Determination of cadmium in lichens by solid sampling graphite furnace atomic absorption spectrometry (SS-GF-AAS)

被引:9
作者
Coufalik, Pavel [1 ]
Uher, Antonin [2 ]
Zverina, Ondrej [3 ]
Komarek, Josef [2 ]
机构
[1] Czech Acad Sci, Inst Analyt Chem, Veveri 97, Brno 60200, Czech Republic
[2] Masaryk Univ, Fac Sci, Dept Chem, Kotlarska 2, CS-61137 Brno, Czech Republic
[3] Masaryk Univ, Fac Med, Dept Publ Hlth, Kamenice 5, Brno 62500, Czech Republic
关键词
Cadmium; Lichen; Solid sampling; AAS; BIOLOGICAL SAMPLES; CHROMIUM; METALS; LEAD;
D O I
10.1007/s10661-020-8186-5
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The determination of trace metal contents directly from a solid sample is a trend in modern atomic spectrometry. The aim of this study was to develop an analytical method for the routine determination of Cd in lichens using solid sampling graphite furnace atomic absorption spectrometry (SS-GF-AAS). For the determination of Cd, the temperature program of the graphite furnace was optimized using a mixed matrix modifier (Pd + Mg (NO3)(2) + Triton X-100). The limit of detection and the limit of quantification were 0.9 mu g/kg and 3 mu g/kg, respectively. The analytical method for Cd determination in the plant matrix was verified by the analysis of certified reference materials of lichens, seaweed, and rye grass. The developed procedure was applied to the study of Cd distribution in thalli of Usnea antarctica lichen from James Ross Island, Antarctica. The SS-GF-AAS analytical method is particularly suited for use in environmental studies and plant physiology (the microanalysis of anatomical structures).
引用
收藏
页数:7
相关论文
共 17 条
[1]   Determination of cadmium and lead in biological samples by Zeeman ETAAS using various chemical modifiers [J].
Acar, O .
TALANTA, 2001, 55 (03) :613-622
[2]   Direct solid sampling with electrothermal vaporization/atomization: what for and how? [J].
Belarra, MA ;
Resano, M ;
Vanhaecke, F ;
Moens, L .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2002, 21 (12) :828-839
[3]   SOLID SAMPLING IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY USING COMMERCIAL ATOMIZERS - A REVIEW [J].
BENDICHO, C ;
DELOOSVOLLEBREGT, MTC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (05) :353-374
[4]   DIRECT SOLID VS SLURRY ANALYSIS OF TOBACCO-LEAVES FOR SOME TRACE-METALS BY GRAPHITE-FURNACE AAS - A COMPARATIVE-STUDY [J].
DOBROWOLSKI, R ;
MIERZWA, J .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 344 (7-8) :340-344
[5]   SIGNAL SUMMATION FOR THE IMPROVEMENT OF THE DETECTION PERFORMANCE AND INDIRECT INCREASE OF THE SAMPLE QUANTITY IN DIRECT AAS OF SOLIDS [J].
HENNIG, W ;
BERNDT, H ;
SCHALDACH, G .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 337 (03) :275-279
[6]  
Kurfurst U., 1998, SOLID SAMPLE ANAL DI
[7]  
Pozzatti M, 2017, ANAL METHODS-UK, V9, P329, DOI [10.1039/c6ay02594c, 10.1039/C6AY02594C]
[8]  
Raats M. M., 1992, Food Quality and Preference, V3, P89, DOI 10.1016/0950-3293(91)90028-D
[9]   Electrothermal vaporization for sample introduction in atomic absorption, atomic emission and plasma mass spectrometry-a critical review with focus on solid sampling and slurry analysis [J].
Resano, M. ;
Vanhaecke, F. ;
de Loos-Vollebregt, M. T. C. .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2008, 23 (11) :1450-1475
[10]   High-resolution continuum source graphite furnace atomic absorption spectrometry for direct analysis of solid samples and complex materials: a tutorial review [J].
Resano, Martin ;
Aramendia, Maite ;
Belarra, Miguel A. .