic systems into a single port combined with diagnostic cost reduction is an important advance that can impact measurement availability on all fusion devices. This article presents preliminary data from a unique, low cost and reliable millimeter-wave interferometer that has been integrated to operate simultaneously with a collective Thomson scattering system. The interferometer uses only one commercially available frequency chirped source and only one detector. The DIII-D system design will be described together with preliminary design of a tangentially viewing system for the Pegasus spherical torus experiment. (C) 1999 American Institute of Physics. [S0034-6748(99)75601-7].
引用
收藏
页码:1082 / 1084
页数:3
相关论文
共 8 条
[1]   THE HELICALLY SYMMETRICAL EXPERIMENT, (HSX) GOALS, DESIGN AND STATUS [J].
ANDERSON, FSB ;
ALMAGRI, AF ;
ANDERSON, DT ;
MATTHEWS, PG ;
TALMADGE, JN ;
SHOHET, JL .
FUSION TECHNOLOGY, 1995, 27 :273-277
[2]  
CHRZANOWSKI JH, 1995, P 16 INT S FUS ENG C, V2, P1430
[3]   THE MADISON SYMMETRICAL TORUS [J].
DEXTER, RN ;
KERST, DW ;
LOVELL, TW ;
PRAGER, SC ;
SPROTT, JC .
FUSION TECHNOLOGY, 1991, 19 (01) :131-139
[4]   1-MILLIMETER WAVE INTERFEROMETER FOR THE MEASUREMENT OF LINE INTEGRAL ELECTRON-DENSITY ON TFTR [J].
EFTHIMION, PC ;
TAYLOR, G ;
ERNST, W ;
GOLDMAN, M ;
MCCARTHY, M ;
ANDERSON, H ;
LUHMANN, NC .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1985, 56 (05) :908-910
[5]  
HUTCHINSON IH, 1987, PRINCIPLES PLASMA DI
[6]   A quasi-optical 1 mm microwave heterodyne interferometer for plasma diagnostics using a frequency-tripled Gunn oscillator [J].
Niemoller, N ;
Gathen, VSV ;
Stampa, A ;
Dobele, HF .
PLASMA SOURCES SCIENCE & TECHNOLOGY, 1997, 6 (04) :478-483
[7]   A 2-COLOR MM-WAVE INTERFEROMETER FOR THE JET DIVERTOR [J].
PRENTICE, R ;
EDLINGTON, T ;
SMITH, RTC ;
TROTMAN, DL ;
WYLDE, RJ ;
ZIMMERMANN, P .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02) :1154-1158
[8]   DEVELOPMENT AND OPERATION OF A BACKWARD-WAVE OSCILLATOR BASED FIR SCATTERING SYSTEM FOR DIII-D [J].
RETTIG, CL ;
BURNS, S ;
PHILIPONA, R ;
PEEBLES, WA ;
LUHMANN, NC .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10) :3010-3012