Per-organ assessment of subject-induced susceptibility distortion for MR-only male pelvis treatment planning

被引:12
作者
Glide-Hurst, Carri [1 ]
Nejad-Davarani, Siamak [1 ]
Weiss, Steffen [2 ]
Zheng, Weili [3 ]
Chetty, Indrin J. [1 ]
Renisch, Steffen [2 ]
机构
[1] Henry Ford Canc Inst, Dept Radiat Oncol, Detroit, MI 48202 USA
[2] Philips Res Labs, Dept Digital Imaging, D-22335 Hamburg, Germany
[3] William Beaumont Hosp, Dept Radiat Oncol, Royal Oak, MI 48073 USA
来源
RADIATION ONCOLOGY | 2018年 / 13卷
基金
美国国家卫生研究院;
关键词
CONFORMAL RADIATION-THERAPY; PROSTATE-CANCER; GEOMETRIC DISTORTION; BRAIN IMAGES; REGISTRATION; OPTIMIZATION; RADIOTHERAPY; SIMULATION; SCANNER; PHANTOM;
D O I
10.1186/s13014-018-1090-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Patient-specific distortions, particularly near tissue/air interfaces, require assessment for magnetic resonance (MR) only radiation treatment planning (RTP). However, patients are dynamic due to changes in physiological status during imaging sessions. This work investigated changes in subject-induced susceptibility distortions to pelvic organs at different bladder states to support pelvis MR-only RTP. Methods: Pelvises of 9 healthy male volunteers were imaged at 1.0 Tesla (T),1.5 T, and 3.0 T. Subject-induced susceptibility distortion field maps were generated using a dual-echo gradient-recalled echo (GRE) sequence with B-0 field maps obtained from the phase difference between the two echoes acquired at several bladder volume states (3-4/subject, 32 overall). T2 turbo spin echo images were also acquired at each bladder state for organ delineation. Magnet central frequency was tracked over time. Distortion map differences and boxplots were computed to characterize changes within the clinical target volume (CTV), bladder, seminal vesicles, and prostate volumes. Results: The time between the initial and final BO maps was 42.6 +/- 13.9 (range: 13.2-62.1) minutes with minimal change in magnet central frequency (0.02 +/- 0.05 mm (range: - 0.06 - 0.12 mm)). Subject-induced susceptibility distortion across all bladder states, field strengths, and subjects was relatively small (1.4-1.9% of all voxels in the prostate and seminal vesicles were distorted > 0.5 mm). In the bladder, no voxels exhibited distortions > 1 mm. An extreme case acquired at 3.0 T with a large volume of rectal air yielded 27.4-34.6% of voxels within the CTVs had susceptibility-induced distortions > 0.5 mm across all time points. Conclusions: Our work suggests that subject-induced susceptibility distortions caused by bladder/rectal conditions are generally small and subject-dependent Local changes may be non-negligible within the CTV, thus proper management of filling status is warranted. Future work evaluating the impact of multiple models to accommodate for extreme status changes may be advantageous.
引用
收藏
页数:9
相关论文
共 29 条
[1]   Quantifying the Effect of 3T Magnetic Resonance Imaging Residual System Distortions and Patient-Induced Susceptibility Distortions on Radiation Therapy Treatment Planning for Prostate Cancer [J].
Adjeiwaah, Mary ;
Bylund, Mikael ;
Lundman, Josef A. ;
Karlsson, Camilla Thellenberg ;
Jonsson, Joakim H. ;
Nyholm, Tufve .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 100 (02) :317-324
[2]   Characterization, prediction, and correction of geometric distortion in 3 T MR images [J].
Baldwin, Lesley N. ;
Wachowicz, Keith ;
Thomas, Steven D. ;
Rivest, Ryan ;
Fallone, B. Gino .
MEDICAL PHYSICS, 2007, 34 (02) :388-399
[3]   A two-step scheme for distortion rectification of magnetic resonance images [J].
Baldwin, Lesley N. ;
Wachowicz, Keith ;
Fallone, B. Gino .
MEDICAL PHYSICS, 2009, 36 (09) :3917-3926
[4]   Real-time RF pulse adjustment for B0 drift correction [J].
Benner, Thomas ;
van der Kouwe, Andre J. W. ;
Kirsch, John E. ;
Sorensen, A. Gregory .
MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (01) :204-209
[5]   An evaluation of four CT-MRI co-registration techniques for radiotherapy treatment planning of prone rectal cancer patients [J].
Dean, C. J. ;
Sykes, J. R. ;
Cooper, R. A. ;
Hatfield, P. ;
Carey, B. ;
Swift, S. ;
Bacon, S. E. ;
Thwaites, D. ;
Sebag-Montefiore, D. ;
Morgan, A. M. .
BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1009) :61-68
[6]   A complete distortion correction for MR images: I. Gradient warp correction [J].
Doran, SJ ;
Charles-Edwards, L ;
Reinsberg, SA ;
Leach, MO .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (07) :1343-1361
[7]   A review of substitute CT generation for MRI-only radiation therapy [J].
Edmund, Jens M. ;
Nyholm, Tufve .
RADIATION ONCOLOGY, 2017, 12
[8]   Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group* (I):: concepts and terms in 3D image based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV [J].
Haie-Meder, C ;
Pötter, R ;
Van Limbergen, E ;
Briot, E ;
De Brabandere, M ;
Dimopoulos, J ;
Dumas, I ;
Hellebust, TP ;
Kirisits, C ;
Lang, SF ;
Muschitz, S ;
Nevinson, J ;
Nulens, A ;
Petrow, P ;
Wachter-Gerstner, N .
RADIOTHERAPY AND ONCOLOGY, 2005, 74 (03) :235-245
[9]   Phantom-based characterization of distortion on a magnetic resonance imaging simulator for radiation oncology [J].
Huang, Ke ;
Cao, Yue ;
Baharom, Umar ;
Balter, James M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (02) :774-790
[10]   Improved optimization for the robust and accurate linear registration and motion correction of brain images [J].
Jenkinson, M ;
Bannister, P ;
Brady, M ;
Smith, S .
NEUROIMAGE, 2002, 17 (02) :825-841