Inhibitive potentiometric detection of trace metals with ultrathin polypyrrole glucose oxidase biosensor

被引:31
作者
Ayenimo, Joseph G. [1 ]
Adeloju, Samuel B. [1 ]
机构
[1] Monash Univ, Sch Chem, NanoSci & Sensor Technol Res Grp, Clayton, Vic 3800, Australia
关键词
Ultrathin polypyrrole film; Glucose oxidase; Biosensor; Trace metals; Potentiometry; Inhibition; CONDUCTOMETRIC BIOSENSOR; ENZYME; IONS; UREASE; MERCURY; SENSITIVITY; CALCIUM; ASSAYS;
D O I
10.1016/j.talanta.2015.01.006
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method, based on the inhibition of an ultrathin polypyrrole-glucose oxidase (PPy-GOx) potentiometric biosensor response, is described for the detection of Cu2+, Hg2+, Cd2+ and Pb2+ ions. Based on experimental conditions (0.2 M pyrrole, 500 U mL(-1) GOx, and an applied current density of 0.05 mA cm(-2) and a polymerization period of 500 s) previously published by us, PPy-GOx films of approximately 55 nm thick were used to demonstrate the inhibitive potentiometric detection of selected trace metals down to 0.079 mu M Cu2+, 0.025 mu M Hg2+, 0.024 mu M Pb2+ and 0.044 mu M Cd2+. Furthermore, good linear concentration ranges were achieved for Cu2+ (0.079-16 mu M), Hg2+ (0.025-5 mu M), Pb2+ (0.10-15 mu M) and Cd2+ (0.04-62 mu M). The analysis of the nature of the inhibition of glucose oxidase in the PPy-GOx biosensor by these metals was achieved by Dixon and Cornish-Bowden plots. The shapes of the curves (exponential decay, parabolic and linear) obtained for the inhibitors suggest that the inhibition by the metal ions may not be exclusively directed at the essential -SH group, but involve additional binding sites of the enzyme. Dixon and Cornish-Bowden plots suggest that the inhibition is competitive for Cd2+, while non-competitive inhibition was observed for other metal ions. The ultrathin PPy-GOx film enabled improved permeability to the metal inhibitors than possible with conventional biosensors with thicker films and, hence, better reflects the actual inhibition effect of the trace metals on the enzyme activity. The use of the ultra-thin film also eliminated the usual need for incubation of the enzyme electrode for a long period in the presence of the inhibitors. Furthermore, a rapid recovery of the enzyme activity was achieved by simply washing the electrode with water and storing in phosphate buffer for 10-15 min. The proposed biosensing approach was successfully used for the detection of individual trace metals in tap water, achieving a 98-101% recovery. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:62 / 70
页数:9
相关论文
共 47 条
[1]   Heavy metal-induced inhibition of Aspergillus niger nitrate reductase:: applications for rapid contaminant detection in aqueous samples [J].
Aiken, AM ;
Peyton, BM ;
Apel, WA ;
Petersen, JN .
ANALYTICA CHIMICA ACTA, 2003, 480 (01) :131-142
[2]   Determination of calcium in milk and water samples by using catalase enzyme electrode [J].
Akyilmaz, Erol ;
Kozgus, Ozge .
FOOD CHEMISTRY, 2009, 115 (01) :347-351
[3]  
Alexander PW, 2000, ELECTROANAL, V12, P343
[4]   Determination of mercury(II), methylmercury and ethylmercury in the ng/ml range with an electrochemical enzyme glucose probe [J].
Amine, A ;
Cremisini, C ;
Palleschi, G .
MIKROCHIMICA ACTA, 1995, 121 (1-4) :183-190
[5]   EFFECT OF THE COUPLING REAGENT ON THE METAL INHIBITION OF IMMOBILIZED UREASE IN AN OPTICAL BIOSENSOR [J].
ANDRES, RT ;
NARAYANASWAMY, R .
ANALYST, 1995, 120 (05) :1549-1554
[6]   Improved potentiometric glucose detection with ultra-thin polypyrrole-glucose oxidase films [J].
Ayenimo, Joseph G. ;
Adeloju, Samuel B. .
ANALYTICAL METHODS, 2014, 6 (22) :8996-9006
[7]   Detection of heavy metal ions at femtomolar levels using protein-based biosensors [J].
Bontidean, I ;
Berggren, C ;
Johansson, G ;
Csöregi, E ;
Mattiasson, B ;
Lloyd, JA ;
Jakeman, KJ ;
Brown, NL .
ANALYTICAL CHEMISTRY, 1998, 70 (19) :4162-4169
[8]   Experimental Platform to Study Heavy Metal Ion-Enzyme Interactions and Amperometric Inhibitive Assay of Ag+ Based on Solution State and Immobilized Glucose Oxidase [J].
Chen, Chao ;
Xie, Qingji ;
Wang, Lihua ;
Qin, Cong ;
Xie, Fangyun ;
Yao, Shouzhuo ;
Chen, Jinhua .
ANALYTICAL CHEMISTRY, 2011, 83 (07) :2660-2666
[9]  
Cornish-Bowden A., 1976, Principles of Enzyme Kinetics, P52
[10]  
Del Carlo M., 2005, BIOSENSORS FOOD QUAL