Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-LASIK eyes for myopia

被引:12
作者
Al-Mezaine, Hani S. [1 ]
Al-Amro, Saleh A. [1 ]
Kangave, Dustan [2 ]
Al-Obeidan, Saleh [1 ]
Al-Jubair, Khaled M. [3 ]
机构
[1] King Saud Univ, Coll Med, Dept Ophthalmol, Riyadh 11321, Saudi Arabia
[2] King Saud Univ, Coll Med, Res Dept, Riyadh 11321, Saudi Arabia
[3] Natl Guard Hosp, Dept Ophthalmol, Riyadh, Saudi Arabia
关键词
Corneal thickness; Laser in situ keratomileusis; Pachymetry; Pentacam; Ultrasound; NONCONTACT SPECULAR MICROSCOPY; IN-SITU KERATOMILEUSIS; ANTERIOR-CHAMBER DEPTH; REFRACTIVE SURGERY; ORBSCAN-II; SCHEIMPFLUG PHOTOGRAPHY; OCULUS PENTACAM; REPEATABILITY; PACHOMETRY; OXYBUPROCAINE;
D O I
10.1177/112067211002000507
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To compare central corneal thickness (CCT) measurements obtained using the Pentacam Scheimpflug system with those obtained using DGH ultrasound pachymetry (UP) in post-laser in situ keratomileusis (LASIK) eyes for myopia. METHODS. In a prospective study, measurement agreement was assessed in 143 eyes of 72 post-LASIK patients using both the Pentacam and UP at the Eye Consultants Center in Riyadh, Saudi Arabia. RESULTS. The mean CCT was 522+/-42.2 mu m with the Oculus Pentacam and 516.2+/-40.6 mu m with UP. The Bland-Altman plot showed that the mean +/-SD for the differences between the 2 devices was 5.8+/-13.6 mu m, with 95% confidence interval limits ranging from -20.9 mu m to 32.6 mu m. A test of statistical significance indicated that the mean differences of 5.8+/-13.6 mu m differed significantly from zero (p<0.001; Wilcoxon signed-rank test), thus indicating that the Pentacam measurements tended to overestimate CCT compared with UP. Analysis of regression showed a high correlation between the values obtained with both devices (r=0.947, p<0.001). CONCLUSIONS. In post-LASIK myopic eyes, although a high correlation has been shown between Pentacam and UP measurements, Pentacam tends to overestimate CCT compared to UP. Pentacam probably cannot be used interchangeably with UP in post-LASIK eyes for myopia. (Eur J Ophthalmol 2010; 20: 852-7)
引用
收藏
页码:852 / 857
页数:6
相关论文
共 22 条
[1]   Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry [J].
Al-Mezaine H.S. ;
Al-Amro S.A. ;
Kangave D. ;
Sadaawy A. ;
Wehaib T.A. ;
Al-Obeidan S. .
International Ophthalmology, 2008, 28 (5) :333-338
[2]   Corneal thickness values before and after oxybuprocaine 0.4% eye drops [J].
Asensio, I ;
Rahhal, SM ;
Alonso, L ;
Palanca-Sanfrancisco, JM ;
Sanchis-Gimeno, JA .
CORNEA, 2003, 22 (06) :527-532
[3]   Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry [J].
Barkana, Y ;
Gerber, Y ;
Elbaz, U ;
Schwartz, S ;
Ken-Dror, G ;
Avni, I ;
Zadok, D .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09) :1729-1735
[4]   Comparison of three methods of measuring corneal thickness and anterior chamber depth [J].
Buehl, W ;
Stojanac, D ;
Sacu, S ;
Drexler, W ;
Findl, O .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 141 (01) :7-12
[5]   Comparison of corneal thickness measurements by ultrasound and scheimpflug photography in eyes that have undergone laser in situ keratomileusis [J].
Ciolino, Joseph B. ;
Khachikian, Steven S. ;
Belin, Michael W. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 145 (01) :75-80
[6]   Comparison of Pentacam Scheimpflug camera with ultrasound pachymetry and 0 noncontact specular Microscopy in measuring central corneal thickness [J].
Fujioka, Miyuki ;
Nakamura, Makoto ;
Tatsumi, Yasuko ;
Kusuhara, Azusa ;
Maeda, Hidetaka ;
Negi, Akira .
CURRENT EYE RESEARCH, 2007, 32 (02) :89-94
[7]   VARIABILITY OF ULTRASONIC PACHOMETRY [J].
GORDON, A ;
BOGGESS, EA ;
MOLINARI, JF .
OPTOMETRY AND VISION SCIENCE, 1990, 67 (03) :162-165
[8]   Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia [J].
Hashemi, Hassan ;
Mehravaran, Shiva .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (10) :1701-1707
[9]  
HERSE P, 1992, ACTA OPHTHALMOL, V70, P740
[10]   Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia [J].
Ho, Thomas ;
Cheng, Arthur C. K. ;
Rao, Srinivas K. ;
Lau, Silvania ;
Leung, Chris K. S. ;
Lam, Dennis S. C. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (07) :1177-1182