Optimizing the frame duration for data-driven rigid motion estimation in brain PET imaging

被引:10
|
作者
Spangler-Bickell, Matthew G. [1 ,2 ]
Hurley, Samuel A. [1 ]
Deller, Timothy W. [2 ]
Jansen, Floris [2 ]
Bettinardi, Valentino [3 ]
Carlson, Mackenzie [4 ]
Zeineh, Michael [4 ]
Zaharchuk, Greg [4 ]
McMillan, Alan B. [1 ]
机构
[1] Univ Wisconsin, Dept Radiol, Madison, WI 53706 USA
[2] GE Healthcare, PET MR Engn, Waukesha, WI 53188 USA
[3] Nucl Med Unit, IRCCS Osped San Raffaele, Milan, Italy
[4] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
关键词
brain imaging; data-driven motion estimation; list-mode; PET reconstruction; rigid motion correction; ultrashort frames; HEAD MOTION; REGISTRATION; RECONSTRUCTION; PERFORMANCE; IMAGES;
D O I
10.1002/mp.14889
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Data-driven rigid motion estimation for PET brain imaging is usually performed using data frames sampled at low temporal resolution to reduce the overall computation time and to provide adequate signal-to-noise ratio in the frames. In recent work it has been demonstrated that list-mode reconstructions of ultrashort frames are sufficient for motion estimation and can be performed very quickly. In this work we take the approach of using image-based registration of reconstructions of very short frames for data-driven motion estimation, and optimize a number of reconstruction and registration parameters (frame duration, MLEM iterations, image pixel size, post-smoothing filter, reference image creation, and registration metric) to ensure accurate registrations while maximizing temporal resolution and minimizing total computation time. Methods: Data from F-18-fluorodeoxyglucose (FDG) and F-18-florbetaben (FBB) tracer studies with varying count rates are analyzed, for PET/MR and PET/CT scanners. For framed reconstructions using various parameter combinations interframe motion is simulated and image-based registrations are performed to estimate that motion. Results: For FDG and FBB tracers using 4 x 10(5) true and scattered coincidence events per frame ensures that 95% of the registrations will be accurate to within 1 mm of the ground truth. This corresponds to a frame duration of 0.5-1 sec for typical clinical PET activity levels. Using four MLEM iterations with no subsets, a transaxial pixel size of 4 mm, a post-smoothing filter with 4-6 mm full width at half maximum, and averaging two or more frames to create the reference image provides an optimal set of parameters to produce accurate registrations while keeping the reconstruction and processing time low. Conclusions: It is shown that very short frames (<= 1 sec) can be used to provide accurate and quick data-driven rigid motion estimates for use in an event-by-event motion corrected reconstruction. (C) 2021 American Association of Physicists in Medicine
引用
收藏
页码:3031 / 3041
页数:11
相关论文
共 50 条
  • [1] Evaluation of Data-Driven Rigid Motion Correction in Clinical Brain PET Imaging
    Spangler-Bickell, Matthew G.
    Hurley, Samuel A.
    Pirasteh, Ali
    Perlman, Scott B.
    Deller, Timothy
    McMillan, Alan B.
    JOURNAL OF NUCLEAR MEDICINE, 2022, 63 (10) : 1604 - 1610
  • [2] Verification of the effect of data-driven brain motion correction on PET imaging
    Odagiri, Hayato
    Watabe, Hiroshi
    Takanami, Kentaro
    Akimoto, Kazuma
    Usui, Akihito
    Kawakami, Hirofumi
    Katsuki, Akie
    Uetake, Nozomu
    Dendo, Yutaka
    Tanaka, Yoshitaka
    Kodama, Hiroyasu
    Takase, Kei
    Kaneta, Tomohiro
    PLOS ONE, 2024, 19 (07):
  • [3] Improved frame-based estimation of head motion in PET brain imaging
    Mukherjee, J. M.
    Lindsay, C.
    Mukherjee, A.
    Olivier, P.
    Shao, L.
    King, M. A.
    Licho, R.
    MEDICAL PHYSICS, 2016, 43 (05) : 2443 - 2454
  • [4] Adaptive data-driven motion detection and optimized correction for brain PET
    Revilla, Enette Mae
    Gallezot, Jean-Dominique
    Naganawa, Mika
    Toyonaga, Takuya
    Fontaine, Kathryn
    Mulnix, Tim A.
    Onofrey, John E.
    Carson, Richard
    Lu, Yihuan
    NEUROIMAGE, 2022, 252
  • [5] Real-Time Data-Driven Rigid Motion Detection and Correction for Brain Scan with Listmode PET
    Feng, Tao
    Yang, Defu
    Zhu, Wentao
    Dong, Yun
    Li, Hongdi
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [6] Data-driven, projection-based respiratory motion compensation of PET data for cardiac PET/CT and PET/MR imaging
    Lassen, Martin Lyngby
    Beyer, Thomas
    Berger, Alexander
    Beitzke, Dietrich
    Rasul, Sazan
    Buether, Florian
    Hacker, Marcus
    Cal-Gonzalez, Jacobo
    JOURNAL OF NUCLEAR CARDIOLOGY, 2020, 27 (06) : 2216 - 2230
  • [7] Rigid motion tracking using moments of inertia in TOF-PET brain studies
    Rezaei, Ahmadreza
    Spangler-Bickell, Matthew
    Schramm, Georg
    Van Laere, Koen
    Nuyts, Johan
    Defrise, Michel
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (18)
  • [8] Zero-Extra-Dose PET Delayed Imaging with Data-Driven Attenuation Correction Estimation
    Pang, Lifang
    Zhu, Wentao
    Dong, Yun
    Lv, Yang
    Shi, Hongcheng
    MOLECULAR IMAGING AND BIOLOGY, 2019, 21 (01) : 149 - 158
  • [9] Rigid motion correction of PET and CT for PET/CT brain imaging
    Bickell, Matthew
    Kim, Jung-Ha
    Rezaei, Ahmadreza
    Nuyts, Johan
    Fulton, Roger
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [10] Data-driven, projection-based respiratory motion compensation of PET data for cardiac PET/CT and PET/MR imaging
    Martin Lyngby Lassen
    Thomas Beyer
    Alexander Berger
    Dietrich Beitzke
    Sazan Rasul
    Florian Büther
    Marcus Hacker
    Jacobo Cal-González
    Journal of Nuclear Cardiology, 2020, 27 : 2216 - 2230