Accuracy of S2 Alar-Iliac Screw Placement Under the Guidance of a 3D-Printed Surgical Guide Template

被引:5
|
作者
Zhao, Yonghui [1 ,2 ]
Luo, Haotian [1 ]
Ma, Yulong [2 ]
Liang, Jinlong [2 ]
Han, Gonghai [1 ]
Xu, Yongqing [2 ]
Lu, Sheng [1 ]
机构
[1] Kunming Univ Sci & Technol, Peoples Hosp Yunnan Prov 1, Dept Orthoped, Key Lab Digital Orthoped Yunnan Prov,Affiliated H, Kunming, Yunnan, Peoples R China
[2] 920 Hosp Joint Logist Support Force, Dept Orthoped, Kunming, Yunnan, Peoples R China
基金
中国国家自然科学基金;
关键词
3D printing technology; Accuracy; S2AI screw; Surgical guide template; MINIMUM 5-YEAR ANALYSIS; SACROPELVIC FIXATION; PELVIC FIXATION; LONG FUSIONS; SPINE; ADULT; SACRUM; COMPLICATIONS; S1;
D O I
10.1016/j.wneu.2020.10.063
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: To evaluate the safety and accuracy of S2 alar-iliac (S2AI) screw placement guided by a 3-dimensional (3D)-printed surgical guide template. - METHODS: The data of 27 patients treated with S2AI screws were analyzed. S2AI surgical guide templates were designed and printed, and S2AI screw placement was completed intraoperatively with the guide templates. Postoperative computed tomography was performed to measure screw path parameters, namely the sagittal angle (SA), the transverse angle (TA), the horizontal distance (HD) between the entry point of the screw and the median sacral crest, and the vertical distance (VD) between the entry point of the screw and the lower edge of the first posterior sacral foramen. Screw placement was graded according to the Oh grading criteria. RESULTS: A total of 54 S2AI screws were placed. The screw grades were as follows: 52 screws were considered grade 0, 2 were grade 1, none were grade 2, and none were grade 3. Thus grade 0 accounted for 96.3% of the screws. When the preoperatively planned SA (32.3 degrees +/- 2.0 degrees), TA (42.1 degrees +/- 3.9 degrees), HD (5.1 +/- 1.1) mm, and VD (19.0 +/- 2.4) mm were compared with the corresponding postoperative SA ( 31.9 degrees +/- 3.8 degrees), TA (42.5 degrees +/- 4.0 degrees), HD (4.9 +/- 1.1) mm, and VD (19.1 +/- 2.3) mm, no significant differences were identified (P > 0.05). CONCLUSIONS: S2AI screw placement assisted by a 3D-printed surgical guide is safe and accurate.
引用
收藏
页码:E161 / E167
页数:7
相关论文
共 43 条
  • [1] S2 Alar-Iliac Screw Insertion: Technical Note with Pictorial Guide
    Yilmaz, Emre
    Abdul-Jabbar, Amir
    Tawfik, Tamir
    Iwanaga, Joe
    Schmidt, Cameron K.
    Chapman, Jens
    Blecher, Ronen
    Tubbs, R. Shane
    Oskouian, Rod J.
    WORLD NEUROSURGERY, 2018, 113 : E296 - E301
  • [2] Iliac screw versus S2 alar-iliac screw fixation in adults: a meta-analysis
    Ramos, Rafael De la Garza
    Nakhla, Jonathan
    Sciubba, Daniel M.
    Yassari, Reza
    JOURNAL OF NEUROSURGERY-SPINE, 2019, 30 (02) : 253 - 258
  • [3] S2 alar iliac screw placement under robotic guidance for adult spinal deformity patients: technical note
    Hyun, Seung-Jae
    Kim, Ki-Jeong
    Jahng, Tae-Ahn
    EUROPEAN SPINE JOURNAL, 2017, 26 (08) : 2198 - 2203
  • [4] Comparison of the 3D-printed operation guide template technique and the free-hand technique for S2-alar-iliac screw placement
    Zhao, Yonghui
    Ma, Yulong
    Liang, Jinlong
    Luo, Haotian
    Cai, Xingbo
    Xu, Yongqing
    Lu, Sheng
    BMC SURGERY, 2020, 20 (01)
  • [5] Freehand S2 Alar-Iliac Screw Placement Using K-Wire and Cannulated Screw : Technical Case Series
    Choi, Ho Yong
    Hyun, Seung-Jae
    Kim, Ki-Jeong
    Jahng, Tae-Ahn
    Kim, Hyun-Jib
    JOURNAL OF KOREAN NEUROSURGICAL SOCIETY, 2018, 61 (01) : 75 - 80
  • [6] The degenerative changes of the sacroiliac joint after S2 alar-iliac screw placement
    Ikeda, Norimasa
    Fujibayashi, Shunsuke
    Otsuki, Bungo
    Masamoto, Kazutaka
    Shimizu, Takayoshi
    Shimizu, Yu
    Murata, Koichi
    Matsuda, Shuichi
    JOURNAL OF NEUROSURGERY-SPINE, 2022, 36 (02) : 287 - 293
  • [7] In-Human Report of S2 Alar-Iliac Screw Placement Using Augmented Reality Assistance
    Judy, Brendan F.
    Liu, Ann
    Jin, Yike
    Ronkon, Charles
    Khan, Majid
    Cottrill, Ethan
    Ehresman, Jeff
    Pennington, Zach
    Bydon, Ali
    Lo, Sheng-Fu L.
    Sciubba, Daniel M.
    Molina, Camilo A.
    Witham, Timothy F.
    OPERATIVE NEUROSURGERY, 2023, 24 (01) : 68 - 73
  • [8] Efficacy and safety of lumbopelvic fixation in spinal metastasis comparing S2 Alar-iliac screw and conventional iliac screw
    Akkarawanit, Pawin
    Suvithayasiri, Siravich
    Santipas, Borriwat
    Wilartratsami, Sirichai
    Luksanapruksa, Panya
    EUROPEAN SPINE JOURNAL, 2025,
  • [9] The three-dimensional printed template guided technique for S2 alar iliac screw placement and a comparison with freehand technique
    Zhou, Zhenhai
    Zeng, Zhimin
    Yu, Honggui
    Xiong, Jiachao
    Liu, Zhiming
    Zhou, Rongping
    Wan, Wenbing
    Pan, Zhimin
    Chen, Lu
    Cao, Kai
    JOURNAL OF ORTHOPAEDIC SURGERY, 2020, 28 (03)
  • [10] Fixation Strength of Modified Iliac Screw Trajectory Compared to Traditional Iliac and S2 Alar-Iliac Trajectories: A Cadaveric Study
    Von Glinski, Alexander
    Pierre, Clifford
    Frieler, Sven
    Mahoney, Jonathan M.
    Harris, Jonathan A.
    Amin, Dhara B.
    Allall, May
    Bucklen, Brandon S.
    Schildhauer, Thomas A.
    Oskouian, Rod J.
    Chapman, Jens R.
    WORLD NEUROSURGERY, 2021, 154 : E481 - E487