Intra- and interrater reliabilities of the myotonometer when assessing the spastic condition of children with cerebral palsy

被引:38
作者
Aarrestad, DD
Williams, MD
Fehrer, SC
Mikhailenok, E
Leonard, CT [1 ]
机构
[1] Univ Montana, Dept Phys Therapy, Missoula, MT 59812 USA
[2] Neurogen Technol Inc, Missoula, MT 59812 USA
关键词
D O I
10.1177/08830738040190110801
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The purposes of this study were to assess intra- and interrater reliabilities by novice users of the Myotonometer (R) (Neurogenic Technologies, Inc., Missoula, MT), a portable electronic device that. quantifies muscle tone (stiffness) and paresis, in assessing children with cerebral palsy. Two raters used the Myotonometer to assess the biceps brachii and medial gastrocnemius muscles of 10 children with spastic-type cerebral palsy. Muscles were measured in a relaxed state and during a voluntary isometric contraction. Intraclass correlation coefficients and repeatability coefficients were calculated for each muscle and for each condition (relaxed and contracted). Intrarater reliabilities ranged front 0.82 to 0.99 (biceps brachii muscles) and 0.88 to 0.99 (medial gastrocnemius muscles). Interrater reliabilities ranged from 0.74 to 0.99 (biceps brachii muscles) and 0.84 to 0.99 (medial gastrocnemius muscles). Repeatability coefficients indicated a 98% level of agreement between raters across all conditions. Novice users of the Myotonometer, with few exceptions, had high to very high intra- and interrater reliabilities for measurements of the biceps brachii and medial gastrocnemius muscles of children with spastic-type cerebral palsy.