Evaluation of silicone as an endovascular stent membrane: In vivo canine studies

被引:4
|
作者
Fontaine, AB [1 ]
Borsa, JJ [1 ]
Hoffer, E [1 ]
Bloch, R [1 ]
So, C [1 ]
机构
[1] Univ Washington, Med Ctr, Dept Radiol, Seattle, WA 98195 USA
关键词
stents; stent grafts; prostheses; silicone;
D O I
10.1007/s00270-001-0014-x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose: Comparative evaluation of the biological effects of a silicone-covered stent versus a bare-metal stent, in an,animal model. Methods: Twelve. stent implants were placed in the iliac arteries of six adult dogs. Each animal received one 8-min x 20-mm silicone-covered stent (Permalume; Boston Scientific Vascular, Watertown, MA, USA), in the right iliac artery and one Wallstent (Boston Scientific Vascular) of the same diameter and length in the left iliac artery, during systemic anticoagulation. Angiography was performed before and after implantations. Animals were then allowed to recover and no platelet suppression was given during a 6-week interval, after which the animals were euthanized. The stented arteries were isolated and pressure-fixed in situ with 10% buffered formalin at a pressure of approximately 100 mmHg for a period of I hr. Two of 12 stented specimens were opened lengthwise and the luminal surfaces were photographed. Ten of 12 stented arterial segments were encased in methacrylate, then stained with hematoxylin and eosin. Neointimal thickness was quantified on histologic cross-section, for both bare and covered stents. The mean neointimal thicknesses were compared for significant difference using a student t-test. Results: All implants were widely patent at 6-week follow-up angiography. Histologic analysis showed bare metal stents covered by a thin uniform lining of neointima composed of, smooth muscle cells in a hyaline matrix (mean thickness of 189 +/- 47 mum). Silicone covered stents were devoid of neointima. There was no chronic thrombus or mature endothelium, noted anywhere upon the internal silicone surfaces of any of the specimens. There was no foreign body reaction to the silicone cover. Conclusion: Short-term implantation of a silicone-lined Wallstent in canine iliac arteries is well tolerated. Silicone appears to be inert at 6 weeks in this experimental application.
引用
收藏
页码:324 / 328
页数:5
相关论文
共 50 条
  • [1] Evaluation of Silicone as an Endovascular Stent Membrane: In Vivo Canine Studies
    Arthur B. Fontaine
    John J. Borsa
    Eric Hoffer
    Robert Bloch
    Corali So
    CardioVascular and Interventional Radiology, 2001, 24 : 324 - 328
  • [2] Evaluation of local abciximab delivery from the surface of a polymer-coated covered stent: In vivo canine studies
    Fontaine, AB
    Borsa, JJ
    Dos Passos, S
    Hoffer, EK
    Bloch, RD
    Starr, F
    So, C
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (04) : 487 - 492
  • [3] Thrombogenicity of various endovascular stent types: An in vitro evaluation
    Tepe, G
    Wendel, HP
    Khorchidi, S
    Schmehl, J
    Wiskirchen, J
    Pusich, B
    Claussen, CD
    Duda, SH
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 13 (10) : 1029 - 1035
  • [4] The mechanical characteristics and performance evaluation of a newly developed silicone airway stent (GINA stent)
    Hwa Sik Jung
    Ganghee Chae
    Jin Hyoung Kim
    Chui Yong Park
    Soyeoun Lim
    Soon Eun Park
    Ho Chang Kim
    Young Jae Lee
    Sung Kwon Kang
    Don Han Kim
    Yongjik Lee
    Taehoon Lee
    Scientific Reports, 11
  • [5] The mechanical characteristics and performance evaluation of a newly developed silicone airway stent (GINA stent)
    Jung, Hwa Sik
    Chae, Ganghee
    Kim, Jin Hyoung
    Park, Chui Yong
    Lim, Soyeoun
    Park, Soon Eun
    Kim, Ho Chang
    Lee, Young Jae
    Kang, Sung Kwon
    Kim, Don Han
    Lee, Yongjik
    Lee, Taehoon
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [6] Acute stent recoil: In vivo evaluation of new stent designs
    Danzi, GB
    Fiocca, L
    Capuano, C
    Predolini, S
    La Terza, C
    Baviera, L
    CIRCULATION, 2000, 102 (18) : 641 - 641
  • [7] Acute stent recoil: In vivo evaluation of different stent designs
    Danzi, GB
    Fiocca, L
    Capuano, C
    Predolini, S
    Quaini, E
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2001, 52 (02) : 147 - 153
  • [8] LABORATORY EVALUATION OF A NEW SILICONE MEMBRANE OXYGENATOR
    BIRNBAUM, D
    EISEMAN, B
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1972, 64 (03): : 441 - &
  • [9] MECHANICAL EVALUATION OF A CANINE INTERVERTEBRAL DISC SPACER - IN-SITU AND IN-VIVO STUDIES
    VUONOHAWKINS, M
    ZIMMERMAN, MC
    LEE, CK
    CARTER, FM
    PARSONS, JR
    LANGRANA, NA
    JOURNAL OF ORTHOPAEDIC RESEARCH, 1994, 12 (01) : 119 - 127
  • [10] Tissue Responses to Endovascular Stent Grafts for Saccular Abdominal Aortic Aneurysms in a Canine Model
    Kim, Hyun Beom
    Choi, Young Ho
    So, Young Ho
    Min, Seung-Kee
    Kim, Hyo-Cheol
    Kim, Young Il
    Park, Jae Hyung
    Chung, Jin Wook
    JOURNAL OF KOREAN MEDICAL SCIENCE, 2012, 27 (10) : 1170 - 1176