Usefulness of a Metal Artifact Reduction Algorithm for Orthopedic Implants in Abdominal CT: Phantom and Clinical Study Results

被引:43
作者
Jeong, Seonji [1 ]
Kim, Se Hyung [1 ,2 ]
Hwang, Eui Jin [1 ]
Shin, Cheong-Il [1 ,2 ]
Han, Joon Koo [1 ,2 ]
Choi, Byung Ihn [1 ,2 ]
机构
[1] Seoul Natl Univ Hosp, Dept Radiol, Seoul 110744, South Korea
[2] Seoul Natl Univ Hosp, Inst Radiat Med, Seoul 110744, South Korea
基金
新加坡国家研究基金会;
关键词
CT; iterative reduction algorithm; metal artifact; phantom; TUBE CURRENT CT; COMPUTED-TOMOGRAPHY; SUPPRESSION; RECONSTRUCTION; ABDOMEN;
D O I
10.2214/AJR.14.12745
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to evaluate the usefulness of a metal artifact reduction (MAR) algorithm for orthopedic prostheses in phantom and clinical CT. MATERIALS AND METHODS. An agar phantom with two sets of spinal screws was scanned at various tube voltage (80-140 kVp) and tube current-time (34-1032 mAs) settings. The orthopedic MAR algorithm was combined with filtered back projection (FBP) or iterative reconstruction. The mean SDs in three ROIs were compared among four datasets (FBP, iterative reconstruction, FBP with orthopedic MAR, and iterative reconstruction with orthopedic MAR). For the clinical study, the mean SDs of three ROIs and 4-point scaled image quality in 52 patients with metallic orthopedic prostheses were compared between CT images acquired with and without orthopedic MAR. The presence and type of image quality improvement with orthopedic MAR and the presence of orthopedic MAR-related new artifacts were also analyzed. RESULTS. In the phantom study, the mean SD with orthopedic MAR was significantly lower than that without orthopedic MAR regardless of dose settings and reconstruction algorithms (FBP versus iterative reconstruction). The mean SD near the metallic prosthesis in 52 patients was significantly lower on CT images with orthopedic MAR (28.04 HU) than those without it (49.21 HU). Image quality regarding metallic artifact was significantly improved with orthopedic MAR (rating of 2.60 versus 1.04). Notable reduction of metallic artifacts and better depiction of abdominal organs were observed in 45 patients. Diagnostic benefit was achieved in six patients, but orthopedic MAR-related new artifacts were seen in 30 patients. CONCLUSION. Use of the orthopedic MAR algorithm significantly reduces metal artifacts in CT of both phantoms and patients and has potential for improving diagnostic performance in patients with severe metallic artifacts.
引用
收藏
页码:307 / 317
页数:11
相关论文
共 31 条
  • [21] Scibelli A, IDOSE4 ITERATIVE REC
  • [22] Stereotactic body radiation therapy in multiple organ sites
    Timmerman, Robert D.
    Kavanagh, Brian D.
    Cho, L. Chinsoo
    Papiez, Lech
    Xing, Lei
    [J]. JOURNAL OF CLINICAL ONCOLOGY, 2007, 25 (08) : 947 - 952
  • [23] Iterative correction of beam hardening artifacts in CT
    Van Gompel, G.
    Van Slambrouck, K.
    Defrise, M.
    Batenburg, K. J.
    de Mey, J.
    Sijbers, J.
    Nuyts, J.
    [J]. MEDICAL PHYSICS, 2011, 38 : S36 - S49
  • [24] Development and validation of segmentation and interpolation techniques in sinograms for metal artifact suppression in CT
    Veldkamp, Wouter J. H.
    Joemai, Raoul M. S.
    van der Molen, Aart J.
    Geleijns, Jacob
    [J]. MEDICAL PHYSICS, 2010, 37 (02) : 620 - 628
  • [25] Iterative X-ray cone-beam tomography for metal artifact reduction and local region reconstruction
    Wang, G
    Vannier, MW
    Cheng, PC
    [J]. MICROSCOPY AND MICROANALYSIS, 1999, 5 (01) : 58 - 65
  • [26] Metal artifacts reduction using monochromatic images from spectral CT: Evaluation of pedicle screws in patients with scoliosis
    Wang, Yang
    Qian, Bangping
    Li, Baoxin
    Qin, Guochu
    Zhou, Zhengyang
    Qiu, Yong
    Sun, Xizhao
    Zhu, Bin
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (08) : E360 - E366
  • [27] A pragmatic approach to metal artifact reduction in CT: merging of metal artifact reduced images
    Watzke, O
    Kalender, WA
    [J]. EUROPEAN RADIOLOGY, 2004, 14 (05) : 849 - 856
  • [28] Comparison of intensity-modulated radiotherapy planning based on manual and automatically generated contours using deformable image registration in four-dimensional computed tomography of lung cancer patients
    Weiss, Elisabeth
    Wijesooriya, Krishni
    Ramakrishnan, Viswanathan
    Keall, Paul J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (02): : 572 - 581
  • [29] A methodology for image quality evaluation of advanced CT systems
    Wilson, Joshua M.
    Christianson, Olav I.
    Richard, Samuel
    Samei, Ehsan
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [30] Metal Artifact Reduction From Reformatted Projections for Hip Prostheses in Multislice Helical Computed Tomography Techniques and Initial Clinical Results
    Yu, Lifeng
    Li, Hua
    Mueller, Jan
    Kofler, James M.
    Liu, Xin
    Primak, Andrew N.
    Fletcher, Joel G.
    Guimaraes, Luis S.
    Macedo, Thanila
    McCollough, Cynthia H.
    [J]. INVESTIGATIVE RADIOLOGY, 2009, 44 (11) : 691 - 696