Optimizing the frame duration for data-driven rigid motion estimation in brain PET imaging

被引:10
作者
Spangler-Bickell, Matthew G. [1 ,2 ]
Hurley, Samuel A. [1 ]
Deller, Timothy W. [2 ]
Jansen, Floris [2 ]
Bettinardi, Valentino [3 ]
Carlson, Mackenzie [4 ]
Zeineh, Michael [4 ]
Zaharchuk, Greg [4 ]
McMillan, Alan B. [1 ]
机构
[1] Univ Wisconsin, Dept Radiol, Madison, WI 53706 USA
[2] GE Healthcare, PET MR Engn, Waukesha, WI 53188 USA
[3] Nucl Med Unit, IRCCS Osped San Raffaele, Milan, Italy
[4] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
关键词
brain imaging; data-driven motion estimation; list-mode; PET reconstruction; rigid motion correction; ultrashort frames; HEAD MOTION; REGISTRATION; RECONSTRUCTION; PERFORMANCE; IMAGES;
D O I
10.1002/mp.14889
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Data-driven rigid motion estimation for PET brain imaging is usually performed using data frames sampled at low temporal resolution to reduce the overall computation time and to provide adequate signal-to-noise ratio in the frames. In recent work it has been demonstrated that list-mode reconstructions of ultrashort frames are sufficient for motion estimation and can be performed very quickly. In this work we take the approach of using image-based registration of reconstructions of very short frames for data-driven motion estimation, and optimize a number of reconstruction and registration parameters (frame duration, MLEM iterations, image pixel size, post-smoothing filter, reference image creation, and registration metric) to ensure accurate registrations while maximizing temporal resolution and minimizing total computation time. Methods: Data from F-18-fluorodeoxyglucose (FDG) and F-18-florbetaben (FBB) tracer studies with varying count rates are analyzed, for PET/MR and PET/CT scanners. For framed reconstructions using various parameter combinations interframe motion is simulated and image-based registrations are performed to estimate that motion. Results: For FDG and FBB tracers using 4 x 10(5) true and scattered coincidence events per frame ensures that 95% of the registrations will be accurate to within 1 mm of the ground truth. This corresponds to a frame duration of 0.5-1 sec for typical clinical PET activity levels. Using four MLEM iterations with no subsets, a transaxial pixel size of 4 mm, a post-smoothing filter with 4-6 mm full width at half maximum, and averaging two or more frames to create the reference image provides an optimal set of parameters to produce accurate registrations while keeping the reconstruction and processing time low. Conclusions: It is shown that very short frames (<= 1 sec) can be used to provide accurate and quick data-driven rigid motion estimates for use in an event-by-event motion corrected reconstruction. (C) 2021 American Association of Physicists in Medicine
引用
收藏
页码:3031 / 3041
页数:11
相关论文
共 50 条
  • [41] Dual respiratory and cardiac motion estimation in PET imaging: Methods design and quantitative evaluation
    Feng, Tao
    Wang, Jizhe
    Tsui, Benjamin M. W.
    MEDICAL PHYSICS, 2018, 45 (04) : 1481 - 1490
  • [42] Markerless motion estimation for motion-compensated clinical brain imaging
    Kyme, Andre Z.
    Se, Stephen
    Meikle, Steven R.
    Fulton, Roger R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (10)
  • [43] Sign determination methods for the respiratory signal in data-driven PET gating
    Bertolli, Ottavia
    Arridge, Simon
    Wollenweber, Scott D.
    Stearns, Charles W.
    Hutton, Brian F.
    Thielemans, Kris
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08) : 3204 - 3220
  • [44] Data-driven gated PET/CT: implications for lesion segmentation and quantitation
    Thomas, M. Allan
    Pan, Tinsu
    EJNMMI PHYSICS, 2021, 8 (01)
  • [45] Carpe Datum: A Consideration of the Barriers and Potential of Data-Driven PET Innovation
    Kesner, Adam Leon
    Daou, Doumit
    Schindler, Thomas Hellmut
    Koo, Phillip J.
    JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2016, 13 (01) : 106 - 108
  • [46] Spherical Image Inpainting with Frame Transformation and Data-Driven Prior Deep Networks
    Li, Jianfei
    Huang, Chaoyan
    Chan, Raymond
    Feng, Han
    Ng, Michael K.
    Zeng, Tieyong
    SIAM JOURNAL ON IMAGING SCIENCES, 2023, 16 (03) : 1179 - 1196
  • [47] A Globally Optimal Data-Driven Approach for Image Distortion Estimation
    Tian, Yuandong
    Narasimhan, Srinivasa G.
    2010 IEEE CONFERENCE ON COMPUTER VISION AND PATTERN RECOGNITION (CVPR), 2010, : 1277 - 1284
  • [48] Development and Evaluation of Data-driven Respiratory Gating Methods with Simulated Listmode PET Data
    Wang, Jizhe
    Feng, Tao
    Tsui, Benjamin M. W.
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [49] On Data-Driven Sparse Sensing and Linear Estimation of Fluid Flows
    Jayaraman, Balaji
    Al Mamun, S. M. Abdullah
    SENSORS, 2020, 20 (13) : 1 - 31
  • [50] Evaluation of Frame-Based and Event-by-Event Motion-Correction Methods for Awake Monkey Brain PET Imaging
    Jin, Xiao
    Mulnix, Tim
    Sandiego, Christine M.
    Carson, Richard E.
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55 (02) : 287 - 293