Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films

被引:73
作者
Wen, Ning [1 ]
Lu, Siming [1 ]
Kim, Jinkoo [1 ]
Qin, Yujiao [1 ]
Huang, Yimei [1 ]
Zhao, Bo [1 ]
Liu, Chang [1 ]
Chetty, Indrin J. [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Radiat Oncol, 2799 West Grand Blvd, Detroit, MI 48202 USA
来源
RADIATION ONCOLOGY | 2016年 / 11卷
关键词
Stereotactic radiosurgery; Stereotactic body radiation therapy; Gafchromic films; Quality assurance; IONIZATION-CHAMBER ARRAYS; RADIOCHROMIC FILM; PHOTON BEAMS; SCANNERS; VERIFICATION;
D O I
10.1186/s13014-016-0709-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this study is to evaluate the dosimetric uncertainty associated with Gafchromic (TM) (EBT3) films and establish a practical and efficient film dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). Method and materials: EBT3 films were irradiated at each of seven different dose levels between 1 and 15 Gy with open fields and standard deviations of dose maps were calculated at each color channel for evaluation. A scanner non-uniform response correction map was built by registering and comparing film doses to the reference ion chamber array-based dose map delivered with the same doses. To determine the temporal dependence of EBT3 films, the average correction factors of different dose levels as a function of time were evaluated up to 4 days after irradiation. An integrated film dosimetry protocol was developed for dose calibration, calibration curve fitting, dose mapping, and profile/gamma analysis. Patient specific quality assurance (PSQA) was performed for 83 SRS/SBRT treatment plans, and analysis of the measurements and calculations are presented here. Results: The scanner response varied within 1 % for the field sizes less than 5 x 5 cm(2), and up to 5 % for the field sizes of 10 x 10 cm(2) for all color channels. The scanner correction method was able to remove visually evident, irregular detector responses for larger field sizes. The dose response of the film changed rapidly (similar to 10 %) in the first two hours and became smooth plateaued afterwards, similar to 3 % change between 2 and 24 h. The uncertainties were approximately 1.5, 1.7 and 4.8 % over the dose range of 3 similar to 15 Gy for the red, green and blue channels. The green channel showed very high sensitivity and low uncertainty in the dose range between 10 and 15 Gy, which is suitable for SRS/SBRT commissioning and PSQA. The difference between the calculated dose and measured dose of ion chamber measurement at isocenter was -0.64 +/- 2.02 for all plans, corresponding to a 95 % confidence interval of (-1.09, -0.26). The percentage of points passing the 3 %/1 mm gamma criteria in absolute dose, averaged over all tests was 95.0 +/- 4.2. Conclusion: We have developed the EBT3 films based dosimetry protocol to obtain absolute dose values. The overall uncertainty has been established to be 1.5 % for SRS and SBRT PSQA.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] A quality assurance program in stereotactic radiosurgery using the gamma knife unit
    Stuecklschweiger, GE
    Feichtinger, K
    STRAHLENTHERAPIE UND ONKOLOGIE, 1998, 174 : 43 - 46
  • [32] High precision film dosimetry with GAFCHROMIC® films for quality assurance especially when using small fields
    Mack, A
    Mack, G
    Weltz, D
    Scheib, SG
    Böttcher, HD
    Seifert, V
    MEDICAL PHYSICS, 2003, 30 (09) : 2399 - 2409
  • [33] Evaluation of patient-specific IMRT quality assurance and point dose measurement for complex head and neck and brain cancer using Gafchromic EBT3 film
    Gedam, Varsha R.
    Pradhan, Anirudh
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (02) : 164 - 170
  • [34] Characterisation of a synthetic diamond detector for end-to-end dosimetry in stereotactic body radiotherapy and radiosurgery
    Shaw, Maddison
    Lye, Jessica
    Alves, Andrew
    Keehan, Stephanie
    Lehmann, Joerg
    Hanlon, Maximilian
    Kenny, John
    Baines, John
    Porumb, Claudiu
    Geso, Moshi
    Brown, Rhonda
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2021, 20 : 40 - 45
  • [35] A feasibility study of stereotactic radiosurgery/stereotactic body radiotherapy/stereotactic ablative radiotherapy practice using tomoedge in helical tomotherapy for lung, liver, and spine targets
    Sresty, N. V. N. Madhusudhana
    Raju, A. Krishnam
    Kumar, G. Deleep
    Rohit, S.
    Reddy, B. Nagarjuna
    Sahithya, V. C.
    Reddy, B. Devender
    Mohd, Yakub
    Rushdi, Tasneem
    Bajwa, Harjoth
    Aparna, S.
    JOURNAL OF MEDICAL PHYSICS, 2021, 46 (03) : 204 - 210
  • [36] In vivo dosimetry using EBT3 and EBT-XD radiochromic films for high-energy electron beam in hypofractionation keloid radiotherapy
    Lin, Yu -Fang
    Hsieh, Chen-Hsi
    Tien, Hui-Ju
    Lai, Lu-Han
    Hsu, Shih-Ming
    Shueng, Pei-Wei
    RADIATION PHYSICS AND CHEMISTRY, 2024, 218
  • [37] PATIENT SPECIFIC QUALITY ASSURANCE FOR LUNG CANCER STEREOTACTIC ABLATIVE BODY RADIOTHERAPY
    Hardcastle, Nicholas
    Clements, Natalie
    Siva, Shankar
    Ball, David
    Kron, Tomas
    JOURNAL OF THORACIC ONCOLOGY, 2013, 8 : S1139 - S1140
  • [38] In-vivo dosimetry in total body irradiation using GafchromicTM EBT-3 film: A retrospective analysis for quality assurance
    Dhivya, S.
    Anuradha, C.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2023, 21 (02): : 311 - 317
  • [39] Beam quality and dose rate dependency of Gafchromic EBT3 film irradiated with therapeutic megavolt photon beams
    Sharma, Megha
    Singh, Ranjit
    Robert, N.
    Trivedi, G.
    Tomar, P.
    Oinam, A. S.
    RADIATION MEASUREMENTS, 2021, 146
  • [40] In-vivo dosimetry with EBT-3 gafchromic films and thermoluminescent dosimeters in breast cancer radiotherapy treatments
    Fokou, Mvoufo
    Hasford, Francis
    Ndontchueng, Maurice M.
    Tagoe, Samuel A.
    Njantang, Ruth N.
    Dery, Theresa B.
    Simo, Augustin
    Larbi, Kofi O. A.
    Mantheaw, Philip
    Eduful, Ernest
    Motapon, Ousmanou
    HEALTH AND TECHNOLOGY, 2023, 13 (04) : 585 - 600