The effect of tube voltage combination on image artefact and radiation dose in dual-source dual-energy CT: comparison between conventional 80/140 kV and 80/150 kV plus tin filter for gout protocol

被引:22
|
作者
Jeon, Ji Young [1 ]
Lee, Sheen-Woo [1 ]
Jeong, Yu Mi [1 ]
Baek, Han Joo [2 ]
机构
[1] Gachon Univ, Coll Med, Gil Med Ctr, Dept Radiol, 21 Namdong Daero,774beon Gil, Incheon 21565, South Korea
[2] Gachon Univ, Gil Med Ctr, Dept Internal Med, Div Rheumatol,Coll Med, 21 Namdong Daero,774beon Gil, Incheon 21565, South Korea
关键词
Gout; Tomography; X-ray computed; Tin; Artefacts; Radiation dosage; COMPUTED-TOMOGRAPHY; METAL ARTIFACTS; CLASSIFICATION; IDENTIFICATION; REDUCTION; AGREEMENT; DIAGNOSIS; QUALITY; UTILITY; BEAM;
D O I
10.1007/s00330-018-5622-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: In dual-source CT, dual-energy (DE) performance is affected by various X-ray tube voltage combinations with and without tin filter (Sn). The purpose of this study was to assess the utility of the 80/150 Sn kV voltage combination in terms of image artefact and radiation dose for DECT gout protocol, compared with the conventional 80/140 kV. Methods: Seventy-four patients with suspected gout who underwent dual-source DECT examinations scanned at 80/140 kV (n = 37) and at 80/150 Sn kV (n = 37) were included. Patients' age, sex, and serum uric acid levels were matched between the two groups. The types and incidence of image artefacts and radiation dose were evaluated. Results: The 80/150 Sn kV group had significantly fewer patients with artefacts, compared to the 80/140 kV group [11 (30 %) of 37 vs 35 (94.6 %) of 37, p < 0.001]. Except for the motion artefact, the rest of the artefactsskin, nail bed, submillimetre, motion, vascular, beam-hardening, clumpy artefact along tendonwere significantly less observed in the 80/150 Sn kV acquisitions. The dose-length product (DLP) and effective dose were significantly lower for the 80/150 Sn kV acquisitions compared with the 8s0/140 kV scans (DLP: 104.46 10.66 mGy<bold>cm vs </bold>344.70 56.39 mGy<bold>cm</bold>, p < 0.001; effective dose: 1.04 0.11 mSv vs 3.45 0.56 mSv, p < 0.001). Conclusions: The 80/150 Sn kV voltage combination in dual-source DECT system could be used as one of the artefact reduction methods while reducing radiation dose for gout protocol when compared to the conventional 80/140 kV. Key Points: DECT has emerged as the leading modality for non-invasive diagnosis of gout. Various X-ray tube voltage combinations are now feasible in dual-source DECT. The 80/150 Sn kV acquisition could facilitate artefact reduction in gout protocol.
引用
收藏
页码:1248 / 1257
页数:10
相关论文
共 17 条
  • [1] The effect of tube voltage combination on image artefact and radiation dose in dual-source dual-energy CT: comparison between conventional 80/140 kV and 80/150 kV plus tin filter for gout protocol
    Ji Young Jeon
    Sheen-Woo Lee
    Yu Mi Jeong
    Han Joo Baek
    European Radiology, 2019, 29 : 1248 - 1257
  • [2] Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter
    Choi, Seung Joon
    Ahn, Su Joa
    Park, So Hyun
    Park, Seong Ho
    Pak, Seong Yong
    Choi, Jae Won
    Shim, Young Sup
    Jeong, Yu Mi
    Kim, Bohyun
    PLOS ONE, 2020, 15 (09):
  • [3] Tube Current Modulation Between Single- and Dual-Energy CT With a Second-Generation Dual-Source Scanner: Radiation Dose and Image Quality
    Matsubara, Kosuke
    Takata, Tadanori
    Kobayashi, Masanao
    Kobayashi, Satoshi
    Koshida, Kichiro
    Gabata, Toshifumi
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 207 (02) : 354 - 361
  • [4] Dual energy CT angiography of the carotid arteries: quality, bone subtraction, and radiation dosage using tube voltage 80/140kV versus 100/140kV
    Santos Armentia, E.
    Tardaguila de la Fuente, G.
    Castellon Plaza, D.
    Delgado Sanchez-Gracian, C.
    Prada Gonzalez, R.
    Fernandez Fernandez, L.
    Tardaguila Montero, F.
    RADIOLOGIA, 2014, 56 (03): : 241 - 246
  • [5] Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT
    Wichmann, Julian L.
    Hardie, Andrew D.
    Schoepf, U. Joseph
    Felmly, Lloyd M.
    Perry, Jonathan D.
    Varga-Szemes, Akos
    Mangold, Stefanie
    Caruso, Damiano
    Canstein, Christian
    Vogl, Thomas J.
    De Cecco, Carlo N.
    EUROPEAN RADIOLOGY, 2017, 27 (02) : 642 - 650
  • [6] Single- and dual-energy CT pulmonary angiography using second- and third-generation dual-source CT systems: comparison of radiation dose and image quality
    Lenga, Lukas
    Trapp, Franziska
    Albrecht, Moritz H.
    Wichmann, Julian L.
    Johnson, Addison A.
    Yel, Ibrahim
    D'Angelo, Tommaso
    Booz, Christian
    Vogl, Thomas J.
    Martin, Simon S.
    EUROPEAN RADIOLOGY, 2019, 29 (09) : 4603 - 4612
  • [7] Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT
    Julian L. Wichmann
    Andrew D. Hardie
    U. Joseph Schoepf
    Lloyd M. Felmly
    Jonathan D. Perry
    Akos Varga-Szemes
    Stefanie Mangold
    Damiano Caruso
    Christian Canstein
    Thomas J. Vogl
    Carlo N. De Cecco
    European Radiology, 2017, 27 : 642 - 650
  • [8] Single- and dual-energy CT pulmonary angiography using second- and third-generation dual-source CT systems: comparison of radiation dose and image quality
    Lukas Lenga
    Franziska Trapp
    Moritz H. Albrecht
    Julian L. Wichmann
    Addison A. Johnson
    Ibrahim Yel
    Tommaso D’Angelo
    Christian Booz
    Thomas J. Vogl
    Simon S. Martin
    European Radiology, 2019, 29 : 4603 - 4612
  • [9] Third-generation dual-source CT of the neck using automated tube voltage adaptation in combination with advanced modeled iterative reconstruction: evaluation of image quality and radiation dose
    Scholtz, Jan-Erik
    Wichmann, Julian L.
    Huesers, Kristina
    Albrecht, Moritz H.
    Beeres, Martin
    Bauer, Ralf W.
    Vogl, Thomas J.
    Bodelle, Boris
    EUROPEAN RADIOLOGY, 2016, 26 (08) : 2623 - 2631
  • [10] Comparison of image quality and radiation dose between split-filter dual-energy images and single-energy images in single-source abdominal CT
    Euler, Andre
    Obmann, Markus M.
    Szucs-Farkas, Zsolt
    Mileto, Achille
    Zaehringer, Caroline
    Falkowski, Anna L.
    Winkel, David J.
    Marin, Daniele
    Stieltjes, Bram
    Krauss, Bernhard
    Schindera, Sebastian T.
    EUROPEAN RADIOLOGY, 2018, 28 (08) : 3405 - 3412