Pleth variability index to monitor the respiratory variations in the pulse oximeter plethysmographic waveform amplitude and predict fluid responsiveness in the operating theatre

被引:236
作者
Cannesson, M. [2 ]
Desebbe, O. [2 ]
Rosamel, P. [2 ]
Delannoy, B. [2 ]
Robin, J. [1 ]
Bastien, O. [2 ]
Lehot, J. -J. [2 ]
机构
[1] Univ Lyon 1, Louis Pradel Hosp, Hospices Civils Lyon, Dept Cardiac Surg,ERI 22, F-69365 Lyon, France
[2] Univ Lyon 1, Louis Pradel Hosp, Hospices Civils Lyon, Dept Anaesthesiol & Intens Care,ERI 22, F-69365 Lyon, France
关键词
equipment; pulse oximeter; fluids; i.v; heart; cardiac output; monitoring; cardiopulmonary; intraoperative;
D O I
10.1093/bja/aen133
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. Respiratory variations in pulse oximetry plethysmographic waveform amplitude (Delta POP) can predict fluid responsiveness in mechanically ventilated patients but cannot be easily assessed at the bedside. Pleth variability index (PVI) is a new algorithm allowing for automated and continuous monitoring of Delta POP. We hypothesized that PVI can predict fluid responsiveness in mechanically ventilated patients under general anaesthesia. Methods. Twenty-five patients were studied after induction of general anaesthesia. Haemodynamic data [cardiac index (CI), respiratory variations in arterial pulse pressure (Delta PP), Delta POP, and PVI] were recorded before and after volume expansion (500 ml of hetastarch 6%). Fluid responsiveness was defined as an increase in CI >= 15%. Results. Volume expansion induced changes in CI [2.0 (SD 0.9) to 2.5 (1.2) litre min(-1) m(-2); P < 0.01], Delta POP [15 (7)% to 8 (3)%; P < 0.01], and PVI [14 (7)% to 9 (3)%; P < 0.01]. Delta POP and PVI were higher in responders than in non-responders [19 (9)% vs 9 (4)% and 18 (6)% vs 8 (4)%, respectively; P < 0.01 for both]. A PVI > 14% before volume expansion discriminated between responders and non-responders with 81% sensitivity and 100% specificity. There was a significant relationship between PVI before volume expansion and change in CI after volume expansion (r=0.67; P < 0.01). Conclusions. PVI, an automatic and continuous monitor of Delta POP, can predict fluid responsiveness non-invasively in mechanically ventilated patients during general anaesthesia. This index has potential clinical applications.
引用
收藏
页码:200 / 206
页数:7
相关论文
共 37 条
  • [1] A novel algorithm to estimate the pulse pressure variation index ΔPP
    Aboy, M
    McNames, J
    Thong, T
    Phillips, CR
    Ellenby, MS
    Goldstein, B
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2004, 51 (12) : 2198 - 2203
  • [2] Awad AA, 2001, ANESTH ANALG, V92, P1483
  • [3] COMPLICATIONS OF PERCUTANEOUS RADIAL-ARTERY CANNULATION - OBJECTIVE PROSPECTIVE STUDY IN MAN
    BEDFORD, RF
    WOLLMAN, H
    [J]. ANESTHESIOLOGY, 1973, 38 (03) : 228 - 236
  • [4] Respiratory variations in pulse oximeter waveform amplitude are influenced by venous return in mechanically ventilated patients under general anaesthesia
    Cannesson, M.
    Desebbe, O.
    Hachemi, M.
    Jacques, D.
    Bastien, O.
    Lehot, J. -J
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2007, 24 (03) : 245 - 251
  • [5] Relation between respiratory variations in pulse oximetry plethysmographic waveform amplitude and arterial pulse pressure in ventilated patients
    Cannesson, M
    Besnard, C
    Durand, PG
    Bohé, J
    Jacques, D
    [J]. CRITICAL CARE, 2005, 9 (05): : R562 - R568
  • [6] Does the pleth variability index indicate the respiratory-induced variation in the plethysmogram and arterial pressure waveforms?
    Cannesson, Maxime
    Delannoy, Bertrand
    Morand, Antoine
    Rosamel, Pascal
    Attof, Yassin
    Bastien, Olivier
    Lehot, Jean-Jacques
    [J]. ANESTHESIA AND ANALGESIA, 2008, 106 (04) : 1189 - 1194
  • [7] Respiratory variations in pulse oximetry plethysmographic waveform amplitude to predict fluid responsiveness in the operating room
    Cannesson, Maxime
    Attof, Yassin
    Rosamel, Pascal
    Desebbe, Olivier
    Joseph, Pierre
    Metton, Olivier
    Bastien, Olivier
    Lehot, Jean-Jacques
    [J]. ANESTHESIOLOGY, 2007, 106 (06) : 1105 - 1111
  • [8] Randomised controlled trial investigating the influence of intravenous fluid titration using oesophageal Doppler monitoring during bowel surgery
    Conway, DH
    Mayall, R
    Abdul-Latif, MS
    Gilligan, S
    Tackaberry, C
    [J]. ANAESTHESIA, 2002, 57 (09) : 845 - 849
  • [9] THE INADEQUACY OF BINARY MODELS FOR THE CLINICAL REALITY OF 3-ZONE DIAGNOSTIC DECISIONS
    FEINSTEIN, AR
    [J]. JOURNAL OF CLINICAL EPIDEMIOLOGY, 1990, 43 (01) : 109 - 113
  • [10] Respiratory changes in aortic blood velocity as an indicator of fluid responsiveness in ventilated patients with septic shock
    Feissel, M
    Michard, F
    Mangin, I
    Ruyer, O
    Faller, JP
    Teboul, JL
    [J]. CHEST, 2001, 119 (03) : 867 - 873