Target dose accuracy of single-isocenter stereotactic body radiation therapy for multiple lung lesions

被引:1
作者
Shao, Y. [1 ,2 ]
Chen, H. [1 ,2 ]
Wang, H. [1 ,2 ]
Cheng, Y. [2 ]
Zhu, Z. [2 ]
Zhuo, W. [3 ]
Kong, Q. [1 ]
Xu, Z. [2 ]
机构
[1] Fudan Univ, Inst Modern Phys, Shanghai 200433, Peoples R China
[2] Shanghai Jiao Tong Univ, Shanghai Chest Hosp, Dept Radiat Oncol, Shanghai 20030, Peoples R China
[3] Fudan Univ, Inst Radiat Med, Shanghai 20032, Peoples R China
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2022年 / 20卷 / 01期
关键词
Multiple lung lesions; single-isocenter; SBRT; dose accuracy; ARC THERAPY; RADIOSURGERY; QUALITY; CANCER; SBRT;
D O I
10.52547/ijrr.20.1.37
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: To investigate the maximum distance from lesions to the single-isocenter that can ensure the accuracy of dose delivery for single-isocenter/two-lesions intensity modulated radiotherapy (IMRT) lung stereotactic body radiation therapy (SBRT). Materials and Methods: We employed a lung phantom made of acrylic material. Sixteen gross tumor volumes (GTVs) of the same shape and size were delineated on the CT images. The single-isocenter was placed on the center of the first GTV (GTV1). Six GTV centers were located at left and were 2-7 cm away from the single-isocenter. Nine GTV centers were in the 45 degrees direction and were 2-10 cm away from the single-isocenter. Plans were created for the first planning target volume (PTV1) and other fifteen PTVs separately with the same isocenter. Compass was used to verify the treatment plans.. analysis was carried out with criteria of 2% / 2 mm. The passing rate shall over than 90%. If the passing rate of. analysis was lower than 90%, two-isocenter plan was created and carried out. analysis as well. Results: According to the. analysis, when the distance from lesions to the single-isocenter was more than 6cm, both in the horizontal or 45 degrees direction, the passing rate was lower than 90%. Based on. analysis, two-isocenter plans were created. After verification, the passing rate was higher than 90%. Conclusions: The maximum distance of using singleisocenter to treat multiple lung lesions is 5 cm. The result provides a reference for our center and other centers when using this technique to treat multiple lung lesions.
引用
收藏
页码:241 / 244
页数:4
相关论文
共 50 条
  • [31] Monte Carlo evaluation of target dose coverage in lung stereotactic body radiation therapy with flattening filter-free beams
    Vassiliev, Oleg N.
    Peterson, Christine B.
    Chang, Joe Y.
    Mohan, Radhe
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2022, 21 (01) : 81 - 87
  • [32] Comparison of doses received by the hippocampus in patients treated with single isocenter- vs multiple isocenter-based stereotactic radiation therapy to the brain for multiple brain metastases
    Algan, Ozer
    Giem, Jared
    Young, Julie
    Ali, Imad
    Ahmad, Salahuddin
    Hossain, Sabbir
    MEDICAL DOSIMETRY, 2015, 40 (04) : 314 - 317
  • [33] Stereotactic radiosurgery for multiple small brain metastases using gamma knife versus single-isocenter VMAT: Normal brain dose based on lesion number and size
    Abdou, Abram
    Almeida, Timoteo
    Bossart, Elizabeth
    Monterroso, Irene
    Mellon, Eric A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025,
  • [34] Predicting voxel-level dose distributions of single-isocenter volumetric modulated arc therapy treatment plan for multiple brain metastases
    Huang, Peng
    Shang, Jiawen
    Hu, Zhihui
    Liu, Zhiqiang
    Yan, Hui
    FRONTIERS IN ONCOLOGY, 2024, 14
  • [35] Potential reduction of lung dose via VMAT with jaw tracking in the treatment of single-isocenter/two-lesion lung SBRT
    Pokhrel, Damodar
    Sanford, Lana
    Halfman, Matthew
    Molloy, Janelle
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (05): : 55 - 63
  • [36] Stereotactic Body Radiation Therapy for Oligometastases to the Lung: A Phase 2 Study
    Nuyttens, Joost J.
    van Zyp, Noelle C. M. G. van der Voort
    Verhoef, Cornelis
    Maat, A.
    van Klaveren, Robertus J.
    van der Holt, Bronno
    Aerts, Joachim
    Hoogeman, Mischa
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 91 (02): : 337 - 343
  • [37] Stereotactic body radiation therapy in the treatment of multiple primary lung cancers
    Creach, Kimberly M.
    Bradley, Jeffrey D.
    Mahasittiwat, Pawinee
    Robinson, Clifford G.
    RADIOTHERAPY AND ONCOLOGY, 2012, 104 (01) : 19 - 22
  • [38] Multicomponent mathematical model for tumor volume calculation with setup error using single-isocenter stereotactic radiotherapy for multiple brain metastases
    Nakano, Hisashi
    Shiinoki, Takehiro
    Tanabe, Satoshi
    Nakano, Toshimichi
    Takizawa, Takeshi
    Utsunomiya, Satoru
    Sakai, Madoka
    Tanabe, Shunpei
    Ohta, Atsushi
    Kaidu, Motoki
    Nishio, Teiji
    Ishikawa, Hiroyuki
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (02) : 945 - 953
  • [39] Variation of the prescription dose using the analytical anisotropic algorithm in lung stereotactic body radiation therapy
    Kawai, Daisuke
    Takahashi, Ryo
    Kamima, Tatsuya
    Baba, Hiromi
    Yamamoto, Toshijiro
    Kubo, Yoko
    Ishibashi, Satoru
    Higuchi, Yoshihiro
    Tani, Kensuke
    Mizuno, Norifumi
    Jinno, Shunta
    Tachibana, Hidenobu
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 38 : 98 - 104
  • [40] Dose Tolerance for Stereotactic Body Radiation Therapy
    Grimm, Jimm
    SEMINARS IN RADIATION ONCOLOGY, 2016, 26 (02) : 87 - 88