Optimisation and evaluation of an electromagnetic tracking device for high-accuracy three-dimensional ultrasound imaging of the carotid arteries

被引:41
作者
Barratt, DC
Davies, AH
Hughes, AD
Thom, SA
Humphries, KN
机构
[1] Univ London Imperial Coll Sci Technol & Med, Sch Med, Dept Clin Pharmacol & Therapeut, London, England
[2] Univ London Imperial Coll Sci Technol & Med, Sch Med, Dept Surg, London, England
[3] Univ London Imperial Coll Sci Technol & Med, Sch Med, Dept Imaging, London, England
关键词
three-dimensional ultrasound; magnetic tracking device; carotid artery; accuracy;
D O I
10.1016/S0301-5629(01)00395-7
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Electromagnetic tracking devices provide a flexible, low cost solution for three-dimensional ultrasound (3-D US) imaging. They are, however, susceptible to interference. A commercial device (Ascension pcBIRD) was evaluated to assess the accuracy in locating the scan probe as part of a digital, freehand 3-D US imaging system aimed at vascular applications. The device was optimised by selecting a measurement rate and filter setting that minimised the mean deviation in repeated position and orientation measurements. Experimental evaluation of accuracy indicated that, overall, absolute errors were small: the RMS absolute error was 0.2 mm (range: -0.7 to 0.5 mm) for positional measurements over translations up to 90 rum, and 0.2 degrees (range: -0.8 to 0.9 degrees) for rotational measurements up to 30 degrees. In the case of position measurements, the absolute errors were influenced by the location of the scanner relative to the scan volume. We conclude that the device tested provides an accuracy sufficient for use within a freehand 3-D US system for carotid artery imaging.
引用
收藏
页码:957 / 968
页数:12
相关论文
共 31 条
  • [1] Three-dimensional freehand ultrasound: Image reconstruction and volume analysis
    Barry, CD
    Allott, CP
    John, NW
    Mellor, PM
    Arundel, PA
    Thomson, DS
    Waterton, JC
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (08) : 1209 - 1224
  • [2] Systematic distortions in magnetic position digitizers
    Birkfellner, W
    Watzinger, F
    Wanschitz, F
    Enislidis, G
    Kollmann, C
    Rafolt, D
    Nowotny, R
    [J]. MEDICAL PHYSICS, 1998, 25 (11) : 2242 - 2248
  • [3] Calibration of tracking systems in a surgical environment
    Birkfellner, W
    Watzinger, F
    Wanschitz, F
    Ewers, R
    Bergmann, H
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) : 737 - 742
  • [4] WORK IN PROGRESS - COMMON CAROTID-ARTERY CONTOURS RECONSTRUCTED IN 3 DIMENSIONS FROM PARALLEL ULTRASONIC IMAGES
    BLANKENHORN, DH
    CHIN, HP
    STRIKWERDA, S
    BAMBERGER, J
    HESTENES, JD
    [J]. RADIOLOGY, 1983, 148 (02) : 533 - 537
  • [5] Accuracy of an electromagnetic measurement device and application to the measurement and description of knee joint motion
    Bull, AMJ
    Berkshire, FH
    Amis, AA
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1998, 212 (H5) : 347 - 355
  • [6] CURTIS WD, 1993, IEEE VIRTUAL REALITY ANNUAL INTERNATIONAL SYMPOSIUM, P377, DOI 10.1109/VRAIS.1993.380755
  • [7] Influence of ECG-triggered data acquisition on reliability for carotid plaque volume measurements with a magnetic sensor three-dimensional ultrasound system
    Delcker, A
    Tegeler, C
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (04) : 601 - 605
  • [8] QUANTIFICATION OF ATHEROSCLEROTIC PLAQUES IN CAROTID ARTERIES BY 3-DIMENSIONAL ULTRASOUND
    DELCKER, A
    DIENER, HC
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (799) : 672 - 678
  • [9] Delcker A, 1994, EUR J ULTRASOUND, V1, P337
  • [10] DELCKER A, 1996, CEREBROVASC DIS, V6, P2