In Vivo Measurement of Blood Velocity in Human Major Retinal Vessels Using the Laser Speckle Method

被引:77
作者
Nagahara, Miyuki [1 ]
Tamaki, Yasuhiro [1 ]
Tomidokoro, Atsuo [1 ]
Araie, Makoto [1 ]
机构
[1] Univ Tokyo, Grad Sch Med, Dept Ophthalmol, Bunkyo Ku, Tokyo 1138655, Japan
关键词
OPTIC-NERVE HEAD; 2-DIMENSIONAL MEASUREMENT; FLUORESCEIN ANGIOGRAMS; CHOROIDAL CIRCULATION; ABSOLUTE MEASUREMENT; FLOW VISUALIZATION; REPRODUCIBILITY; PHENYLEPHRINE; PHOTOGRAPHY; ANTAGONIST;
D O I
10.1167/iovs.09-4422
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To develop a simple, noninvasive method of determining blood velocity and flow through human retinal vessels (RVs), by using the laser speckle method and validating the results by bidirectional laser Doppler velocimetry (LDV). METHOD. The square blur rate (SBR), a quantitative blurring index of the laser speckle pattern that parallels the velocity of moving substances, obtained from blood flowing through glass capillary tubes (RV analogues), correlated with tube diameter, background reflectance and absorption, flow velocity, and the SBR obtained from blood flowing through underlying glass capillary tubes (choroidal vessel analogues). A nomogram was constructed to calculate the blood velocity in human RVs from the SBR values obtained in vivo. Blood velocities in RVs were determined in 12 normal eyes by using the laser speckle method and bidirectional LDV. Measurements were performed twice at the same site at 1-hour intervals. RESULTS. Measurements from a temporal superior artery (n = 12; mean +/- SD) were blood velocity (V-mean), 41.7 +/- 4.2 mm/s; flow, 13.0 +/- 3.2 mu L/min; and diameter, 119.5 +/- 15.7 mu m and time to complete one measurement, 65 +/- 18 seconds, with the laser speckle apparatus; and V-mean, 37.7 +/- 6.7 mm/s; flow, 11.7 +/- 3.0 mu L/min; diameter, 111.1 +/- 16.6 mu m; and measurement time, 112 +/- 25 seconds, with the bidirectional LDV apparatus. The results obtained by the two methods correlated with each other (V-mean, r = 0.59, P = 0.023; flow, r = 0.83, P = 0.005; and diameter, r = 0.56, P = 0.032). The coefficients of reproducibility for V mean, blood flow, and diameter measurement were 9.5% +/- 2.5%, 10.5% +/- 3.2%, and 5.3% +/- 2.7% for the former and 15.3% +/- 4.2%, 18.5% +/- 4.1%, and 6.2% +/- 2.2% for the latter, respectively. CONCLUSIONS. The laser speckle method accurately and reproducibly determines blood velocity in human RVs in less time than the LDV apparatus requires. (Invest Ophthalmol Vis Sci. 2011; 52: 87-92) DOI:10.1167/iovs.09-4422
引用
收藏
页码:87 / 92
页数:6
相关论文
共 44 条
[1]  
American National Standards Institute, 2000, AM NAT STAND CENTR P
[2]  
AREND O, 1994, ACTA OPHTHALMOL, V72, P332
[3]  
BRIERS JD, 1982, INVEST OPHTH VIS SCI, V22, P255
[4]   LASER DOPPLER TECHNIQUE FOR ABSOLUTE MEASUREMENT OF BLOOD SPEED IN RETINAL-VESSELS [J].
FEKE, GT ;
GOGER, DG ;
TAGAWA, H ;
DELORI, FC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1987, 34 (09) :673-680
[5]   LASER DOPPLER MEASUREMENTS OF BLOOD VELOCITY IN HUMAN RETINAL-VESSELS [J].
FEKE, GT ;
RIVA, CE .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1978, 68 (04) :526-531
[6]  
FEKE GT, 1989, INVEST OPHTH VIS SCI, V30, P58
[7]   FLOW VISUALIZATION BY MEANS OF SINGLE-EXPOSURE SPECKLE PHOTOGRAPHY [J].
FERCHER, AF ;
BRIERS, JD .
OPTICS COMMUNICATIONS, 1981, 37 (05) :326-330
[8]   PULSATILE FLOW IN THE CHOROIDAL CIRCULATION - A PRELIMINARY INVESTIGATION [J].
FLOWER, RW ;
KLEIN, GJ .
EYE, 1990, 4 :310-318
[9]  
GEERAETS WJ, 1962, INVEST OPHTH VISUAL, V1, P340
[10]  
Harris A., 1994, SURV OPHTHALMOL S, V38, P81, DOI org/10.1016/0039-6257(94)90050-7