Trace anion determination in concentrated hydrofluoric acid solutions by two-dimensional ion chromatography - II. Method performance study with a hydroxide eluent and a low noise suppressor
被引:17
作者:
Vermeiren, K
[1
]
机构:
[1] SOLVAY Tes & Technol, Analyt Technol Competence Ctr, B-1120 Brussels, Belgium
关键词:
two-dimensional ion chromatography;
trace analysis;
hydrofluoric acid;
SEMI;
method validation;
D O I:
10.1016/j.chroma.2005.05.098
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
The two-dimensional ion exclusion chromatography/ion chromatography (ICE-IC) approach is considered to be the method of choice for the determination of trace anions in concentrated hydrofluoric acid (48-50%, w/w). In order to achieve lower detection limits, this method was for the first time used with electrolytically generated and purified hydroxide eluents in combination with a low noise electrochemical suppressor. Compared to carbonate based eluents, the achieved gain in peak height sensitivity for chloride, sulfate, nitrate and phosphate is a factor of 6, 7, 16 and 13, respectively. The instrumental detection limits, based on the background noise, are 2, 0.2, 0.4 and 1.4 mu g/kg HF 50% (w/w) for the same anions. Their method detection limits, calculated according to SEMI, are all within the 6-10 mu g/kg HF 50% (w/w) range and thus at least 10 times lower than the current Tier C grade requirements. The chromatographic run time could be shortened with some 10min by the use of a relatively fast high-capacity hydroxide selective anion exchange column. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:66 / 73
页数:8
相关论文