BLADE Sequences in Transverse T2-weighted MR Imaging of the Cervical Spine. Cut-off for Artefacts?

被引:6
作者
Finkenzeller, T. [1 ]
Wendl, C. M. [2 ]
Lenhart, S. [3 ]
Stroszczynski, C. [2 ]
Schuierer, G. [4 ]
Fellner, C. [2 ]
机构
[1] Klinikum Nuernberg Sued, Inst Radiol & Neuroradiol, D-90471 Nurnberg, Germany
[2] Univ Hosp Regensburg, Inst Radiol, Regensburg, Germany
[3] Klinikum Weiden, Weiden, Germany
[4] Ctr Neuroradiol, Inst Neuroradiol, Regensburg, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2015年 / 187卷 / 02期
关键词
spine; spinal cord; MR-imaging; spinal canal; technical aspects; MAGNETIC-RESONANCE; QUALITY; MOTION; BRAIN; IMPROVEMENT; PROPELLER; FLOW;
D O I
10.1055/s-0034-1385179
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The BLADE (PROPELLER) technique reduces artefacts in imaging of the cervical spine in sagittal orientation, but till now failed to do so in axial orientation, because here it increased through plane CSF-flow artefacts, which spoiled the benefit of BLADE artefact reduction "in plane". The aim of this study was to compare a BLADE sequence with optimised measurement parameters in axial orientation to T2-TSE. Materials and Methods: Both sequences were compared in 58 patients with 31 discal, 16 bony and 11 spinal cord lesions. Image sharpness, reliability of spinal cord depiction, CSF flow artefacts and lesion detection were evaluated by 3 independent observers. Additionally the observers were asked which sequence they would prefer for diagnostic workup. Statistical evaluations were performed using sign and chi(2) test. Results: BLADE was significantly superior concerning image sharpness, spinal cord depiction and overall lesion detection. BLADE was rated better for most pathologies, for bony lesions the differences compared with TSE were statistically significant. Regarding CSF-flow artefacts both sequences showed no difference. All readers preferred BLADE in side by side reading. Conclusion: An optimised axial T2 BLADE sequence decreases the problems of increased through plane CSF-flow artefacts in this orientation. By reducing various other artefacts it yields better image quality and has the potential to reduce the number of non-diagnostic examinations especially in uncooperative patients.
引用
收藏
页码:102 / 108
页数:7
相关论文
共 18 条
[1]   Cerebrospinal fluid dynamics and relation with blood flow -: A magnetic resonance study with semiautomated cerebrospinal fluid segmentation [J].
Balédent, O ;
Henry-Feugeas, MCC ;
Idy-Peretti, I .
INVESTIGATIVE RADIOLOGY, 2001, 36 (07) :368-377
[2]   BLADE in Sagittal T2-Weighted MR Imaging of the Cervical Spine [J].
Fellner, C. ;
Menzel, C. ;
Fellner, F. A. ;
Ginthoer, C. ;
Zorger, N. ;
Schreyer, A. ;
Jung, E. M. ;
Feuerbach, S. ;
Finkenzeller, T. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2010, 31 (04) :674-681
[3]   Novel application of T1-weighted BLADE sequences with fat suppression compared to TSE in contrast-enhanced T1-weighted imaging of the neck: Cutting-edge images? [J].
Finkenzeller, Thomas ;
Zorger, Niels ;
Kuehnel, Thomas ;
Paetzel, Christian ;
Schuierer, Gerhard ;
Stroszczynski, Christian ;
Fellner, Claudia .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 37 (03) :660-668
[4]  
Forbes KP, 2003, AM J NEURORADIOL, V24, P796
[5]   Dispensing with Sedation in Pediatric MR Imaging of the Brain: What is Feasible? [J].
Heyer, C. M. ;
Lemburg, S. P. ;
Sterl, S. ;
Holland-Letz, T. ;
Nicolas, V. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2012, 184 (11) :1034-1042
[6]   MRI Artifact Reduction and Quality Improvement in the Upper Abdomen with PROPELLER and Prospective Acquisition Correction (PACE) Technique [J].
Hirokawa, Yuusuke ;
Isoda, Hiroyoshi ;
Maetani, Yoji S. ;
Arizono, Shigeki ;
Shimada, Kotaro ;
Togashi, Kaori .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 191 (04) :1154-1158
[7]   Elimination of motion, pulsatile flow and cross-talk artifacts using blade sequences in lumbar spine MR imaging [J].
Lavdas, Eleftherios ;
Mavroidis, Panayiotis ;
Kostopoulos, Spiros ;
Glotsos, Dimitrios ;
Roka, Violeta ;
Koutsiaris, Aristotle G. ;
Batsikas, Georgios ;
Sakkas, Georgios K. ;
Tsagkalis, Antonios ;
Notaras, Ioannis ;
Stathakis, Sotirios ;
Papanikolaou, Nikos ;
Vassiou, Katerina .
MAGNETIC RESONANCE IMAGING, 2013, 31 (06) :882-890
[8]   Improvement of image quality using BLADE sequences in brain MR imaging [J].
Lavdas, Eleftherios ;
Mavroidis, Panayiotis ;
Kostopoulos, Spiros ;
Glotsos, Dimitrios ;
Roka, Violeta ;
Topalzikis, Theofilos ;
Bakas, Athanasios ;
Oikonomou, Georgia ;
Papanikolaou, Nikos ;
Batsikas, Georgios ;
Kaffes, Ioannis ;
Kechagias, Dimitrios .
MAGNETIC RESONANCE IMAGING, 2013, 31 (02) :189-200
[9]   Normal MRI appearance and motion-related phenomena of CSF [J].
Lisanti, Christopher ;
Carlin, Carrie ;
Banks, Kevin P. ;
Wang, David .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (03) :716-725
[10]   Magnetic resonance imaging of the cervical spine: comparison of 2D T2-weighted turbo spin echo, 2D T2*weighted gradient-recalled echo and 3D T2-weighted variable flip-angle turbo spin echo sequences [J].
Meindl, T. ;
Wirth, S. ;
Weckbach, S. ;
Dietrich, O. ;
Reiser, M. ;
Schoenberg, S. O. .
EUROPEAN RADIOLOGY, 2009, 19 (03) :713-721