Non-invasive monitoring method for lower-leg compartment syndrome using a wireless sensor system and finite element analysis

被引:3
作者
Chang, Fuh-Yu [1 ]
Teng, Ping-Tun [1 ]
Chen, Liang-Chun [1 ]
机构
[1] Taiwan Univ Sci & Technol, Dept Mech Engn, 43,Sect 4,Keelung Rd, Taipei 106, Taiwan
关键词
Compartment syndrome; non-invasive pressure detect; pressure sensor; finite element method; smartphone;
D O I
10.1177/0954411920981243
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, a non-invasive pressure monitoring system that is portable and convenient was designed for detecting compartment syndrome. The system combines a wireless module and smartphone, which aids in the achievement of mHealth objectives, specifically, the continuous monitoring of the compartment pressure in patients. A compartment syndrome detecting method using a wireless sensor system and finite element analysis is developed and verified with an in vitro lower-leg model by rapid prototyping. The sensor system is designed to measure a five point pressure variation from the outside of the lower leg and transmit the data to a smartphone via Bluetooth. The analysis model based on the finite element method is employed to calculate the change of pressure and volume inside the four compartments of the lower leg. The in vitro experimental results show that the non-invasive detecting method can monitor the compartment pressure and provide a warning for the occurrence of compartment syndrome if the compartment pressure is higher than 30 mmHg. Furthermore, the theoretical simulation of the real lower leg shows similar trends to those of the in vitro experiments and can promptly detect the occurrence of compartment syndrome. Measured pressure values exceeding 6.3, 2.7, and 2.8 kPa for the three sensors contacting the outside centers of the superficial posterior, anterior, and lateral compartments, respectively, can indicate that each compartment contains a pressure higher than 30 mmHg. These results can provide a warning for the risk of compartment syndrome of each compartment. In addition, the measured values from the three sensors contacting the superficial posterior compartment at the outside center, close to the tibia, and close to the lateral compartment exceeding 1.8, 0.7, and 0.7 kPa, respectively, can indicate the risk of deep posterior compartment syndrome.
引用
收藏
页码:346 / 356
页数:11
相关论文
共 19 条
  • [1] [Anonymous], 2014, THESIS
  • [2] Three-dimensional finite element analysis of the foot during standing - a material sensitivity study
    Cheung, JTM
    Zhang, M
    Leung, AKL
    Fan, YB
    [J]. JOURNAL OF BIOMECHANICS, 2005, 38 (05) : 1045 - 1054
  • [3] Dickson KF, 2003, ORTHOPEDICS, V26, P1215
  • [4] Accelerometers and force sensing resistors for optimal control of walking of a hemiplegic
    Dosen, Strahinja
    Popovic, Dejan B.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (08) : 1973 - 1984
  • [5] Finite element modeling of soft tissues: Material models, tissue interaction and challenges
    Freutel, Maren
    Schmidt, Hendrik
    Duerselen, Lutz
    Ignatius, Anita
    Galbusera, Fabio
    [J]. CLINICAL BIOMECHANICS, 2014, 29 (04) : 363 - 372
  • [6] Near-infrared spectroscopy versos compartment pressure for the diagnosis of lower extremity compartmental syndrome using electromyography-determined measurements of neuromuscular function
    Gentilello, LM
    Sanzone, A
    Wang, LL
    Liu, PY
    Robinson, L
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2001, 51 (01): : 1 - 8
  • [7] Determining Level of Postural Control in Young Adults Using Force-Sensing Resistors
    Gopalai, Alpha Agape
    Senanayake, S. M. N. Arosha
    Gouwanda, Darwin
    [J]. IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, 2011, 15 (04): : 608 - 614
  • [8] Prevalence and patterns of foot injuries following motorcycle trauma
    Jeffers, RF
    Tan, HB
    Nicolopoulos, C
    Kamath, R
    Giannoudis, PV
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (02) : 87 - 91
  • [9] Lambert A., 1995, PHYS THER REHABIL J, V81, P234