A dose comparison study between XVI® and OBI® CBCT systems

被引:118
作者
Song, William Y.
Kamath, Srijit
Ozawal, Shuichi
Al Ani, Shlomi
Chvetsov, Alexei
Lhandare, Niranjan
Palta, Jatinder R.
Liu, Chihray
Li, Jonathan G.
机构
[1] Department of Radiation Oncology, University of Florida, Gainesville
[2] Department of Radiation Oncology, Davis Cancer Center, Gainesville, FL 32610-0385
关键词
cone-beam CT; dose measurements; IGRT; XVI; OBI;
D O I
10.1118/1.2825619
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to establish a comprehensive set of dose measurements data obtained from the X-ray Volumetric Imager (XVI (R), Elekta Oncology Systems) and the On-Board Imager (01310, Varian Medical Systems) cone-beam CT (CBCT) systems. To this end, two uniform-density cylindrical acrylic phantoms with diameters of 18 cm (head phantom) and 30 cm (body phantom) were used for all measurements. Both phantoms included ion chamber placement holes in the center and at periphery (2 cm below surface). For the XVI unit, the four standard manufacturer-supplied protocols were measured. For the OBI unit, the full bow tie and half bow tie (and no bow tie) filters were used in combination with the two scanning modes; namely, full-fan and half-fan. The total milliampere X seconds (mA s) setting was also varied for each protocol to establish the linear relationship between the dose deposited and the mA s used (with all other factors being held constant). Half-value layers in aluminum (Al) were also measured for beam characteristic determination. For the XVI unit, the average dose ranged from 0.1 to 3.5 cGy with the highest dose measured using the "prostate" protocol with the body phantom. For the OBI unit, the average dose ranged from 1.1 to 8.3 cGy with the highest dose measured using the full-fan protocol with the head phantom. The measured doses were highly linear as a function of mA s, for both units, where the measurement points followed a linear relationship very closely with R-2 > 0.99 for all cases. Half-value layers were between 4.6- and 7.0-mm-Al for the two CBCT units where XVI generally had more penetrating beams at the similar kVp settings. In conclusion, a comprehensive series of dose measurements were performed on the XVI and the OBI CBCT units. In the process, many of the important similarities and differences between the two systems were observed and summarized in this work. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:480 / 486
页数:7
相关论文
共 33 条
  • [1] A multi-platform approach to image guided radiation therapy (IGRT)
    Amies, C
    Bani-Hashemi, A
    Celi, JC
    Grousset, G
    Ghelmansarai, F
    Hristov, D
    Lane, D
    Mitschke, M
    Singh, A
    Shukla, H
    Stein, J
    Wofford, M
    [J]. MEDICAL DOSIMETRY, 2006, 31 (01) : 12 - 19
  • [2] Is it time to retire the CTDI for CT quality assurance and dose optimization?
    Brenner, DJ
    [J]. MEDICAL PHYSICS, 2005, 32 (10) : 3225 - 3226
  • [3] CONE-BEAM CT FOR RADIOTHERAPY APPLICATIONS
    CHO, PS
    JOHNSON, RH
    GRIFFIN, TW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11) : 1863 - 1883
  • [4] Increased risk of biochemical and local failure in patients with distended rectum on the planning CT for prostate cancer radiotherapy
    De Crevoisier, R
    Tucker, SL
    Dong, L
    Mohan, R
    Cheung, R
    Cox, JD
    Kuban, DA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04): : 965 - 973
  • [5] Characteristics of kilovoltage x-ray beams used for cone-beam computed tomography in radiation therapy
    Ding, George X.
    Duggan, Dennis M.
    Coffey, Charles W.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (06) : 1595 - 1615
  • [6] Dixon RL, 2003, MED PHYS, V30, P1272, DOI 10.1118/1.1576952
  • [7] Dose and image quality for a cone-beam C-arm CT system
    Fahrig, Rebecca
    Dixon, Robert
    Payne, Thomas
    Morin, Richard L.
    Ganguly, Arundhuti
    Strobel, Norbert
    [J]. MEDICAL PHYSICS, 2006, 33 (12) : 4541 - 4550
  • [8] Online image-guided intensity-modulated radiotherapy for prostate cancer: How much improvement can we expect? A theoretical assessment of clinical benefits and potential dose escalation by improving precision and accuracy of radiation delivery
    Ghilezan, M
    Yan, D
    Liang, J
    Jaffray, D
    Wong, J
    Martinez, A
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (05): : 1602 - 1610
  • [9] Compensators for dose and scatter management in cone-beam computed tomography
    Graham, S. A.
    Moseley, D. J.
    Siewerdsen, J. H.
    Jaffray, D. A.
    [J]. MEDICAL PHYSICS, 2007, 34 (07) : 2691 - 2703
  • [10] A study of the effects of internal organ motion on dose escalation in conformal prostate treatments
    Happersett, L
    Mageras, GS
    Zelefsky, MJ
    Burman, CM
    Leibel, SA
    Chui, C
    Fuks, Z
    Bull, S
    Ling, CC
    Kutcher, GJ
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 66 (03) : 263 - 270