Optimizing the frame duration for data-driven rigid motion estimation in brain PET imaging

被引:10
作者
Spangler-Bickell, Matthew G. [1 ,2 ]
Hurley, Samuel A. [1 ]
Deller, Timothy W. [2 ]
Jansen, Floris [2 ]
Bettinardi, Valentino [3 ]
Carlson, Mackenzie [4 ]
Zeineh, Michael [4 ]
Zaharchuk, Greg [4 ]
McMillan, Alan B. [1 ]
机构
[1] Univ Wisconsin, Dept Radiol, Madison, WI 53706 USA
[2] GE Healthcare, PET MR Engn, Waukesha, WI 53188 USA
[3] Nucl Med Unit, IRCCS Osped San Raffaele, Milan, Italy
[4] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
关键词
brain imaging; data-driven motion estimation; list-mode; PET reconstruction; rigid motion correction; ultrashort frames; HEAD MOTION; REGISTRATION; RECONSTRUCTION; PERFORMANCE; IMAGES;
D O I
10.1002/mp.14889
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Data-driven rigid motion estimation for PET brain imaging is usually performed using data frames sampled at low temporal resolution to reduce the overall computation time and to provide adequate signal-to-noise ratio in the frames. In recent work it has been demonstrated that list-mode reconstructions of ultrashort frames are sufficient for motion estimation and can be performed very quickly. In this work we take the approach of using image-based registration of reconstructions of very short frames for data-driven motion estimation, and optimize a number of reconstruction and registration parameters (frame duration, MLEM iterations, image pixel size, post-smoothing filter, reference image creation, and registration metric) to ensure accurate registrations while maximizing temporal resolution and minimizing total computation time. Methods: Data from F-18-fluorodeoxyglucose (FDG) and F-18-florbetaben (FBB) tracer studies with varying count rates are analyzed, for PET/MR and PET/CT scanners. For framed reconstructions using various parameter combinations interframe motion is simulated and image-based registrations are performed to estimate that motion. Results: For FDG and FBB tracers using 4 x 10(5) true and scattered coincidence events per frame ensures that 95% of the registrations will be accurate to within 1 mm of the ground truth. This corresponds to a frame duration of 0.5-1 sec for typical clinical PET activity levels. Using four MLEM iterations with no subsets, a transaxial pixel size of 4 mm, a post-smoothing filter with 4-6 mm full width at half maximum, and averaging two or more frames to create the reference image provides an optimal set of parameters to produce accurate registrations while keeping the reconstruction and processing time low. Conclusions: It is shown that very short frames (<= 1 sec) can be used to provide accurate and quick data-driven rigid motion estimates for use in an event-by-event motion corrected reconstruction. (C) 2021 American Association of Physicists in Medicine
引用
收藏
页码:3031 / 3041
页数:11
相关论文
共 50 条
  • [31] A Data-Driven Daylight Estimation Approach to Lighting Control
    Borile, Stefano
    Pandharipande, Ashish
    Caicedo, David
    Schenato, Luca
    Cenedese, Angelo
    IEEE ACCESS, 2017, 5 : 21461 - 21471
  • [32] Data-driven HRF estimation for encoding and decoding models
    Pedregosa, Fabian
    Eickenberg, Michael
    Ciuciu, Philippe
    Thirion, Bertrand
    Gramfort, Alexandre
    NEUROIMAGE, 2015, 104 : 209 - 220
  • [33] Estimation of and correction for finite motion sampling errors in small animal PET rigid motion correction
    Miranda, A.
    Staelens, S.
    Stroobants, S.
    Verhaeghe, J.
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2019, 57 (02) : 505 - 518
  • [34] Data-driven robust receding horizon fault estimation
    Wan, Yiming
    Keviczky, Tamas
    Verhaegen, Michel
    Gustafsson, Fredrik
    AUTOMATICA, 2016, 71 : 210 - 221
  • [35] Data-driven respiratory phase-matched PET attenuation correction without CT
    Hwang, Donghwi
    Kang, Seung Kwan
    Kim, Kyeong Yun
    Choi, Hongyoon
    Seo, Seongho
    Lee, Jae Sung
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (11)
  • [36] External radioactive markers for PET data-driven respiratory gating in positron emission tomography
    Buether, Florian
    Ernst, Iris
    Hamill, James
    Eich, Hans T.
    Schober, Otmar
    Schaefers, Michael
    Schaefers, Klaus P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 40 (04) : 602 - 614
  • [37] Denoising and reconstruction of 3D seismic data on a data-driven tight frame
    Chen J.
    Niu C.
    Li Y.
    Huang R.
    Chen L.
    Ma Z.
    1600, Science Press (55): : 725 - 732
  • [38] DATA-DRIVEN ILLUMINATION PATTERNS FOR CODED DIFFRACTION IMAGING
    Cai, Zikui
    Hyder, Rakib
    Asif, M. Salman
    2021 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP), 2021, : 2818 - 2822
  • [39] Motion correction of multi-frame PET data in neuroreceptor mapping: Simulation based validation
    Costes, Nicolas
    Dagher, Alain
    Larcher, Kevin
    Evans, Alan C.
    Collins, D. Louis
    Reilhac, Anthonin
    NEUROIMAGE, 2009, 47 (04) : 1496 - 1505
  • [40] Evaluation of motion correction methods in human brain PET imaging-A simulation study based on human motion data
    Jin, Xiao
    Mulnix, Tim
    Gallezot, Jean-Dominique
    Carson, Richard E.
    MEDICAL PHYSICS, 2013, 40 (10)