Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films

被引:72
|
作者
Wen, Ning [1 ]
Lu, Siming [1 ]
Kim, Jinkoo [1 ]
Qin, Yujiao [1 ]
Huang, Yimei [1 ]
Zhao, Bo [1 ]
Liu, Chang [1 ]
Chetty, Indrin J. [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Radiat Oncol, 2799 West Grand Blvd, Detroit, MI 48202 USA
来源
RADIATION ONCOLOGY | 2016年 / 11卷
关键词
Stereotactic radiosurgery; Stereotactic body radiation therapy; Gafchromic films; Quality assurance; IONIZATION-CHAMBER ARRAYS; RADIOCHROMIC FILM; PHOTON BEAMS; SCANNERS; VERIFICATION;
D O I
10.1186/s13014-016-0709-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this study is to evaluate the dosimetric uncertainty associated with Gafchromic (TM) (EBT3) films and establish a practical and efficient film dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). Method and materials: EBT3 films were irradiated at each of seven different dose levels between 1 and 15 Gy with open fields and standard deviations of dose maps were calculated at each color channel for evaluation. A scanner non-uniform response correction map was built by registering and comparing film doses to the reference ion chamber array-based dose map delivered with the same doses. To determine the temporal dependence of EBT3 films, the average correction factors of different dose levels as a function of time were evaluated up to 4 days after irradiation. An integrated film dosimetry protocol was developed for dose calibration, calibration curve fitting, dose mapping, and profile/gamma analysis. Patient specific quality assurance (PSQA) was performed for 83 SRS/SBRT treatment plans, and analysis of the measurements and calculations are presented here. Results: The scanner response varied within 1 % for the field sizes less than 5 x 5 cm(2), and up to 5 % for the field sizes of 10 x 10 cm(2) for all color channels. The scanner correction method was able to remove visually evident, irregular detector responses for larger field sizes. The dose response of the film changed rapidly (similar to 10 %) in the first two hours and became smooth plateaued afterwards, similar to 3 % change between 2 and 24 h. The uncertainties were approximately 1.5, 1.7 and 4.8 % over the dose range of 3 similar to 15 Gy for the red, green and blue channels. The green channel showed very high sensitivity and low uncertainty in the dose range between 10 and 15 Gy, which is suitable for SRS/SBRT commissioning and PSQA. The difference between the calculated dose and measured dose of ion chamber measurement at isocenter was -0.64 +/- 2.02 for all plans, corresponding to a 95 % confidence interval of (-1.09, -0.26). The percentage of points passing the 3 %/1 mm gamma criteria in absolute dose, averaged over all tests was 95.0 +/- 4.2. Conclusion: We have developed the EBT3 films based dosimetry protocol to obtain absolute dose values. The overall uncertainty has been established to be 1.5 % for SRS and SBRT PSQA.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Quality assurance in stereotactic radiosurgery/radiotherapy according to DIN 6875-1
    Mack, A
    Mack, G
    Scheib, S
    Czempiel, H
    Kreiner, HJ
    Lomax, NJ
    Gianolini, S
    Rieker, M
    Weltz, D
    Wolff, R
    Muacevic, A
    Wowra, B
    Böttcher, HD
    Seifert, V
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2004, 82 (5-6) : 235 - 243
  • [22] Surface dose measurement with Gafchromic EBT3 film for intensity modulated radiotherapy technique
    Akbas, Ugur
    Kesen, Nazmiye Donmez
    Koksal, Canan
    Okutan, Murat
    Demir, Bayram
    Becerir, Hatice Bilge
    3RD INTERNATIONAL CONFERENCE ON THEORETICAL AND EXPERIMENTAL STUDIES IN NUCLEAR APPLICATIONS AND TECHNOLOGY (TESNAT 2017), 2017, 154
  • [23] Quality Assurance of Spine Volumetric Modulated Arc Therapy with Flattening Filter Free Beams Using Gafchromic EBT3 Films
    Choi, Y.
    Cho, B.
    Kwak, J.
    Ahn, S.
    Choi, E.
    Kim, J.
    Song, S.
    Yoon, S.
    Kim, Y.
    Kim, S.
    Park, J.
    MEDICAL PHYSICS, 2014, 41 (06) : 285 - 285
  • [24] Measuring diffuse ultraviolet exposures using Gafchromic EBT3 films
    Tsang, P. Y.
    Chan, P. M.
    Yu, K. N.
    RESULTS IN PHYSICS, 2017, 7 : 1492 - 1493
  • [25] GafChromic® EBT3 films for patient specific IMRT QA using amultichannel approach
    Marrazzo, Livia
    Zani, Margherita
    Pallotta, Stefania
    Arilli, Chiara
    Casati, Marta
    Compagnucci, Antonella
    Talamonti, Cinzia
    Bucciolini, Marta
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (08): : 1035 - 1042
  • [26] Point dose verification of Cranial Stereotactic Radiosurgery using micro Ionization Chamber and EBT3 film for 6MV FF and FFF beams in Varian TrueBeam® LINAC
    Mamballikalam, Gopinath
    Senthilkumar, S.
    Basith, Ahamed P. M.
    Inipully, Rohit
    Jayadevan, P. M.
    Clinto, C. O.
    Bos, Jaon R. C.
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2020, 26 (03): : 135 - 142
  • [27] Peripheral and surface dose assessment using diode and Gafchromic EBT3 films dosimeters for different radiotherapy techniques
    AL-Shareef, Jamal M.
    Attalla, Ehab M.
    El-Gebaly, Reem H.
    Deiab, Nashaat A.
    Mokhtar, Maha H.
    Fathy, Mohamed M.
    RADIATION PHYSICS AND CHEMISTRY, 2022, 198
  • [28] A study of surface dosimetry for breast cancer radiotherapy treatments using Gafchromic EBT2 film
    Nakano, Masahiro
    Hill, Robin F.
    Whitaker, May
    Kim, Jung-Ha
    Kuncic, Zdenka
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (03): : 83 - 97
  • [29] Adaptive calibration of Gafchromic EBT3 film using generalized additive neural networks
    Chang, Liyun
    Chen, Pang-Yu
    Yeh, Shyh-An
    Kang, Chen-Lin
    Su, Chun-Ting
    Lee, Tsair-Fwu
    SCIENTIFIC REPORTS, 2025, 15 (01):
  • [30] Radiochromic film dosimetry: Considerations on precision and accuracy for EBT2 and EBT3 type films
    Dreindl, Ralf
    Georg, Dietmar
    Stock, Markus
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2014, 24 (02): : 153 - 163