Housing design and testing of a surgical robot developed for orthopaedic surgery

被引:5
|
作者
Qin, Lai-Yin [1 ,2 ]
Wen, Jing Zhou [1 ]
Chui, Chun-Sing [3 ]
Leung, Kwok-Sui [3 ]
机构
[1] ND Ind Design Ltd, Shenzhen, Peoples R China
[2] Chinese Univ Hong Kong, Fac Engn, Div Biomed Engn, Hong Kong, Hong Kong, Peoples R China
[3] Chinese Univ Hong Kong, Dept Orthopaed & Traumatol, Hong Kong, Hong Kong, Peoples R China
关键词
housing design and test; navigation technology; orthopaedic robots; robotic surgery; surgical robotics; ENDOSCOPIC SURGERY; SYSTEM;
D O I
10.1016/j.jot.2016.02.002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background/Objective: Surgical technology has advanced rapidly with the introduction of robot technology. Apart from mechanical and electronic elements, housing design is an essential component that must be thoughtfully considered, bearing in mind the general requirements for medical devices used in operating theatres. The aim of this study was to design a modern and safe housing for a surgical robotic system for orthopaedic applications in Hong Kong that would meet the general requirements for obtaining local regulatory body approval. Methods: Based on the general requirements for Class II Medical Devices, industrial product designers worked in close collaboration with a robot research team formed by engineers and orthopaedic surgeons to design a modern and safe housing for the HybriDot (R) Surgical Robotic System that performs computer-assisted surgery. Results: The design received local regulatory body approval for its application in operating theatres and was approved for orthopaedic surgery in Hong Kong after fulfilling the general requirements for safety, accuracy, movability and operability. Conclusion: This project demonstrated a good model of multidisciplinary R& D of surgical robotics led by orthopaedic surgeons, in collaboration with mechanical and electronic engineers and industrial designers. Copyright (C) 2016, The Authors. Published by Elsevier (Singapore) Pte Ltd.
引用
收藏
页码:72 / 80
页数:9
相关论文
共 50 条
  • [31] Design of a Sliding-Pin IN eedle Driver for a Continuum Surgical Robot
    Zhang, Zhaoyu
    Wu, Zhonghao
    Zeng, Lingyun
    Zhang, Weihao
    Xu, Kai
    2018 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND BIOMIMETICS (ROBIO), 2018, : 922 - 927
  • [32] Automatic Surgical Field of View Control in Robot-Assisted Nasal Surgery
    He, Yucheng
    Zhao, Baoliang
    Qi, Xiaozhi
    Li, Shibo
    Yang, Yuanyuan
    Hu, Ying
    IEEE ROBOTICS AND AUTOMATION LETTERS, 2021, 6 (01) : 247 - 254
  • [33] Design and modeling of a virtual PA-10 robot for surgical applications
    Eduardo Fernandez-Riomalo, Carlos
    Andres Guastar-Morillo, Hector
    Andres Vivas-Alban, Oscar
    REVISTA FACULTAD DE INGENIERIA, UNIVERSIDAD PEDAGOGICA Y TECNOLOGICA DE COLOMBIA, 2016, 25 (42): : 21 - 32
  • [34] Dimensional synthesis and concept design of a novel minimally invasive surgical robot
    Niu, Guojun
    Pan, Bo
    Zhang, Fuhai
    Feng, Haibo
    Gao, Wenpeng
    Fu, Yili
    ROBOTICA, 2018, 36 (05) : 715 - 737
  • [35] Pilot Study on Effectiveness of Simulation for Surgical Robot Design Using Manipulability
    Kawamura, Kazuya
    Seno, Hiroto
    Kobayashi, Yo
    Fujie, Masakatsu G.
    2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 4538 - 4541
  • [36] Design and Analysis of a Snake-like Surgical Robot with Continuum Joints
    Wang, Teng
    Wang, Ziwen
    Wu, Guangxin
    Lei, Long
    Zhao, Baoliang
    Zhang, Peng
    Shang, Peng
    2020 5TH INTERNATIONAL CONFERENCE ON ADVANCED ROBOTICS AND MECHATRONICS (ICARM 2020), 2020, : 178 - 183
  • [37] The surgical robot Micro Hand S: A cheapest device for an innovative surgery. An invited commentary on "Chinese surgical robot Micro Hand S: A consecutive case series in general surgery" Comment
    Manzia, Tommaso Maria
    Parente, Alessandro
    INTERNATIONAL JOURNAL OF SURGERY, 2020, 76 : 25 - 26
  • [38] Upper limb rehabilitation robot for physical therapy: design, control, and testing
    Akdogan, Erhan
    TURKISH JOURNAL OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCES, 2016, 24 (03) : 911 - 934
  • [39] Design and path tracking control of a continuum robot for maxillary sinus surgery
    Cao, Yongfeng
    Liu, Zefeng
    Liu, Zheng
    Wang, Shuang
    Xie, Le
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2023, 18 (04) : 753 - 761
  • [40] DESIGN OF A QUADRAMANUAL ROBOT FOR SINGLE-NOSTRIL SKULL BASE SURGERY
    Swaney, Philip J.
    Croom, Jordan M.
    Burgner, Jessica
    Gilbert, Hunter B.
    Rucker, D. Caleb
    Webster, Robert J., III
    Weaver, Kyle D.
    Russell, Paul T., III
    PROCEEDINGS OF THE ASME 5TH ANNUAL DYNAMIC SYSTEMS AND CONTROL DIVISION CONFERENCE AND JSME 11TH MOTION AND VIBRATION CONFERENCE, DSCC 2012, VOL 3, 2013, : 387 - 393