Target dose accuracy of single-isocenter stereotactic body radiation therapy for multiple lung lesions

被引:1
作者
Shao, Y. [1 ,2 ]
Chen, H. [1 ,2 ]
Wang, H. [1 ,2 ]
Cheng, Y. [2 ]
Zhu, Z. [2 ]
Zhuo, W. [3 ]
Kong, Q. [1 ]
Xu, Z. [2 ]
机构
[1] Fudan Univ, Inst Modern Phys, Shanghai 200433, Peoples R China
[2] Shanghai Jiao Tong Univ, Shanghai Chest Hosp, Dept Radiat Oncol, Shanghai 20030, Peoples R China
[3] Fudan Univ, Inst Radiat Med, Shanghai 20032, Peoples R China
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2022年 / 20卷 / 01期
关键词
Multiple lung lesions; single-isocenter; SBRT; dose accuracy; ARC THERAPY; RADIOSURGERY; QUALITY; CANCER; SBRT;
D O I
10.52547/ijrr.20.1.37
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: To investigate the maximum distance from lesions to the single-isocenter that can ensure the accuracy of dose delivery for single-isocenter/two-lesions intensity modulated radiotherapy (IMRT) lung stereotactic body radiation therapy (SBRT). Materials and Methods: We employed a lung phantom made of acrylic material. Sixteen gross tumor volumes (GTVs) of the same shape and size were delineated on the CT images. The single-isocenter was placed on the center of the first GTV (GTV1). Six GTV centers were located at left and were 2-7 cm away from the single-isocenter. Nine GTV centers were in the 45 degrees direction and were 2-10 cm away from the single-isocenter. Plans were created for the first planning target volume (PTV1) and other fifteen PTVs separately with the same isocenter. Compass was used to verify the treatment plans.. analysis was carried out with criteria of 2% / 2 mm. The passing rate shall over than 90%. If the passing rate of. analysis was lower than 90%, two-isocenter plan was created and carried out. analysis as well. Results: According to the. analysis, when the distance from lesions to the single-isocenter was more than 6cm, both in the horizontal or 45 degrees direction, the passing rate was lower than 90%. Based on. analysis, two-isocenter plans were created. After verification, the passing rate was higher than 90%. Conclusions: The maximum distance of using singleisocenter to treat multiple lung lesions is 5 cm. The result provides a reference for our center and other centers when using this technique to treat multiple lung lesions.
引用
收藏
页码:241 / 244
页数:4
相关论文
共 50 条
  • [41] Comparative dosimetric analysis of single versus multiple isocenter stereotactic body radiotherapy for liver metastasis
    Jayadevan, P. M.
    Sudesh, N. S.
    Shine, N. S.
    Nithin, K.
    Dhanya
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2025, 11 (01):
  • [42] Comparison of Doses Received by the Hippocampus in Patients Treated with Single Vs Multiple Isocenter Based Stereotactic Radiation Therapy to the Brain for Multiple Brain Metastases
    Algan, O.
    Giem, J.
    Young, J.
    Ali, I.
    Ahmad, S.
    Hossain, S.
    MEDICAL PHYSICS, 2014, 41 (06) : 240 - 240
  • [43] Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D
    McCulloch, James
    Pawlowski, Jamie
    Kirby, Neil
    Rasmussen, Karl
    Shi, Zheng
    Myers, Pamela
    Stathakis, Sotirios
    Papanikolaou, Niko
    Saenz, Daniel L.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (09): : 107 - 115
  • [44] Four-dimensional planning for motion synchronized dose delivery in lung stereotactic body radiation therapy
    Tachibana, Hidenobu
    Sawant, Amit
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 (03) : 467 - 472
  • [45] OBSERVATION OF A DOSE-CONTROL RELATIONSHIP FOR LUNG AND LIVER TUMORS AFTER STEREOTACTIC BODY RADIATION THERAPY
    McCammon, Robert
    Schefter, Tracey E.
    Gaspar, Laurie E.
    Zaemisch, Rebekah
    Gravdahl, Daniel
    Kavanagh, Brian
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (01): : 112 - 118
  • [46] Potentially curative stereotactic body radiation therapy (SBRT) for single or oligometastasis to the lung
    Oh, Dongryul
    Ahn, Yong Chan
    Seo, Jeong Min
    Shin, Eun Hyuk
    Park, Hee Chul
    Lim, Do Hoon
    Pyo, Hongryull
    ACTA ONCOLOGICA, 2012, 51 (05) : 596 - 602
  • [47] Radiosensitivity of Lung Metastases by Primary Histology and Implications for Stereotactic Body Radiation Therapy Using the Genomically Adjusted Radiation Dose
    Ahmed, Kamran A.
    Scott, Jacob G.
    Arrington, John A.
    Naghavi, Arash O.
    Grass, G. Daniel
    Perez, Bradford A.
    Caudell, Jimmy J.
    Berglund, Anders E.
    Welsh, Eric A.
    Eschrich, Steven A.
    Dilling, Thomas J.
    Torres-Roca, Javier F.
    JOURNAL OF THORACIC ONCOLOGY, 2018, 13 (08) : 1121 - 1127
  • [48] SINGLE-ISOCENTER FRAMELESS INTENSITY-MODULATED STEREOTACTIC RADIOSURGERY FOR SIMULTANEOUS TREATMENT OF MULTIPLE BRAIN METASTASES: CLINICAL EXPERIENCE
    Nath, Sameer K.
    Lawson, Joshua D.
    Simpson, Daniel R.
    VanderSpek, Lauren
    Wang, Jia-Zhu
    Alksne, John F.
    Ciacci, Joseph
    Mundt, Arno J.
    Murphy, Kevin T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (01): : 91 - 97
  • [49] Feasibility and dosimetric evaluation of single- and multi-isocentre stereotactic body radiation therapy for multiple liver metastases
    Tang, Chunbo
    Gong, Changfei
    Liu, Biaoshui
    Guo, Hailiang
    Dai, Zhongyang
    Yuan, Jun
    Wang, Xiaoping
    Zhang, Yun
    FRONTIERS IN ONCOLOGY, 2023, 13
  • [50] Impact of lung tumor motion on dose delivered to organ at risk in lung stereotactic body radiation therapy
    Savanovic, Milovan
    Strbac, Bojan
    Jaros, Drazan
    Cazic, Dejan
    Foulquier, Jean Noel
    JOURNAL OF RADIATION ONCOLOGY, 2020, 9 (3-4) : 191 - 198