Agreement in component size between preoperative measurement, navigation and final implant in total knee replacement

被引:4
作者
Hernandez-Vaquero, Daniel [1 ]
Noriega-Fernandez, Alfonso [2 ]
Roncero-Gonzalez, Sergio [2 ]
Perez-Coto, Ivan [2 ]
Sierra-Pereira, Andres A. [2 ]
Sandoval-Garcia, Manuel A. [2 ]
机构
[1] Univ Oviedo, Dept Orthopaed, Sch Med, Asturias, Spain
[2] St Agustin Univ Hosp, Dept Orthopaed, Aviles, Spain
关键词
Component size; Digital templating; Navigation; Total knee replacement; ANTERIOR FEMORAL CORTEX; TIBIAL COMPONENT; TOTAL HIP; ARTHROPLASTY; RISK; RELIABILITY; OVERHANG; INSTRUMENTATION; METAANALYSIS; ALIGNMENT;
D O I
10.1016/j.jot.2018.10.006
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: One of the possible causes of dissatisfaction reported by many patients after total knee replacement (TKR) is the lack of agreement between component size and bone structure. To avoid this complication and facilitate the procedure, preoperative planning with digitized templates is recommended. Surgical navigation indicates the best position and the most adequate size of arthroplasty and may therefore replace preoperative radiographic measurement. The objective of the study was to check agreement between the sizes of TKR components measured before surgery with digitized templates, the size recommended by the navigation and sizes actually implanted. Methods: In 103 patients scheduled for TKR, preoperative full-limb radiography was performed to measure the mechanical and anatomical axes of the limb, femur and tibia. The most adequate size of the femoral and tibial components was planned by superimposing digitized templates. The size recommended in navigation and the size of the finally implanted components were also recorded. Results: A high level of agreement was found between the sizes of femoral and tibial components measured by X-rays and in navigation (0.750 and 0.772, respectively) (intraclass correlation and Cronbach's alpha). Agreement between the sizes recommended by X-rays and navigation and those finally implanted was 0.886 for the femur and 0.891 for the tibia. Agreement levels were not different in cases with prior deformities of limb axis. Conclusions: The high level of agreement found in component sizes between radiographic measurement with digitized templates and navigation suggests that preoperative X-ray measurement is not needed when navigation is used for placement of implants during TKR. (C) 2018 The Authors. Published by Elsevier (Singapore) Pte Ltd on behalf of Chinese Speaking Orthopaedic Society.
引用
收藏
页码:84 / 91
页数:8
相关论文
共 39 条
  • [11] Computer Assistance Increases Precision of Component Placement in Total Knee Arthroplasty with Articular Deformity
    Hernandez-Vaquero, Daniel
    Suarez-Vazquez, Abelardo
    Sandoval-Garcia, Manuel A.
    Noriega-Fernandez, Alfonso
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (05) : 1237 - 1241
  • [12] Meta-Analysis of Navigation vs Conventional Total Knee Arthroplasty
    Hetaimish, Bandar M.
    Khan, M. Moin
    Simunovic, Nicole
    Al-Harhi, Hatem H.
    Bhandari, Mohit
    Zalzal, Paul K.
    [J]. JOURNAL OF ARTHROPLASTY, 2012, 27 (06) : 1177 - 1182
  • [13] Effect of femoro-tibial component size mismatch on outcome in primary total knee replacement
    Heylen, Steven
    Foubert, Knud
    Van Haver, Annemieke
    Nicolai, Paul
    [J]. KNEE, 2016, 23 (03) : 532 - 534
  • [14] Effect of Training Level on Accuracy of Digital Templating in Primary Total Hip and Knee Arthroplasty
    Hsu, Andrew R.
    Kim, Jeffrey D.
    Bhatia, Sanjeev
    Levine, Brett R.
    [J]. ORTHOPEDICS, 2012, 35 (02) : E179 - E183
  • [15] Patient-specific instrumentation does not improve radiographic alignment or clinical outcomes after total knee arthroplasty A meta-analysis
    Huijbregts, Henricus J. T. A. M.
    Khan, Riaz J. K.
    Sorensen, Emma
    Fick, Daniel P.
    Haebich, Samantha
    [J]. ACTA ORTHOPAEDICA, 2016, 87 (04) : 386 - 394
  • [16] Digital templating in total knee and hip replacement: an analysis of planning accuracy
    Kniesel, Bettina
    Konstantinidis, Lukas
    Hirschmueller, Anja
    Suedkamp, Norbert
    Helwig, Peter
    [J]. INTERNATIONAL ORTHOPAEDICS, 2014, 38 (04) : 733 - 739
  • [17] MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA
    LANDIS, JR
    KOCH, GG
    [J]. BIOMETRICS, 1977, 33 (01) : 159 - 174
  • [18] Risk of Anterior Femoral Notching in Navigated Total Knee Arthroplasty
    Lee, Ju Hong
    Wang, Seong-Il
    [J]. CLINICS IN ORTHOPEDIC SURGERY, 2015, 7 (02) : 217 - 224
  • [19] Levine Brett, 2010, Orthopedics, V33, P797, DOI 10.3928/01477447-20100924-04
  • [20] INFLUENCE OF POSITIONING OF PROSTHESIS IN TOTAL KNEE REPLACEMENT
    LOTKE, PA
    ECKER, ML