Usefulness of a Metal Artifact Reduction Algorithm for Orthopedic Implants in Abdominal CT: Phantom and Clinical Study Results

被引:41
作者
Jeong, Seonji [1 ]
Kim, Se Hyung [1 ,2 ]
Hwang, Eui Jin [1 ]
Shin, Cheong-Il [1 ,2 ]
Han, Joon Koo [1 ,2 ]
Choi, Byung Ihn [1 ,2 ]
机构
[1] Seoul Natl Univ Hosp, Dept Radiol, Seoul 110744, South Korea
[2] Seoul Natl Univ Hosp, Inst Radiat Med, Seoul 110744, South Korea
基金
新加坡国家研究基金会;
关键词
CT; iterative reduction algorithm; metal artifact; phantom; TUBE CURRENT CT; COMPUTED-TOMOGRAPHY; SUPPRESSION; RECONSTRUCTION; ABDOMEN;
D O I
10.2214/AJR.14.12745
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to evaluate the usefulness of a metal artifact reduction (MAR) algorithm for orthopedic prostheses in phantom and clinical CT. MATERIALS AND METHODS. An agar phantom with two sets of spinal screws was scanned at various tube voltage (80-140 kVp) and tube current-time (34-1032 mAs) settings. The orthopedic MAR algorithm was combined with filtered back projection (FBP) or iterative reconstruction. The mean SDs in three ROIs were compared among four datasets (FBP, iterative reconstruction, FBP with orthopedic MAR, and iterative reconstruction with orthopedic MAR). For the clinical study, the mean SDs of three ROIs and 4-point scaled image quality in 52 patients with metallic orthopedic prostheses were compared between CT images acquired with and without orthopedic MAR. The presence and type of image quality improvement with orthopedic MAR and the presence of orthopedic MAR-related new artifacts were also analyzed. RESULTS. In the phantom study, the mean SD with orthopedic MAR was significantly lower than that without orthopedic MAR regardless of dose settings and reconstruction algorithms (FBP versus iterative reconstruction). The mean SD near the metallic prosthesis in 52 patients was significantly lower on CT images with orthopedic MAR (28.04 HU) than those without it (49.21 HU). Image quality regarding metallic artifact was significantly improved with orthopedic MAR (rating of 2.60 versus 1.04). Notable reduction of metallic artifacts and better depiction of abdominal organs were observed in 45 patients. Diagnostic benefit was achieved in six patients, but orthopedic MAR-related new artifacts were seen in 30 patients. CONCLUSION. Use of the orthopedic MAR algorithm significantly reduces metal artifacts in CT of both phantoms and patients and has potential for improving diagnostic performance in patients with severe metallic artifacts.
引用
收藏
页码:307 / 317
页数:11
相关论文
共 31 条
  • [1] [Anonymous], MET ART RED ORTH IMP
  • [2] Penalized-Likelihood Reconstruction for Metal Artifact Reduction in Cone-Beam CT
    Aootaphao, Sorapong
    Pintavirooj, Chuchart
    Sotthivirat, Saowapak
    [J]. 2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, : 2733 - 2736
  • [3] Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation
    Bamberg, Fabian
    Dierks, Alexander
    Nikolaou, Konstantin
    Reiser, Maximilian F.
    Becker, Christoph R.
    Johnson, Thorsten R. C.
    [J]. EUROPEAN RADIOLOGY, 2011, 21 (07) : 1424 - 1429
  • [4] Artifacts in CT: Recognition and avoidance
    Barrett, JF
    Keat, N
    [J]. RADIOGRAPHICS, 2004, 24 (06) : 1679 - 1691
  • [5] Evaluation of Two Iterative Techniques for Reducing Metal Artifacts in Computed Tomography
    Boas, F. Edward
    Fleischmann, Dominik
    [J]. RADIOLOGY, 2011, 259 (03) : 894 - 902
  • [6] An iterative maximum-likelihood polychromatic algorithm for CT
    De Man, B
    Nuyts, J
    Dupont, P
    Marchal, G
    Suetens, P
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (10) : 999 - 1008
  • [7] REDUCTION OF CT ARTIFACTS CAUSED BY METALLIC IMPLANTS
    KALENDER, WA
    HEBEL, R
    EBERSBERGER, J
    [J]. RADIOLOGY, 1987, 164 (02) : 576 - 577
  • [8] Comparison of z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis
    Kalra, MK
    Maher, MM
    Toth, TL
    Kamath, RS
    Halpern, EF
    Saini, S
    [J]. RADIOLOGY, 2004, 232 (02) : 347 - 353
  • [9] Reduction of dental metallic artefacts in CT: Value of a newly developed algorithm for metal artefact reduction (O-MAR)
    Kidoh, M.
    Nakaura, T.
    Nakamura, S.
    Tokuyasu, S.
    Osakabe, H.
    Harada, K.
    Yamashita, Y.
    [J]. CLINICAL RADIOLOGY, 2014, 69 (01) : E11 - E16
  • [10] Koehler T., 2011, INT C XRAY COMP TOM