Telemonitoring with implantable electronic devices in young patients with congenital heart diseases

被引:13
|
作者
Zartner, Peter A. [1 ]
Toussaint-Goetz, Nicole [1 ]
Photiadis, Joachim [2 ]
Wiebe, Walter [1 ]
Schneider, Martin B. [1 ]
机构
[1] Deutsch Kinderherzzentrum Sankt Augustin, Dept Cardiol, D-53757 St Augustin, Germany
[2] Deutsch Kinderherzzentrum Sankt Augustin, Dept Cardiothorac Surg, D-53757 St Augustin, Germany
来源
EUROPACE | 2012年 / 14卷 / 07期
关键词
Children; Infants; Congenital heart disease; Pacemaker; Defibrillator; Automated telemonitoring; Network; CARDIAC RESYNCHRONIZATION THERAPY; FOLLOW-UP; CARDIOVERTER-DEFIBRILLATORS; PACEMAKER; CHILDREN; REMOTE; ADULTS; EXPERIENCE; ICD; ADOLESCENTS;
D O I
10.1093/europace/eur434
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In paediatric patients with a pacemaker (PM) or an implantable cardioverter defibrillator (ICD) device, interrogation during in-clinic visits is usually required to obtain information on arrhythmias or system failures. An automated telemonitoring system tracking patient- and system-related parameters provides this information on a daily basis and might assist patient management. This retrospective analysis evaluates telemetric data obtained from 48 devices implanted in 45 patients (median age, 12.4 years; range, 5 weeks to 37.6 years) using an automated system [Home Monitoring (HM)] over the last 5 years. Regular transmissions were received on 72 of all days. The event messages and monitoring data of 34 (71) devices induced system revisions, electrophysiological studies including ablation, programming changes, medication changes, and alterations in sporting activity. Approximately 47 of event messages and 50 of emergency messages arrived within 4 weeks of implantation or latest outpatient visit; 113 (17) emergency messages reported acute changes in lead parameters or tachycardia, which required medical intervention. In four ICD patients, delivered shocks were unreported and only detected by HM. An automated telemonitoring system reporting patient- and system-related parameters, generated within the patients home surrounding, can improve the safety and quality of PM and ICD therapy, especially in children, by allowing early detection of system failure and changes in arrhythmic events.
引用
收藏
页码:1030 / 1037
页数:8
相关论文
共 50 条
  • [1] Performance of an autonomous telemonitoring system in children and young adults with congenital heart diseases
    Zartner, Peter
    Handke, Ronald
    Photiadis, Joachim
    Brecher, Anne M.
    Schneider, Martin B.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 (10): : 1291 - 1299
  • [2] The Effects of Cardiac Implantable Electronic Device Personal Protective Equipment Use in Patients with Pediatric and Adult Congenital Heart Disease
    Kean, Kelly A. A.
    Fitzgerald, Donna J. J.
    Ayers, Mark D. D.
    Kean, Adam C. C.
    PEDIATRIC CARDIOLOGY, 2023, 44 (08) : 1721 - 1725
  • [3] Investigating the Cost-Effectiveness of Telemonitoring Patients With Cardiac Implantable Electronic Devices: Systematic Review
    Raes, Sarah
    Prezzi, Andrea
    Willems, Rik
    Heidbuchel, Hein
    Annemans, Lieven
    JOURNAL OF MEDICAL INTERNET RESEARCH, 2024, 26
  • [4] Effect of telemonitoring of cardiac implantable electronic devices on healthcare utilization: a meta-analysis of randomized controlled trials in patients with heart failure
    Klersy, Catherine
    Boriani, Giuseppe
    De Silvestri, Annalisa
    Mairesse, Georges H.
    Braunschweig, Frieder
    Scotti, Valeria
    Balduini, Anna
    Cowie, Martin R.
    Leyva, Francisco
    EUROPEAN JOURNAL OF HEART FAILURE, 2016, 18 (02) : 195 - 204
  • [5] Investigating the Cost-Effectiveness of Telemonitoring Patients With Cardiac Implantable Electronic Devices: Systematic Review
    Raes, Sarah
    Prezzi, Andrea
    Willems, Rik
    Heidbuchel, Hein
    Annemans, Lieven
    JOURNAL OF MEDICAL INTERNET RESEARCH, 2024, 26
  • [6] MRI in patients with cardiac implantable electronic devices
    Cronin, Edmond M.
    Mahon, Niall
    Wilkoff, Bruce L.
    EXPERT REVIEW OF MEDICAL DEVICES, 2012, 9 (02) : 139 - 146
  • [7] Remote monitoring of cardiac implantable electronic devices (CIED)
    Zeitler, Emily P.
    Piccini, Jonathan P.
    TRENDS IN CARDIOVASCULAR MEDICINE, 2016, 26 (06) : 568 - 577
  • [8] MRI of Patients with Cardiac Implantable Electronic Devices
    Martinez, Jessica A.
    Ennis, Daniel B.
    CURRENT CARDIOVASCULAR IMAGING REPORTS, 2019, 12 (07)
  • [9] Cardiac Implantable Electronic Devices in Ebstein Anomaly: Management and Outcomes
    Tan, Nicholas Y.
    Amin, Mustapha
    Dearani, Joseph A.
    McLeod, Christopher J.
    Stephens, Elizabeth H.
    Cannon, Bryan C.
    Miranda, William R.
    Connolly, Heidi M.
    Egbe, Alexander
    Asirvatham, Samuel J.
    Madhavan, Malini
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2022, 15 (07) : 464 - 472
  • [10] Results of remote follow-up and monitoring in young patients with cardiac implantable electronic devices
    Silvetti, Massimo S.
    Saputo, Fabio A.
    Palmieri, Rosalinda
    Placidi, Silvia
    Santucci, Lorenzo
    Di Mambro, Corrado
    Righi, Daniela
    Drago, Fabrizio
    CARDIOLOGY IN THE YOUNG, 2016, 26 (01) : 53 - 60