Cone beam computed tomography and image registration based on target area for stereotactic body radiation therapy of lung cancer

被引:1
|
作者
Gui, Long-Gang [1 ]
Shi, Miao [1 ]
Li, Jun [1 ]
机构
[1] Northern Jiangsu Peoples Hosp, Radiotherapy Ctr, Yangzhou 225001, Jiangsu, Peoples R China
关键词
Lung cancer; SBRT; CBCT; image registration; dosimetric parameter; INTENSITY-MODULATED RADIOTHERAPY; CLINICAL-OUTCOMES; DIAGNOSTIC-VALUE; CT; IRRADIATION;
D O I
10.1080/16878507.2021.1981754
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
This paper aims to compare the dosimetric parameters of the original plan and the simulated plan to explore the clinical value of cone beam CT and planned CT image registration based on the target area. The results showed that Dmax, Dmin, dmean and V60 of PTV in the original plan were (108.3 +/- 1.1)%, (97.2 +/- 0.3)%, (102.6 +/- 0.6)% and (95.2 +/- 0.6)% respectively. The Dmax of spinal cord was (9.6 +/- 1.8) gy, the V20 of lung was (5.2 +/- 1.9)%, and the dmean of lung was (- 3.7 +/- 1.5) gy; however, when the target area was translated in the three directions of the patient's back, right side and foot, the dosimetric parameters of the target area (Dmax, Dmin and V60), dmean and V20 of lung and Dmax of spinal cord showed slight differences, which all met the limit requirements of target area and dangerous organs (OARs) (RTOG) 0915 report. Therefore, target based image registration is feasible. If the registration translation error is less than or equal to 1 cm, there is no need to re move the treatment bed according to the registration error.
引用
收藏
页码:404 / 411
页数:8
相关论文
共 50 条
  • [1] Using previously registered cone beam computerized tomography images to facilitate online computerized tomography to cone beam computerized tomography image registration in lung stereotactic body radiation therapy
    Liang, Jian
    Liu, Qiang
    Grills, Inga
    Guerrero, Thomas
    Stevens, Craig
    Yan, Di
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (04):
  • [2] Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy
    Oechsner, Markus
    Chizzali, Barbara
    Devecka, Michal
    Combs, Stephanie Elisabeth
    Wilkens, Jan Jakob
    Duma, Marciana Nona
    RADIATION ONCOLOGY, 2016, 11
  • [3] Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy
    Markus Oechsner
    Barbara Chizzali
    Michal Devecka
    Stephanie Elisabeth Combs
    Jan Jakob Wilkens
    Marciana Nona Duma
    Radiation Oncology, 11
  • [4] Liver motion during cone beam computed tomography guided stereotactic body radiation therapy
    Park, Justin C.
    Park, Sung Ho
    Kim, Jong Hoon
    Yoon, Sang Min
    Song, Si Yeol
    Liu, Zhaowei
    Song, Bongyong
    Kauweloa, Kevin
    Webster, Matthew J.
    Sandhu, Ajay
    Mell, Loren K.
    Jiang, Steve B.
    Mundt, Arno J.
    Song, William Y.
    MEDICAL PHYSICS, 2012, 39 (10) : 6431 - 6442
  • [5] "Dose of the day" based on cone beam computed tomography and deformable image registration for lung cancer radiotherapy
    Yuan, Zilong
    Rong, Yi
    Benedict, Stanley H.
    Daly, Megan E.
    Qiu, Jianfeng
    Yamamoto, Tokihiro
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (01): : 88 - 94
  • [6] Influence of tumor characteristics on correction differences between cone-beam computed tomography-guided patient setup strategies in stereotactic body radiation therapy for lung cancer
    Jang, Seong Soon
    Park, Suk Young
    Cho, Eun Youn
    Yang, Po Song
    Huh, Gil Ja
    Yang, Young Jun
    THORACIC CANCER, 2020, 11 (02) : 311 - 319
  • [7] CONE-BEAM COMPUTED TOMOGRAPHIC IMAGE GUIDANCE FOR LUNG CANCER RADIATION THERAPY
    Bissonnette, Jean-Pierre
    Purdie, Thomas G.
    Higgins, Jane A.
    Li, Winnie
    Bezjak, Andrea
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (03): : 927 - 934
  • [8] A novel evaluation model of image registration for cone-beam computed tomography guided lung cancer radiotherapy
    Liu, Yimei
    Chen, Meining
    Fang, Jianlan
    Xiao, Liangjie
    Liu, Songran
    Li, Qiwen
    Qiu, Bo
    Huang, Runda
    Zhang, Jun
    Peng, Yinglin
    THORACIC CANCER, 2024, 15 (17) : 1333 - 1342
  • [9] Cone-beam computed tomography in lung stereotactic ablative radiation therapy: predictive parameters of early response
    Mazzola, Rosario
    Fiorentino, Alba
    Ricchetti, Francesco
    Levra, Niccolo Giaj
    Fersino, Sergio
    Di Paola, Gioacchino
    Lo Casto, Antonio
    Ruggieri, Ruggero
    Alongi, Filippo
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1064)
  • [10] A longitudinal four-dimensional computed tomography and cone beam computed tomography dataset for image-guided radiation therapy research in lung cancer
    Hugo, Geoffrey D.
    Balik, Salim
    Keall, Paul J.
    Lu, Jun
    Williamson, Jeffrey F.
    MEDICAL PHYSICS, 2017, 44 (02) : 762 - 771