Biomechanical Evaluation of Screw-In Femoral Implant in Cementless Total Hip System

被引:13
作者
Kim, James Y. [1 ]
Hayashi, Kei [1 ]
Garcia, Tanya C. [1 ]
Kim, Sun-Young [1 ]
Entwistle, Rachel [1 ]
Kapatkin, Amy S. [1 ]
Stover, Susan M. [1 ]
机构
[1] Univ Calif Davis, Sch Vet Med, Dept Surg & Radiol Sci, JD Wheat Vet Orthoped Res Lab, Davis, CA 95616 USA
关键词
STRAIN DISTRIBUTION; HINDLIMB LAMENESS; GAIT ANALYSIS; CANINE FEMUR; CANAL FILL; DOGS; REPLACEMENT; FORCES; ARTHROPLASTY; PROSTHESIS;
D O I
10.1111/j.1532-950X.2011.00890.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective: To compare (1) proximal femoral axial strains, (2) femoral head deflection, and (3) failure mechanical properties, between Helica head and neck prosthesis implanted femora and normal femora. Study Design: In vitro study. Sample Population: Cadaveric canine femora (n = 5 pair). Methods: Femoral bone strains and head displacement during in vitro simulation of midstance of the gallop were evaluated using cadaveric femurs cyclically loaded in vitro. Strains and displacements were compared within femurs, before and after, prosthesis implantation; and throughout cycling to seek evidence of movement with cyclic loading. Subsequently, implanted femurs and contralateral, intact femurs were loaded to failure to compare failure mechanical properties and modes of failure. Results: Proximal femoral axial strains were significantly different between intact and implanted femora on all 4 cortical surfaces (P < .05). Compressive strains were lower in the implanted femur on all cortical surfaces, except on the caudal surface which was higher. No difference was noted for femoral head angle under an axial load corresponding to gallop (P > .05). Vertical head displacement was similar to 0.1 mm greater for implanted femora than intact femora (P < .05). Yield and failure loads and yield energy of implanted femora were 39-54% lower than those for intact femora (P < .05). Mode of failure for both the intact and implanted femora did not appear to be different. Conclusion: Helica femoral prosthesis alters strain distribution in the proximal aspect of the femur and exhibits initial micromotion. Failure load in axial compression of the Helica-implanted femur is less than that of the normal femur, but greater than that expected in vivo.
引用
收藏
页码:94 / 102
页数:9
相关论文
共 32 条
[1]   Sciatic neurapraxia as a complication in 1000 consecutive canine total hip replacements [J].
Andrews, Chad M. ;
Liska, William D. ;
Roberts, Darryl J. .
VETERINARY SURGERY, 2008, 37 (03) :254-262
[2]  
Bergh MS, 2006, VET COMP ORTHOPAED, V19, P172
[3]   A COMPARISON OF HIP-JOINT FORCES IN SHEEP, DOG AND MAN [J].
BERGMANN, G ;
SIRAKY, J ;
ROHLMANN, A ;
KOELBEL, R .
JOURNAL OF BIOMECHANICS, 1984, 17 (12) :907-&
[4]  
Currey J D, 1970, Clin Orthop Relat Res, V73, P209
[5]   Changes in strain distribution of loaded proximal femora caused by different types of cementless femoral stems [J].
Decking, Ralf ;
Puhl, Wolfhart ;
Simon, Ulrich ;
Claes, Lutz E. .
CLINICAL BIOMECHANICS, 2006, 21 (05) :495-501
[6]   Evaluation of risk factors for luxation after total hip replacement in dogs [J].
Dyce, J ;
Wisner, ER ;
Wang, Q ;
Olmstead, ML .
VETERINARY SURGERY, 2000, 29 (06) :524-532
[7]  
Edwards MR, 1997, J AM VET MED ASSOC, V211, P580
[8]   Initial experience with a newly developed cementless hip endoprosthesis [J].
Hach, V. ;
Delfs, G. .
VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2009, 22 (02) :153-158
[9]   Decrease in canine proximal femoral ultimate strength and stiffness due to fatigue damage [J].
Hoshaw, SJ ;
Cody, DD ;
Saad, AM ;
Fyhrie, DP .
JOURNAL OF BIOMECHANICS, 1997, 30 (04) :323-329
[10]  
JEVENS DJ, 1993, AM J VET RES, V54, P612