Quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPECT:: MRI validation and comparison of the Emory Cardiac Tool Box with QGS and 4D-MSPECT

被引:0
作者
Schaefer, WM
Lipke, CSA
Standke, D
Kühl, HP
Nowak, B
Kaiser, HJ
Koch, KC
Buell, U
机构
[1] Rhein Westfal TH Aachen, Univ Hosp, Dept Nucl Med, D-52074 Aachen, Germany
[2] Rhein Westfal TH Aachen, Univ Hosp, Med Clin Cardiol 1, D-52074 Aachen, Germany
关键词
gated (99m)Tc-MIBI SPECT; Emory Cardiac Tool Box; 4D-MSPECT; Quantitative Gated SPECT; cardiovascular MRI; left ventricular volumes; ejection fraction;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The goal of this study was to validate the accuracy of the Emory Cardiac Tool Box (ECTB) in assessing left ventricular end-diastolic or end-systolic volume (EDV, ESV) and ejection fraction (LVEF) from gated (99m)Tc-methoxyisobutylisonitrile ((99m)Tc-MIBI) SPECT using cardiac MRI (cMRI) as a reference. Furthermore, software-specific characteristics of ECTB were analyzed in comparison with 4D-MSPECT and Quantitative Gated SPECT (QGS) results (all relative to cMRI). Methods: Seventy patients with suspected or known coronary artery disease were examined using gated 99mTc-MIBI SPECT (8 gates/cardiac cycle) 60 min after tracer injection at rest. EDV, ESV, and LVEF were calculated from gated 99mTc-MIBI SPECT using ECTB, 4D-MSPECT, and QGS. Directly before or after gated SPECT, cMRI (20 gates/cardiac cycle) was performed as a reference. EDV, ESV, and LVEF were calculated using Simpson's rule. Results: Correlation between results of gated 99mTc-MIBI SPECT and cMRI was high for EDV (R = 0.90 [ECTB], R = 0.88 [4D-MSPECT], R = 0.92 [QGS]), ESV (R = 0.94 [ECTB], R = 0.96 [4D-MSPECT], R = 0.96 [QGS]), and LVEF (R = 0.85 [ECTB], P = 0.87 [4D-MSPECT], R = 0.89 [QGS]). EDV (ECTB) did not differ significantly from cMRI, whereas 4D-MSPECT and QGS underestimated EDV significantly compared with cMRI (mean +/- SD: 131 +/- 43 mL [ECTB], 127 +/- 42 mL [4D-MSPECT], 120 +/- 38 mL [QGS], 137 +/- 36 mL [cMRI]). For ESV, only ECTB yielded values that were significantly lower than cMRI. For LVEF, ECTB and 4D-MSPECT values did not differ significantly from cMRI, whereas QGS values were significantly lower than cMRI (mean SD: 62.7% +/- 13.7% [ECTB], 59.0% +/- 12.7% [4D-M-SPECT], 53.2% +/- 11.5% [QGS], 60.6% +/- 13.9% [cMRI]). Conclusion: EDV, ESV, and LVEF as determined by ECTB, 4D-MSPECT, and QGS from gated 99mTc-MIBI SPECT agree over a wide range of clinically relevant values with cMR1. Nevertheless, any algorithm-inherent over- or underestimation of volumes and LVEF should be accounted for and an interchangeable use of different software packages should be avoided.
引用
收藏
页码:1256 / 1263
页数:8
相关论文
共 27 条
  • [1] MR evaluation of ventricular function: True fast imaging with steady-state precession versus fast low-angle shot cine MR imaging: Feasibility study
    Barkhausen, J
    Ruehm, SG
    Goyen, M
    Buck, T
    Laub, G
    Debatin, JF
    [J]. RADIOLOGY, 2001, 219 (01) : 264 - 269
  • [2] Left ventricular function: Correlation of quantitative gated SPECT and MR imaging over a wide range of values
    Bavelaar-Croon, CDL
    Kayser, HWM
    van der Wall, EE
    de Roos, A
    Dibbets-Schneider, P
    Pauwels, EKJ
    Germano, G
    Atsma, DE
    [J]. RADIOLOGY, 2000, 217 (02) : 572 - 575
  • [3] Comparison of gated single-photon emission computed tomography with magnetic resonance imaging for evaluation of left ventricular function in ischemic cardiomyopathy
    Bax, JJ
    Lamb, H
    Dibbets, P
    Pelikan, H
    Boersma, E
    Viergever, EP
    Germano, G
    Vliegen, HW
    de Roos, A
    Pauwels, EKJ
    Van der Wall, EE
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2000, 86 (12) : 1299 - 1305
  • [4] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [5] DEPUEY E, 1995, J NUCL MED, V36, P952
  • [6] QUANTIFICATION OF THE LEFT-VENTRICULAR VOLUMES AND FUNCTION WITH CINE MR-IMAGING - COMPARISON OF GEOMETRIC-MODELS WITH 3-DIMENSIONAL DATA
    DULCE, MC
    MOSTBECK, GH
    FRIESE, KK
    CAPUTO, GR
    HIGGINS, CB
    [J]. RADIOLOGY, 1993, 188 (02) : 371 - 376
  • [7] Faber TL, 1999, J NUCL MED, V40, P650
  • [8] Evaluation of left ventricular endocardial volumes and ejection fractions computed from gated perfusion SPECT with magnetic resonance imaging: Comparison of two methods
    Faber, TL
    Vansant, JP
    Pettigrew, RI
    Galt, JR
    Blais, M
    Chatzimavroudis, G
    Cooke, CD
    Folks, RD
    Waldrop, SM
    Gurtler-Krawczynska, E
    Wittry, MD
    Garcia, EV
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2001, 8 (06) : 645 - 651
  • [9] FICARO E, 2003, J NUCL CARDIOL, V10, pS23
  • [10] Ficaro EP, 1999, CIRCULATION, V100, P26