Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-LASIK eyes for myopia

被引:12
|
作者
Al-Mezaine, Hani S. [1 ]
Al-Amro, Saleh A. [1 ]
Kangave, Dustan [2 ]
Al-Obeidan, Saleh [1 ]
Al-Jubair, Khaled M. [3 ]
机构
[1] King Saud Univ, Coll Med, Dept Ophthalmol, Riyadh 11321, Saudi Arabia
[2] King Saud Univ, Coll Med, Res Dept, Riyadh 11321, Saudi Arabia
[3] Natl Guard Hosp, Dept Ophthalmol, Riyadh, Saudi Arabia
关键词
Corneal thickness; Laser in situ keratomileusis; Pachymetry; Pentacam; Ultrasound; NONCONTACT SPECULAR MICROSCOPY; IN-SITU KERATOMILEUSIS; ANTERIOR-CHAMBER DEPTH; REFRACTIVE SURGERY; ORBSCAN-II; SCHEIMPFLUG PHOTOGRAPHY; OCULUS PENTACAM; REPEATABILITY; PACHOMETRY; OXYBUPROCAINE;
D O I
10.1177/112067211002000507
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To compare central corneal thickness (CCT) measurements obtained using the Pentacam Scheimpflug system with those obtained using DGH ultrasound pachymetry (UP) in post-laser in situ keratomileusis (LASIK) eyes for myopia. METHODS. In a prospective study, measurement agreement was assessed in 143 eyes of 72 post-LASIK patients using both the Pentacam and UP at the Eye Consultants Center in Riyadh, Saudi Arabia. RESULTS. The mean CCT was 522+/-42.2 mu m with the Oculus Pentacam and 516.2+/-40.6 mu m with UP. The Bland-Altman plot showed that the mean +/-SD for the differences between the 2 devices was 5.8+/-13.6 mu m, with 95% confidence interval limits ranging from -20.9 mu m to 32.6 mu m. A test of statistical significance indicated that the mean differences of 5.8+/-13.6 mu m differed significantly from zero (p<0.001; Wilcoxon signed-rank test), thus indicating that the Pentacam measurements tended to overestimate CCT compared with UP. Analysis of regression showed a high correlation between the values obtained with both devices (r=0.947, p<0.001). CONCLUSIONS. In post-LASIK myopic eyes, although a high correlation has been shown between Pentacam and UP measurements, Pentacam tends to overestimate CCT compared to UP. Pentacam probably cannot be used interchangeably with UP in post-LASIK eyes for myopia. (Eur J Ophthalmol 2010; 20: 852-7)
引用
收藏
页码:852 / 857
页数:6
相关论文
共 50 条
  • [1] Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes
    Al-Ageel, Saleh
    Al-Muammar, Abdulrahman M.
    SAUDI JOURNAL OF OPHTHALMOLOGY, 2009, 23 (3-4) : 181 - 187
  • [2] Agreement of post-LASIK corneal power and corneal thickness measurements by pentacam and GALILEI corneal tomography systems
    Hsieh, Yun-Hsiu
    Weng, Tzu-Heng
    Chou, Yu-Ching
    Wu, Kuan-Liang
    Liang, Chang-Min
    Tai, Ming-Cheng
    JOURNAL OF THE CHINESE MEDICAL ASSOCIATION, 2019, 82 (01) : 72 - 77
  • [3] Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry
    Al-Mezaine H.S.
    Al-Amro S.A.
    Kangave D.
    Sadaawy A.
    Wehaib T.A.
    Al-Obeidan S.
    International Ophthalmology, 2008, 28 (5) : 333 - 338
  • [4] Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes
    Wu, Wenjing
    Wang, Yan
    Xu, Lulu
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2014, 252 (01) : 91 - 99
  • [5] Comparison of central corneal thickness measurement by scanning slit topography, infrared, and ultrasound pachymetry in normal and post-LASIK eyes
    Ho, Wing-cheung
    Lam, Philip Tsze-ho
    Chiu, Thomas Yee-hang
    Yim, Mandy Ching-man
    Lau, Fion Tung-ching
    INTERNATIONAL OPHTHALMOLOGY, 2020, 40 (11) : 2913 - 2921
  • [6] Comparison of Central Corneal Thickness Measurements in Normal and Keratoconic Eyes Using Ultrasonic Pachymetry and OCULUS Pentacam
    Buyuk, Kursat
    Bozkurt, Banu
    Kamis, Umit
    Ozkagnici, Ahmet
    Okudan, Suleyman
    TURK OFTALMOLOJI DERGISI-TURKISH JOURNAL OF OPHTHALMOLOGY, 2011, 41 (02): : 104 - 107
  • [7] Comparison of Corneal Thickness Measurements Using Galilei, HR Pentacam, and Ultrasound
    Hosseini, Hamid Reza Jahadi
    Katbab, Asadollah
    Khalili, Mohammad Reza
    Abtahi, Mohammad Bagher
    CORNEA, 2010, 29 (10) : 1091 - 1095
  • [8] Reliability of the Corneal Thickness Measurements With the Pentacam HR Imaging System and Ultrasound Pachymetry
    Modis, Laszlo, Jr.
    Szalai, Eszter
    Nemeth, Gabor
    Berta, Andras
    CORNEA, 2011, 30 (05) : 561 - 566
  • [9] Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter
    Yekta, Abbas-Ali
    Hashemi, Hassan
    KhabazKhoob, Mehdi
    Dostdar, Asghar
    Mehravaran, Shiva
    Heravian, Javad
    Fotouhi, Akbar
    IRANIAN JOURNAL OF OPHTHALMOLOGY, 2009, 21 (02): : 51 - 57
  • [10] Comparison between Pentacam-HR and optical coherence tomography central corneal thickness measurements in healthy feline eyes
    Cleymaet, Allison M.
    Hess, Ann M.
    Freeman, Kate S.
    VETERINARY OPHTHALMOLOGY, 2016, 19 : 105 - 114