Cone-beam-CT guided radiation therapy:: technical implementation

被引:198
作者
Létourneau, D
Wong, JW
Oldham, M
Gulam, M
Watt, L
Jaffray, DA
Siewerdsen, JH
Martinez, AA
机构
[1] Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[2] William Beaumont Hosp, Dept Radiat Oncol, Royal Oak, MI USA
[3] Univ Toronto, Dept Med Biophys, Ontario Canc Inst, Toronto, ON, Canada
关键词
cone-beam CT; image quality; image-guided therapy;
D O I
10.1016/j.radonc.2005.03.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: X-ray volumetric imaging system (XVI) mounted on a linear accelerator is available for image guidance applications. In preparation for clinical implementation, phantom and patient imaging studies were conducted to determine the irradiation parameters that would trade-off image quality, patient dose and scanning time. Patients and methods: The XVI image quality and imaging dose were benchmarked against those obtained with a helical CT scanner for a head and body phantom. The irradiation parameters were varied including the total imaging dose, number of projections, field of view, reconstruction resolution and use of a scatter rejection grid. We characterized the image quality based on relative contrast, noise, contrast to noise ratio (CNR) and point spread function (PSF). XVI scans of pelvis, head and neck and lung patients were acquired and submitted to a range of observers to identify the favorable reconstruction parameters. Results: Phantom studies have demonstrated that a scatter rejection grid reduces photon scattering and improves the image uniformity. For the body phantom, the helical CT and the wide field XVI technique produce similar image quality, with surface doses of 0.025 and 0.044 Gy respectively. We have demonstrated that the local tomography technique improves the image contrast and the CNR while reducing the skin dose by 40-50% compared to the wide field technique. Clinical scans of head and neck, lung and prostate patients present good soft tissue contrast and excellent bone definition. Conclusions: With adjustment of irradiation parameters and an imaging surface dose of less than 0.05 Gy, high quality XVI images can be obtained for a phantom simulating the body thickness. XVI is currently feasible for image-guided treatments of head and neck, torso and pelvic areas using soft tissue and bony structures. (C) 2005 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:279 / 286
页数:8
相关论文
共 43 条
  • [1] *AAPM, 1993, 39 AAPM
  • [2] A preliminary investigation of local tomography for megavoltage CT imaging
    Anastasio, MA
    Shi, DX
    Pan, XC
    Pelizzari, C
    Munro, P
    [J]. MEDICAL PHYSICS, 2003, 30 (11) : 2969 - 2980
  • [3] Initial performance evaluation of an indirect-detection, active matrix flat-panel imager (AMFPI) prototype for megavoltage imaging
    Antonuk, LE
    El-Mohri, Y
    Huang, WD
    Jee, KW
    Siewerdsen, JH
    Maolinbay, M
    Scarpine, VE
    Sandler, H
    Yorkston, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (02): : 437 - 454
  • [4] MEASUREMENT OF PROSTATE MOVEMENT OVER THE COURSE OF ROUTINE RADIOTHERAPY USING IMPLANTED MARKERS
    BALTER, JM
    SANDLER, HM
    LAM, K
    BREE, RL
    LICHTER, AS
    TENHAKEN, RK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (01): : 113 - 118
  • [5] Real-time three-dimensional motion analysis for patient positioning verification
    Baroni, G
    Ferrigno, G
    Orecchia, R
    Pedotti, A
    [J]. RADIOTHERAPY AND ONCOLOGY, 2000, 54 (01) : 21 - 27
  • [6] The university of Florida frameless high-precision stereotactic radiotherapy system
    Bova, FJ
    Buatti, JM
    Friedman, WA
    Mendenhall, WM
    Yang, CC
    Liu, C
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04): : 875 - 882
  • [7] CONE-BEAM CT FOR RADIOTHERAPY APPLICATIONS
    CHO, PS
    JOHNSON, RH
    GRIFFIN, TW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11) : 1863 - 1883
  • [8] Characterization of a fluoroscopic imaging system for kV and MV radiography
    Drake, DG
    Jaffray, DA
    Wong, JW
    [J]. MEDICAL PHYSICS, 2000, 27 (05) : 898 - 905
  • [9] Effect of scatter radiation on image noise in cone beam CT
    Endo, M
    Tsunoo, T
    Nakamori, N
    [J]. MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 : 514 - 521
  • [10] Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: Image-based correction of gantry motion nonidealities
    Fahrig, R
    Holdsworth, DW
    [J]. MEDICAL PHYSICS, 2000, 27 (01) : 30 - 38