SPECT collimator dependency of scatter and validation of transmission-dependent scatter compensation methodologies

被引:0
|
作者
Kim, KM [1 ]
Watabe, H
Shidahara, M
Ishida, Y
Iida, H
机构
[1] Natl Cardiovasc Ctr, Res Inst, Dept Invest Radiol, Suita, Osaka 5658565, Japan
[2] Natl Cardiovasc Ctr, Dept Radiol, Suita, Osaka 5658565, Japan
关键词
scatter correction; SPECT collimator;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Scatter correction is important for absolute quantification in SPECT. Among the approaches taken to achieve scatter correction, some methods using transmission information for scatter estimation require the scatter models based on the physical characteristics of the scattered photons. The influence of the SPECT collimator on the characteristics of the scattered photon, however, has not been well studied yet. In this paper, we investigate the relationship between scatter and collimator design for the use of the transmission-dependent convolution subtraction (TDCS) method. Monte Carlo simulations with various collimator acceptance angles and experiments with four collimators of two SPECT cameras were performed to obtain the scatter fraction (SF) and the line-spread function (LSF) of Tc-99m. The experiments with I-123 also used six collimators of three SPECT systems to obtain the SF and the LSF. In the simulations, the SF of Tc-99m did not change with the collimator acceptance angles. The LSFs measured for Tc-99m showed an excellent agreement between collimators. In the experiments for I-123, the SF was practically the same for the six collimators, although a small difference could be observed due to the septal penetration. Also, the shape of the LSF was very similar between the collimators, and the counts in the background region of reconstructed images did not differ when using different sets of TDCS parameters. These results reveal that scatter is independent of SPECT collimator design for Tc-99m even I-123, and that the TDCS could be applicable to scatter correction of I-123 using a unique collimator-independent set of parameters.
引用
收藏
页码:689 / 696
页数:8
相关论文
共 12 条
  • [1] Three-dimensional SPECT reconstruction with transmission-dependent scatter correction
    Antti Sohlberg
    Hiroshi Watabe
    Hidehiro Iida
    Annals of Nuclear Medicine, 2008, 22 : 549 - 556
  • [2] Three-dimensional SPECT reconstruction with transmission-dependent scatter correction
    Sohlberg, Antti
    Watabe, Hiroshi
    Iida, Hidehiro
    ANNALS OF NUCLEAR MEDICINE, 2008, 22 (07) : 549 - 556
  • [3] Evaluation of collimator choice and scatter correction on 123I SPECT images
    Yang, Yu-Wen
    Chen, Jyh-Cheng
    Chang, Chia-Jung
    Cheng, Cheng-Yi
    Wang, Shyh-Jen
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 584 (01) : 204 - 211
  • [4] Contribution of scatter and attenuation compensation to SPECT images of nonuniformly distributed brain activities
    Kim, KM
    Varrone, A
    Watabe, H
    Shidahara, M
    Fujita, M
    Innis, RB
    Iida, H
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (04) : 512 - 519
  • [5] A New Approach for Scatter Removal and Attenuation Compensation from SPECT/CT Images
    Oloomi, Shabnam
    Eskandari, Hadi Noori
    Zakavi, Seyed Rasoul
    Knoll, Peter
    Kalantari, Faraz
    Saffar, Mohsen Hajizadeh
    IRANIAN JOURNAL OF BASIC MEDICAL SCIENCES, 2013, 16 (11) : 1181 - 1189
  • [6] Evaluation of scatter compensation methods by their effects on parameter estimation from SPECT projections
    Moore, SC
    Kijewski, MF
    Müller, SP
    Rybicki, F
    Zimmerman, RE
    MEDICAL PHYSICS, 2001, 28 (02) : 278 - 287
  • [7] Impact of attenuation, scatter and depth dependent collimator response correction in 3-D cardiac SPECT imaging: a Monte Carlo investigation
    Ma, TY
    Jin, YJ
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3594 - 3598
  • [8] A TRANSMISSION-MAP BASED SCATTER CORRECTION TECHNIQUE FOR SPECT IN INHOMOGENEOUS-MEDIA
    WELCH, A
    GULLBERG, GT
    CHRISTIAN, PE
    DATZ, FL
    MORGAN, T
    MEDICAL PHYSICS, 1995, 22 (10) : 1627 - 1635
  • [9] Transmission-dependent convolution subtraction of 99mTc-HMPAO rCBF SPECT -: A Monte Carlo study
    Larsson, A
    Johansson, L
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (01) : 231 - 237
  • [10] A method to correct myocardial SPECT with Tc-99m compounds for scatter and attenuation without transmission
    Kupferschlager, J
    Muller, B
    Schulz, G
    Kaiser, HJ
    Wagenknecht, G
    Bares, R
    Bull, U