The interaction of subglottic drainage, cuff pressure, and oral care on endotracheal tube fluid leakage: A benchtop study

被引:7
作者
Chair, Sek Ying [1 ]
Chan, David Wing Keung [2 ]
Cao, Xi [1 ]
机构
[1] Chinese Univ Hong Kong, Fac Med, Nethersole Sch Nursing, Shatin, Hong Kong, Peoples R China
[2] Caritas Inst Higher Educ, Sch Hlth Sci, Hong Kong, Peoples R China
关键词
Critical care; Ventilator-associated pneumonia; Endotracheal tube; Subglottic secretion drainage; Fluid leakage; VENTILATOR-ASSOCIATED PNEUMONIA; SECRETION DRAINAGE; EXPIRATORY PRESSURE; TRACHEAL CUFF; PREVENTION; MICROASPIRATION; BUNDLE;
D O I
10.1016/j.aucc.2019.05.002
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The types of endotracheal tube, positive end-expiratory pressure (PEEP) level, endotracheal tube cuff pressure level, and nursing activity may influence the occurrence of pulmonary aspiration in ventilated patients with an endotracheal tube, but the evidence on their degree of influence is still inconclusive. Aim: The aim of this study was to examine the effect of endotracheal tubes with or without subglottic secretion drainage on fluid leakage across endotracheal tube cuffs under different PEEP levels, different cuff pressures, and with or without oral care. Method: This was a benchtop study with a multifactorial experimental design using two sets of airway. The two types of endotracheal tubes were tested through 12 scenarios with 240 combinations of different PEEP levels (0, 5, and 10 cmH(2)O), endotracheal tube cuff pressures (15 and 25 cmH(2)O), and oral care (with or without) using two sets of airway manikins. Each scenario lasted for 30 min, and fluid leakage was measured at the end of each scenario. Generalised Linear Model test was used to analyse fluid leak at 30 min (with and without interaction effect). Results: A total of 100 cases showed fluid leakage, with more fluid leakage occurring in low cuff pressure (15 cmH(2)O), no PEEP, and with oral care. Results of the Generalised Linear Model revealed that endotracheal tubes with subglottic secretion drainage, high PEEP (at 10 cmH(2)O), normal cuff pressure (at 25 cmH(2)O), and no oral care demonstrated a significant effect in reducing fluid leakage than endotracheal tubes without subglottic secretion drainage, low PEEP (at 5 or 0 cmH(2)O), and low cuff pressure (at 15 cmH(2)O) (all p < 0.001). However, only the interaction effect of endotracheal tubes with subglottic secretion drainage*high PEEP showed a significant effect on fluid leakage (p < 0.001), with the combination of endotracheal tube*no PEEP producing the greatest volume of fluid leak. Conclusion: Using endotracheal tubes with subglottic secretion drainage, high PEEP, and normal cuff pressure and avoiding excessive endotracheal tube movement during oral care reduced fluid leakage. This study provided strong evidence to inform practice on reducing microaspiration in ventilated patients. (C) 2019 Australian College of Critical Care Nurses Ltd. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:358 / 363
页数:6
相关论文
共 19 条
[1]   Assessment of the effectiveness of a ventilator associated pneumonia prevention bundle that contains endotracheal tube with subglottic drainage and cuff pressure monitorization [J].
Akdogan, Ozlem ;
Ersoy, Yasemin ;
Kuzucu, Cigdem ;
Gedik, Ender ;
Togal, Turkan ;
Yetkin, Funda .
BRAZILIAN JOURNAL OF INFECTIOUS DISEASES, 2017, 21 (03) :276-281
[2]   Prevention of ventilator-associated pneumonia: Use of the care bundle approach [J].
Alcan, Aliye Okgun ;
Korkmaz, Fatma Demir ;
Uyar, Mehmet .
AMERICAN JOURNAL OF INFECTION CONTROL, 2016, 44 (10) :E173-E176
[4]   The value of polyurethane-cuffed endotracheal tubes to reduce microaspiration and intubation-related pneumonia: a systematic review of laboratory and clinical studies [J].
Blot, Stijn I. ;
Rello, Jordi ;
Koulenti, Despoina .
CRITICAL CARE, 2016, 20
[5]   Subglottic Secretion Drainage and Objective Outcomes: A Systematic Review and Meta-Analysis [J].
Caroff, Daniel A. ;
Li, Lingling ;
Muscedere, John ;
Klompas, Michael .
CRITICAL CARE MEDICINE, 2016, 44 (04) :830-840
[6]   VENTILATOR-ASSOCIATED PNEUMONIA: CURRENT STATUS AND FUTURE RECOMMENDATIONS [J].
Efrati, Shai ;
Deutsch, Israel ;
Antonelli, Massimo ;
Hockey, Peter M. ;
Rozenblum, Ronen ;
Gurman, Gabriel M. .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2010, 24 (02) :161-168
[7]  
Ho K, 2009, ANAESTH INTENSIVE CA, V37, P680
[8]   Use of endotracheal tubes with subglottic secretion drainage reduces ventilator-associated pneumonia in trauma patients [J].
Hubbard, Jennifer L. ;
Veneman, Wade L. ;
Dirks, Rachel C. ;
Davis, James W. ;
Kaups, Krista L. .
JOURNAL OF TRAUMA AND ACUTE CARE SURGERY, 2016, 80 (02) :218-221
[9]  
Kim JT, 2009, CAN J ANAESTH, V56, P751, DOI 10.1007/s12630-009-9158-y
[10]   Intermittent Subglottic Secretion Drainage and Ventilator-associated Pneumonia A Multicenter Trial [J].
Lacherade, Jean-Claude ;
De Jonghe, Bernard ;
Guezennec, Pierre ;
Debbat, Karim ;
Hayon, Jan ;
Monsel, Antoine ;
Fangio, Pascal ;
de Vecchi, Corinne Appere ;
Ramaut, Cedric ;
Outin, Herve ;
Bastuji-Garin, Sylvie .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2010, 182 (07) :910-917