MCNP Simulation of Radiation Doses Distributions in a Water Phantoms Simulating Interventional Radiology Patients

被引:0
作者
He, Wenjun [1 ]
Mah, Eugene [2 ]
Huda, Walter [2 ]
Selby, Bayne [2 ]
Yao, Hai [1 ]
机构
[1] Clemson Univ, Clemson MUSC Bioengn Program, Dept Bioengn, Charleston, SC 29425 USA
[2] Med Univ South Carolina, Dept Radiol & Radiol Sci, Charleston, SC 29425 USA
来源
MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING | 2011年 / 7961卷
关键词
Interventional radiology; radiation dose; Monte Carlo; Air kerma; primary radiation; scatter radiation; energy imparted; ALARA; FLUOROSCOPICALLY GUIDED PROCEDURES; MONTE-CARLO; SKIN INJURIES; PART I; RAD-IR; TRANSMISSION; SCATTER;
D O I
10.1117/12.878271
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Purpose: To investigate the dose distributions in water cylinders simulating patients undergoing Interventional Radiological examinations. Method: The irradiation geometry consisted of an x-ray source, dose-area-product chamber, and image intensifier as currently used in Interventional Radiology. Water cylinders of diameters ranging between 17 and 30 cm were used to simulate patients weighing between 20 and 90 kg. X-ray spectra data with peak x-ray tube voltages ranging from 60 to 120 kV were generated using XCOMP3R. Radiation dose distributions inside the water cylinder (D-w) were obtained using MCNP5. The depth dose distribution along the x-ray beam central axis was normalized to free-in-air air kerma (AK) that is incident on the phantom. Scattered radiation within the water cylinders but outside the directly irradiated region was normalized to the dose at the edge of the radiation field. The total absorbed energy to the directly irradiated volume (E-p) and indirectly irradiated volume (E-s) were also determined and investigated as a function of x-ray tube voltage and phantom size. Results: At 80 kV, the average D-w/AK near the x-ray entrance point was 1.3. The ratio of D-w near the entrance point to Dw near the exit point increased from similar to 26 for the 17 cm water cylinder to similar to 290 for the 30 cm water cylinder. At 80 kV, the relative dose for a 17 cm water cylinder fell to 0.1% at 49 cm away from the central ray of the x-ray beam. For a 30 cm water cylinder, the relative dose fell to 0.1% at 53 cm away from the central ray of the x-ray beam. At a fixed x-ray tube voltage of 80 kV, increasing the water cylinder diameter from 17 to 30 cm increased the E-s/(E-p+E-s) ratio by about 50%. At a fixed water cylinder diameter of 24 cm, increasing the tube voltage from 60 kV to 120 kV increased the E-s/(E-p+E-s) ratio by about 12%. The absorbed energy from scattered radiation was between 20-30% of the total energy absorbed by the water cylinder, and was affected more by patient size than x-ray beam energy. Conclusion: MCNP offers a powerful tool to study the absorption and transmission of x-ray energy in phantoms that can be designed to represent patients undergoing Interventional Radiological procedures. This ability will permit a systematic investigation of the relationship between patient dose and diagnostic image quality, and thereby keep patient doses As Low As Reasonably Achievable (ALARA).
引用
收藏
页数:10
相关论文
共 18 条
  • [1] Patient Radiation Doses in Interventional Radiology Procedures: Comparison of Fluoroscopy Dose Indices between the American College of Radiology Dose Index Registry-Fluoroscopy Pilot and the Radiation Doses in Interventional Radiology Study
    Jones, A. Kyle
    Wunderle, Kevin A.
    Fruscello, Tom
    Simanowith, Michael
    Cline, Brendan
    Dharmadhikari, Shalmali
    Duan, Xinhui
    Durack, Jeremy C.
    Hirschl, David
    Kim, Don-Soo
    Mahmood, Usman
    Mann, Steve D.
    Martin, Charles
    Metwalli, Zeyad
    Moirano, Jeffrey M.
    Neill, Rebecca A.
    Newsome, Janice
    Padua, Horacio
    Schoenfeld, Alan H.
    Miller, Donald L.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2023, 34 (04)
  • [2] Estimating Radiation Doses to the Skin from Interventional Radiology Procedures for a Patient Population with Cancer
    Dauer, Lawrence T.
    Thornton, Raymond
    Erdi, Yusuf
    Ching, Hung
    Hamacher, Klaus
    Boylan, Daniel C.
    Williamson, Matthew J.
    Balter, Stephen
    Germain, Jean St.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 20 (06) : 782 - 788
  • [3] Staff Radiation Doses to the Lower Extremities in Interventional Radiology
    C. P. Shortt
    H. Al-Hashimi
    L. Malone
    M. J. Lee
    CardioVascular and Interventional Radiology, 2007, 30 : 1206 - 1209
  • [4] Staff radiation doses to the lower extremities in interventional radiology
    Shortt, C. P.
    Al-Hashimi, H.
    Malone, L.
    Lee, M. J.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 30 (06) : 1206 - 1209
  • [5] Patient radiation doses and reference levels in pediatric interventional radiology
    Bouchra Habib Geryes
    Adeline Bak
    Julie Lachaux
    Augustin Ozanne
    Nathalie Boddaert
    Francis Brunelle
    Olivier Naggara
    Guillaume Saliou
    European Radiology, 2017, 27 : 3983 - 3990
  • [6] Patient radiation doses and reference levels in pediatric interventional radiology
    Geryes, Bouchra Habib
    Bak, Adeline
    Lachaux, Julie
    Ozanne, Augustin
    Boddaert, Nathalie
    Brunelle, Francis
    Naggara, Olivier
    Saliou, Guillaume
    EUROPEAN RADIOLOGY, 2017, 27 (09) : 3983 - 3990
  • [7] Radiation protection of patients in interventional radiology
    Roche, A.
    JOURNAL DE RADIOLOGIE, 2010, 91 (11): : 1231 - 1235
  • [8] An audit of patient radiation doses in interventional radiology at a South African hospital
    Slave, Oneile
    Mahomed, Nasreen
    SA JOURNAL OF RADIOLOGY, 2023, 27 (01):
  • [9] Exposures in interventional radiology using Monte Carlo simulation coupled with virtual anthropomorphic phantoms
    Santos, William S.
    Neves, Lucio P.
    Perini, Ana P.
    Belinato, Walmir
    Caldas, Linda V. E.
    Carvalho, Alberico B., Jr.
    Maia, Ana F.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (08): : 929 - 933
  • [10] Mixed Reality Visualization of Radiation Dose for Health Professionals and Patients in Interventional Radiology
    Takata, Takeshi
    Nakabayashi, Susumu
    Kondo, Hiroshi
    Yamamoto, Masayoshi
    Furui, Shigeru
    Shiraishi, Kenshiro
    Kobayashi, Takenori
    Oba, Hiroshi
    Okamoto, Takahide
    Kotoku, Jun'ichi
    JOURNAL OF MEDICAL SYSTEMS, 2021, 45 (04)