Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-LASIK eyes for myopia

被引:12
|
作者
Al-Mezaine, Hani S. [1 ]
Al-Amro, Saleh A. [1 ]
Kangave, Dustan [2 ]
Al-Obeidan, Saleh [1 ]
Al-Jubair, Khaled M. [3 ]
机构
[1] King Saud Univ, Coll Med, Dept Ophthalmol, Riyadh 11321, Saudi Arabia
[2] King Saud Univ, Coll Med, Res Dept, Riyadh 11321, Saudi Arabia
[3] Natl Guard Hosp, Dept Ophthalmol, Riyadh, Saudi Arabia
关键词
Corneal thickness; Laser in situ keratomileusis; Pachymetry; Pentacam; Ultrasound; NONCONTACT SPECULAR MICROSCOPY; IN-SITU KERATOMILEUSIS; ANTERIOR-CHAMBER DEPTH; REFRACTIVE SURGERY; ORBSCAN-II; SCHEIMPFLUG PHOTOGRAPHY; OCULUS PENTACAM; REPEATABILITY; PACHOMETRY; OXYBUPROCAINE;
D O I
10.1177/112067211002000507
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To compare central corneal thickness (CCT) measurements obtained using the Pentacam Scheimpflug system with those obtained using DGH ultrasound pachymetry (UP) in post-laser in situ keratomileusis (LASIK) eyes for myopia. METHODS. In a prospective study, measurement agreement was assessed in 143 eyes of 72 post-LASIK patients using both the Pentacam and UP at the Eye Consultants Center in Riyadh, Saudi Arabia. RESULTS. The mean CCT was 522+/-42.2 mu m with the Oculus Pentacam and 516.2+/-40.6 mu m with UP. The Bland-Altman plot showed that the mean +/-SD for the differences between the 2 devices was 5.8+/-13.6 mu m, with 95% confidence interval limits ranging from -20.9 mu m to 32.6 mu m. A test of statistical significance indicated that the mean differences of 5.8+/-13.6 mu m differed significantly from zero (p<0.001; Wilcoxon signed-rank test), thus indicating that the Pentacam measurements tended to overestimate CCT compared with UP. Analysis of regression showed a high correlation between the values obtained with both devices (r=0.947, p<0.001). CONCLUSIONS. In post-LASIK myopic eyes, although a high correlation has been shown between Pentacam and UP measurements, Pentacam tends to overestimate CCT compared to UP. Pentacam probably cannot be used interchangeably with UP in post-LASIK eyes for myopia. (Eur J Ophthalmol 2010; 20: 852-7)
引用
收藏
页码:852 / 857
页数:6
相关论文
共 50 条
  • [11] Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post–LASIK or PRK, and keratoconic or keratoconus-suspect eyes
    Wenjing Wu
    Yan Wang
    Lulu Xu
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2014, 252 : 91 - 99
  • [12] The Shape of Posterior Corneal Surface in Normal, Post-LASIK, and Post-epi-LASIK Eyes
    Zhang, Lin
    Wang, Yan
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (07) : 3468 - 3475
  • [13] Measurement of central corneal thickness using Orbscan 3, Pentacam HR and ultrasound pachymetry in normal eyes
    Gharieb, Hesham Mohamed
    Ashour, Doaa Maamoun
    Saleh, Mohamed Ibrahim
    Othman, Ihab Saad
    INTERNATIONAL OPHTHALMOLOGY, 2020, 40 (07) : 1759 - 1764
  • [14] Comparison of corneal thickness measurements using ultrasound pachymetry, noncontact tonopachy, Pentacam HR, and Fourier-domain OCT
    Kim, Jung Sub
    Rho, Chang Rae
    Cho, Yeon Woo
    Shin, Jeongah
    MEDICINE, 2021, 100 (16) : E25638
  • [15] Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes
    Yu, Ayong
    Zhao, Weiqi
    Savini, Giacomo
    Huang, Zixu
    Bao, Fangjun
    Lu, Weicong
    Wang, Qinmei
    Huang, Jinhai
    JOURNAL OF OPHTHALMOLOGY, 2015, 2015
  • [16] Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia
    Ho, Thomas
    Cheng, Arthur C. K.
    Rao, Srinivas K.
    Lau, Silvania
    Leung, Chris K. S.
    Lam, Dennis S. C.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (07) : 1177 - 1182
  • [17] Comparison of Central Corneal Thickness Measurement Using Ultrasonic Pachymetry, Rotating Scheimpflug Camera, and Scanning-Slit Topography
    Sedaghat, Mohammad Reza
    Daneshvar, Ramin
    Kargozar, Abbas
    Derakhshan, Akbar
    Daraei, Mona
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2010, 150 (06) : 780 - 789
  • [18] Measurement of Central Corneal Thickness With Optical Low-Coherence Reflectometry and Ultrasound Pachymetry in Normal and Post-Femtosecond Laser in Situ Keratomileusis Eyes
    Huang, Jinhai
    Liao, Na
    Savini, Giacomo
    Li, Yuanguang
    Bao, Fangjun
    Yu, Ye
    Yu, Ayong
    Wang, Qinmei
    CORNEA, 2015, 34 (02) : 204 - 208
  • [19] Comparison of central corneal thickness measurements with standard ultrasonic pachymetry and optical devices
    Dogan, Mustafa
    Ertan, Elif
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2019, 102 (02) : 126 - 130
  • [20] Repeatability of Corneal Epithelial Thickness Measurements Using Fourier-Domain Optical Coherence Tomography in Normal and Post-LASIK Eyes
    Ma, Xingxuan Jack
    Wang, Li
    Koch, Douglas D.
    CORNEA, 2013, 32 (12) : 1544 - 1548