The Effects of High Insertion Torque Versus Low Insertion Torque on Marginal Bone Resorption and Implant Failure Rates: A Systematic Review With Meta-Analyses

被引:47
作者
Berardini, Marco [1 ]
Trisi, Paolo [2 ,3 ]
Sinjari, Bruna
Rutjes, Anne W. S. [4 ]
Caputi, Sergio [3 ]
机构
[1] Biomat Clin & Histol Res Assoc, Pescara, Italy
[2] Biomat Clin Res Assoc, Pescara, Italy
[3] Univ G DAnnunzio, Dept Med Oral & Biotechnol Sci, Chieti, Italy
[4] Univ G DAnnunzio, Ctr Systemat Reviews, Chieti, Italy
关键词
peri-implant bone remodeling; osseointegration; implant primary stability; torque review; titanium implants; LOADING PROTOCOLS; STABILITY; DENSITY; SURVIVAL; DESIGN; TRIAL; DOGS;
D O I
10.1097/ID.0000000000000422
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The aim was to analyze the data about the effects on marginal bone resorption and implant failure rates between implants inserted with high or low insertion torque values. Materials and Methods: A systematic literature search until July 2015 was conducted. Data were summarized qualitatively in descriptive tables and quantitatively by performing random effects meta-analyses of effect sizes (ESs) for bone resorption and bone-to-implant contact (BIC) and relative risks (RRs) for implant failures. Risk of bias assessments were performed using the Cochrane tool for human studies and the SYRCLE's tool for animal studies. Results: Four studies in humans and 6 quasirandomized animal studies were included. A total of 591 implants were evaluated qualitatively: 348 installed with high insertion torque (> 25 Ncm, up to 176 Ncm) and 243 implants inserted with low torque values (<30-35 Ncm). No significant differences were detected for bone resorption (ES, 0.13; 95% confidence interval [CI], -0.12 to 0.38 in human studies; ES predictive interval from 35.03 to 34.50 in animal studies), implant failure (RR, 0.39; 95% CI, 0.01-20.77 in human studies; RR, 2.05; 95% CI, 0.19-21.71 in animal studies), or BIC (ES predictive interval from -3.84 to 5.13 in animal studies). Conclusion: The current review indicated that there is no significant difference in marginal bone resorption and implant failure rate between implants inserted with high or low insertion torque values.
引用
收藏
页码:532 / 540
页数:9
相关论文
共 28 条
[1]   Undersized Implant Site Preparation to Enhance Primary Implant Stability in Poor Bone Density: A Prospective Clinical Study [J].
Alghamdi, Hamdan ;
Anand, Pradeep S. ;
Anil, Sukumaran .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2011, 69 (12) :E506-E512
[2]   Loading Protocols for Single-Implant Crowns: A Systematic Review and Meta-Analysis [J].
Benic, Goran I. ;
Mir-Mari, Javier ;
Hammerle, Christoph H. F. .
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2014, 29 :222-238
[3]   Are insertion torque and early osseointegration proportional? A histologic evaluation [J].
Campos, Felipe E. B. ;
Jimbo, Ryo ;
Bonfante, Estevam A. ;
Barbosa, Darceny Z. ;
Oliveira, Maiolino T. F. ;
Janal, Malvin N. ;
Coelho, Paulo G. .
CLINICAL ORAL IMPLANTS RESEARCH, 2015, 26 (11) :1256-1260
[4]  
Cannizzaro G, 2012, EUR J ORAL IMPLANTOL, V5, P333
[5]   Biomechanical Evaluation of Undersized Drilling on Implant Biomechanical Stability at Early Implantation Times [J].
Coelho, Paulo G. ;
Marin, Charles ;
Teixeira, Hellen S. ;
Campos, Felipe E. ;
Gomes, Julio B. ;
Guastaldi, Fernando ;
Anchieta, Rodolfo B. ;
Silveira, Lucas ;
Bonfante, Estevam A. .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2013, 71 (02) :E69-E75
[6]  
Cohen J., 1988, STAT POWER ANAL BEHA, V8, P273
[7]   Histologic and Biomechanical Evaluation of the Effects of Implant Insertion Torque on Peri-Implant Bone Healing [J].
Consolo, Ugo ;
Travaglini, Domenico ;
Todisco, Marzio ;
Trisi, Paolo ;
Galli, Silvia .
JOURNAL OF CRANIOFACIAL SURGERY, 2013, 24 (03) :860-865
[8]  
Degidi M, 2012, CLIN IMPLANT DENT R, p[14, e39]
[9]   Effect of insertion torque on titanium implant osseointegration: an animal experimental study [J].
Duyck, Joke ;
Roesems, Rutger ;
Cardoso, Marcio V. ;
Ogawa, Toru ;
Camargos, Germana De Villa ;
Vandamme, Katleen .
CLINICAL ORAL IMPLANTS RESEARCH, 2015, 26 (02) :191-196
[10]   Histological, histomorphometrical, and radiological evaluation of an experimental implant design with a high insertion torque [J].
Duyck, Joke ;
Corpas, Livia ;
Vermeiren, Stephanie ;
Ogawa, Toru ;
Quirynen, Marc ;
Vandamme, Katleen ;
Jacobs, Reinhilde ;
Naert, Ignace .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (08) :877-884