Telemonitoring with implantable electronic devices in young patients with congenital heart diseases

被引:13
|
作者
Zartner, Peter A. [1 ]
Toussaint-Goetz, Nicole [1 ]
Photiadis, Joachim [2 ]
Wiebe, Walter [1 ]
Schneider, Martin B. [1 ]
机构
[1] Deutsch Kinderherzzentrum Sankt Augustin, Dept Cardiol, D-53757 St Augustin, Germany
[2] Deutsch Kinderherzzentrum Sankt Augustin, Dept Cardiothorac Surg, D-53757 St Augustin, Germany
来源
EUROPACE | 2012年 / 14卷 / 07期
关键词
Children; Infants; Congenital heart disease; Pacemaker; Defibrillator; Automated telemonitoring; Network; CARDIAC RESYNCHRONIZATION THERAPY; FOLLOW-UP; CARDIOVERTER-DEFIBRILLATORS; PACEMAKER; CHILDREN; REMOTE; ADULTS; EXPERIENCE; ICD; ADOLESCENTS;
D O I
10.1093/europace/eur434
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In paediatric patients with a pacemaker (PM) or an implantable cardioverter defibrillator (ICD) device, interrogation during in-clinic visits is usually required to obtain information on arrhythmias or system failures. An automated telemonitoring system tracking patient- and system-related parameters provides this information on a daily basis and might assist patient management. This retrospective analysis evaluates telemetric data obtained from 48 devices implanted in 45 patients (median age, 12.4 years; range, 5 weeks to 37.6 years) using an automated system [Home Monitoring (HM)] over the last 5 years. Regular transmissions were received on 72 of all days. The event messages and monitoring data of 34 (71) devices induced system revisions, electrophysiological studies including ablation, programming changes, medication changes, and alterations in sporting activity. Approximately 47 of event messages and 50 of emergency messages arrived within 4 weeks of implantation or latest outpatient visit; 113 (17) emergency messages reported acute changes in lead parameters or tachycardia, which required medical intervention. In four ICD patients, delivered shocks were unreported and only detected by HM. An automated telemonitoring system reporting patient- and system-related parameters, generated within the patients home surrounding, can improve the safety and quality of PM and ICD therapy, especially in children, by allowing early detection of system failure and changes in arrhythmic events.
引用
收藏
页码:1030 / 1037
页数:8
相关论文
共 50 条
  • [31] Perioperative Management of Patients with Cardiac Implantable Electronic Devices and Utility of Magnet Application
    oezkartal, Tardu
    Demarchi, Andrea
    Caputo, Maria Luce
    Baldi, Enrico
    Conte, Giulio
    Auricchio, Angelo
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (03)
  • [32] Outcome of cardiac implantable electronic devices in pediatric heart transplant recipients
    Baskar, Shankar
    O'Leary, Edward T.
    Whitehill, Robert
    Jackson, Lanier
    Chin, Clifford
    Mah, Douglas Y.
    Pham, Tam Dan N.
    HEART RHYTHM, 2025, 22 (02) : 416 - 417
  • [33] Implantable devices in young patients: Hitting the reset button on risk versus benefit
    Hamilton, Robert M.
    HEART RHYTHM, 2016, 13 (02) : 455 - 456
  • [34] Long-term effects of physical training on cardiopulmonary exercise parameters in young patients with congenital heart diseases
    Ferrero, Paolo
    Piazza, Isabelle
    Poggioli, Giacomo
    Chessa, Massimo
    Lorenzelli, Fabio
    CARDIOLOGY IN THE YOUNG, 2024, 34 (04) : 831 - 837
  • [35] Cardiac Implantable Electronic Devices Following Heart Transplantation
    DeFilippis, Ersilia M.
    Rubin, Geoffrey
    Farr, Maryjane A.
    Biviano, Angelo
    Wan, Elaine Y.
    Takeda, Koji
    Garan, Hasan
    Topkara, Veli K.
    Yarmohammadi, Hirad
    JACC-CLINICAL ELECTROPHYSIOLOGY, 2020, 6 (08) : 1028 - 1042
  • [36] Extraction of Select Secure leads compared to conventional pacing leads in patients with congenital heart disease and congenital atrioventricular block
    Shepherd, Emma
    Stuart, Graham
    Martin, Rob
    Walsh, Mark A.
    HEART RHYTHM, 2015, 12 (06) : 1227 - 1232
  • [37] Magnetic resonance imaging in patients with cardiac implantable electronic devices
    Maass, A. H.
    Hemels, M. E. W.
    Allaart, C. P.
    NETHERLANDS HEART JOURNAL, 2018, 26 (12) : 584 - 590
  • [38] Electrosurgery in Patients With Implantable Electronic Cardiac Devices (Pacemakers and Defibrillators)
    Garcia Bracamonte, B.
    Rodriguez, J.
    Casado, R.
    Vanaclocha, F.
    ACTAS DERMO-SIFILIOGRAFICAS, 2013, 104 (02): : 128 - 132
  • [39] Trends in cardiac implantable electronic device utilization in adults with congenital heart disease: a US nationwide analysis
    Bhalla, Jaideep Singh
    Majmundar, Monil
    Patel, Kunal N.
    Deshmukh, Abhishek J.
    Connolly, Heidi M.
    Chirac, Anca
    Egbe, Alexander C.
    Miranda, William R.
    Madhavan, Malini
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2024, 67 (02) : 319 - 328
  • [40] Magnetic resonance imaging in patients with cardiac implantable electronic devices
    A. H. Maass
    M. E. W. Hemels
    C. P. Allaart
    Netherlands Heart Journal, 2018, 26 : 584 - 590