Low-dose computed tomography with 4th-generation iterative reconstruction algorithm in assessment of oncologic patients

被引:3
作者
Ippolito, Davide [1 ,2 ]
Casiraghi, Alessandra Silvia [1 ,2 ]
Franzesi, Cammillo Talei [1 ,2 ]
Fior, Davide [1 ,2 ]
Meloni, Franca [2 ]
Sironi, Sandro [2 ,3 ]
机构
[1] San Gerardo Hosp, Dept Diagnost Radiol, I-20900 Monza, Italy
[2] Univ Milano Bicocca, Sch Med, Via Pergolesi 33, I-20900 Monza, Italy
[3] Papa Giovanni XXIII Hosp, Dept Diagnost Radiol, I-24127 Bergamo, Italy
关键词
Computed tomography; Low-dose computed tomography; Tube current modulation; Oncologic imaging; Radiation dose; FILTERED BACK-PROJECTION; IMAGE-QUALITY; RADIATION-EXPOSURE; CT; REDUCTION;
D O I
10.4251/wjgo.v9.i10.423
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
AIM To compare radiation dose and image quality of low-dose computed tomography (CT) protocol combined with hybrid-iterative reconstruction algorithm with standard-dose CT examinations for follow-up of oncologic patients. METHODS Fifty-one patients with known malignant diseases which underwent, during clinical follow-up, both standard-dose and low-dose whole-body CT scans were enrolled. Low-dose CT was performed on 256-row scanner, with 120 kV and automated mA modulation, and iterative reconstruction algorithm. Standard-dose CT was performed on 16-rows scanner, with 120 kV, 200-400 mAs (depending on patient weight). We evaluated density values and signal-to-noise ratio, along with image noise (SD), sharpness and diagnostic quality with 4-point scale. RESULTS Density values in liver, spleen and aorta were higher in low-dose images (liver 112.55 HU vs 103.90 HU, P < 0.001), as SD values in liver and spleen (liver 16.81 vs 14.41). Volumetric-Computed-Tomographic-Dose-Index (CTDIvol) and Dose-Length-Product (DLP) were significantly lower in low-dose CT as compared to standard-dose (DLP 1025.6 mGy*cm vs 1429.2 mGy*cm, P < 0.001) with overall dose reduction of 28.9%. Qualitative analysis did not reveal significant differences in image noise and diagnostic quality. CONCLUSION Automatic tube-current modulation combined with hybriditerative algorithm allows radiation dose reduction of 28.9% without loss of diagnostic quality, being useful in reducing dose exposure in oncologic patients.
引用
收藏
页码:423 / 430
页数:8
相关论文
共 26 条
  • [1] Adams J.P., 2009, Practical Methods for Biocatalysis and Biotransformations, P1
  • [2] Using "iDose4" iterative reconstruction algorithm in adults' chest-abdomen-pelvis CT examinations: effect on image quality in relation to patient radiation exposure
    Arapakis, I.
    Efstathopoulos, E.
    Tsitsia, V.
    Kordolaimi, S.
    Economopoulos, N.
    Argentos, S.
    Ploussi, A.
    Alexopoulou, E.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2014, 87 (1036)
  • [3] Benefits of sinogram-affirmed iterative reconstruction in 0.4 mSv ultra-low-dose CT of the upper abdomen following transarterial chemoembolisation: comparison to low-dose and standard-dose CT and filtered back projection technique
    Bodelle, B.
    Isler, S.
    Scholtz, J. -E.
    Frellesen, C.
    Luboldt, W.
    Vogl, T. J.
    Beeres, M.
    [J]. CLINICAL RADIOLOGY, 2016, 71 (01) : E11 - E15
  • [4] Current concepts - Computed tomography - An increasing source of radiation exposure
    Brenner, David J.
    Hall, Eric J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) : 2277 - 2284
  • [5] Brown GR M., 2014, RECOMMENDATIONS CROS, V2nd
  • [6] Assessment of a Model-Based, Iterative Reconstruction Algorithm (MBIR) Regarding Image Quality and Dose Reduction in Liver Computed Tomography
    Chang, Won
    Lee, Jeong Min
    Lee, Kyunghee
    Yoon, Jeong Hee
    Yu, Mi Hye
    Han, Joon Koo
    Choi, Byung Ihn
    [J]. INVESTIGATIVE RADIOLOGY, 2013, 48 (08) : 598 - 606
  • [7] ESR, 2011, INSIGHTS IMAGING, V2, P357, DOI 10.1007/s13244-011-0108-1
  • [8] Reduction of radiation exposure and improvement of image quality with BMI-adapted prospective cardiac computed tomography and iterative reconstruction
    Hosch, Waldemar
    Stiller, Wolfram
    Mueller, Dirk
    Gitsioudis, Gitsios
    Welzel, Johanna
    Dadrich, Monika
    Buss, Sebastian J.
    Giannitsis, Evangelos
    Kauczor, Hans U.
    Katus, Hugo A.
    Korosoglou, Grigorios
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (11) : 3568 - 3576
  • [9] iDose4 Iterative Reconstruction Technique, 2011, IDOSE4 IT REC TECHN
  • [10] Dose reduction in oncological staging multidetector CT: effect of iterative reconstruction
    Karpitschka, M.
    Augart, D.
    Becker, H-C
    Reiser, M.
    Graser, A.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1021)