Dynamic Contrast-Enhanced MRI in Mice: An Investigation of Model Parameter Uncertainties

被引:3
作者
Rukat, Tammo [1 ,2 ]
Walker-Samuel, Simon [3 ]
Reinsberg, Stefan A. [1 ]
机构
[1] Univ British Columbia, Dept Phys & Astron, Vancouver, BC V6T 1Z1, Canada
[2] Humboldt Univ, Dept Phys, Berlin, Germany
[3] UCL Ctr Adv Biomed Imaging, Div Med, London, England
基金
加拿大自然科学与工程研究理事会; 英国惠康基金;
关键词
dynamic contrast-enhanced-MRI; perfusion; permeability; pharmacokinetic models; tracer-kinetic models; ARTERIAL INPUT FUNCTION; DCE-MRI; TRACER KINETICS; TEMPORAL-RESOLUTION; T-1-WEIGHTED MRI; WATER EXCHANGE; TISSUE; PERMEABILITY; CANCER; REQUIREMENTS;
D O I
10.1002/mrm.25319
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo establish the experimental factors that dominate the uncertainty of hemodynamic parameters in commonly used pharmacokinetic models. MethodsBy fitting simulation results from a multiregion tissue exchange model (Multiple path, Multiple tracer, Indicator Dilution, 4 region), the precision and accuracy of hemodynamic parameters in dynamic contrast-enhanced MRI with four tracer kinetic models is investigated. The impact of various injection rates as well as imprecise knowledge of the arterial input functions is examined. ResultsFast injections are beneficial for K-trans precision within the extended Tofts model and within the two-compartment exchange model but do not affect the other models under investigation. Biases from errors in the arterial input functions are mostly consistent in size and direction for the simple and the extended Tofts model, while they are hardly predictable for the other models. Errors in the hematocrit introduce the greatest loss in parameter accuracy, amounting to an average K-trans bias of 40% for a 30% overestimation throughout all models. ConclusionThis simulation study allows the detailed inspection of the isolated impact from various experimental conditions on parameter uncertainty. Because parameter uncertainty comparable to human studies was found, this study represents a validation of preclinical dynamic contrast-enhanced MRI for modeling human tumor physiology. Magn Reson Med 73:1979-1987, 2015. (c) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:1979 / 1987
页数:9
相关论文
共 50 条
[41]   Fundamentals of Tracer Kinetics for Dynamic Contrast-Enhanced MRI [J].
Koh, Tong San ;
Bisdas, Sotirios ;
Koh, Dow Mu ;
Thng, Choon Hua .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 34 (06) :1262-1276
[42]   Dynamic contrast-enhanced MRI for oncology drug development [J].
Sung, Yu Sub ;
Park, Bumwoo ;
Choi, Yoonseok ;
Lim, Hyeong-Seok ;
Woo, Dong-Cheol ;
Kim, Kyung Won ;
Kim, Jeong Kon .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2016, 44 (02) :251-264
[43]   Operator dependency of arterial input function in dynamic contrast-enhanced MRI [J].
Magne Kleppestø ;
Atle Bjørnerud ;
Inge Rasmus Groote ;
Minjae Kim ;
Jonas Vardal ;
Christopher Larsson .
Magnetic Resonance Materials in Physics, Biology and Medicine, 2022, 35 :105-112
[44]   Modified MR dispersion imaging in prostate dynamic contrast-enhanced MRI [J].
Sung, Kyunghyun .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2019, 50 (04) :1307-1317
[45]   Diagnosis of Spinal Lesion Using Heuristic and Pharmacokinetic Parameter Measured by Dynamic Contrast-Enhanced MRI [J].
Lang, Ning ;
Yuan, Huishu ;
Yu, Hon J. ;
Su, Min-Ying .
ACADEMIC RADIOLOGY, 2017, 24 (07) :867-875
[46]   Optimization of Two-Compartment-Exchange- Model Analysis for Dynamic Contrast-Enhanced MRI Incorporating Bolus Arrival Time [J].
Nadav, Guy ;
Liberman, Gilad ;
Artzi, Moran ;
Kiryati, Nahum ;
Ben Bashat, Dafna .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2017, 45 (01) :237-249
[47]   Comparison of Dynamic Contrast-Enhanced MRI and Dynamic Contrast-Enhanced CT Biomarkers in Bladder Cancer [J].
Naish, J. H. ;
McGrath, D. M. ;
Bains, L. J. ;
Passera, K. ;
Roberts, C. ;
Watson, Y. ;
Cheung, S. ;
Taylor, M. B. ;
Logue, J. P. ;
Buckley, D. L. ;
Tessier, J. ;
Young, H. ;
Waterton, J. C. ;
Parker, G. J. M. .
MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (01) :219-226
[48]   A Multi-Institutional Comparison of Dynamic Contrast-Enhanced Magnetic Resonance Imaging Parameter Calculations [J].
Ger, Rachel B. ;
Mohamed, Abdallah S. R. ;
Awan, Musaddiq J. ;
Ding, Yao ;
Li, Kimberly ;
Fave, Xenia J. ;
Beers, Andrew L. ;
Driscoll, Brandon ;
Elhalawani, Hesham ;
Hormuth, David A., II ;
van Houdt, Petra J. ;
He, Renjie ;
Zhou, Shouhao ;
Mathieu, Kelsey B. ;
Li, Heng ;
Coolens, Catherine ;
Chung, Caroline ;
Bankson, James A. ;
Huang, Wei ;
Wang, Jihong ;
Sandulache, Vlad C. ;
Lai, Stephen Y. ;
Howell, Rebecca M. ;
Stafford, R. Jason ;
Yankeelov, Thomas E. ;
van der Heide, Uulke A. ;
Frank, Steven J. ;
Barboriak, Daniel P. ;
Hazle, John D. ;
Court, Laurence E. ;
Kalpathy-Cramer, Jayashree ;
Fuller, Clifton D. .
SCIENTIFIC REPORTS, 2017, 7
[49]   Model selection in dynamic contrast enhanced MRI: The Akaike Information Criterion [J].
Ingrisch, M. ;
Sourbron, S. ;
Reiser, M. F. ;
Peller, M. .
WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 4: IMAGE PROCESSING, BIOSIGNAL PROCESSING, MODELLING AND SIMULATION, BIOMECHANICS, 2010, 25 :356-358
[50]   Improved parameter extraction and classification for dynamic contrast enhanced MRI of prostate [J].
Haq, Nandinee Fariah ;
Kozlowski, Piotr ;
Jones, Edward C. ;
Chang, Silvia D. ;
Goldenberg, S. Larry ;
Moradi, Mehdi .
MEDICAL IMAGING 2014: COMPUTER-AIDED DIAGNOSIS, 2014, 9035