Long-term implant-bone fixation of the femoral component in total knee replacement

被引:29
作者
Cristofolini, L. [1 ,2 ]
Affatato, S. [1 ]
Erani, P. [1 ]
Leardini, W. [1 ]
Tigani, D. [3 ]
Viceconti, M. [1 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Bologna, Fac Engn, I-40126 Bologna, Italy
[3] Ist Ortoped Rizzoli, Div 7, Bologna, Italy
关键词
cemented total knee replacement; TKR femoral component loosening; long-term implant fixation; bone-prosthesis micromotion; in vitro fatigue testing; migration pattern;
D O I
10.1243/09544119JEIM328
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Success of total knee replacement (TKR) depends on the prosthetic design. Aseptic loosening of the femoral component is a significant failure mode that has received little attention. Despite the clinical relevance of failures, no protocol is available to test long-term implant-bone fixation of TKR in vitro. The scope of this work was to develop and validate a protocol to assess pre-clinically the fixation of TKR femoral components. An in vitro protocol was designed to apply a simplified but relevant loading profile using a 6-degrees-of-freedom knee simulator for 1 000 000 cycles. Implant-bone inducible micromotions and permanent migrations were measured at three locations throughout the test. After test completion, fatigue damage in the cement was quantified. The developed protocol was successfully applied to a commercial TKR. Additional tests were performed to exclude artefacts due to swelling or creep of the composite femur models. The components migrated distally; they tilted towards valgus in the frontal plane and in extension in the sagittal plane. The migration patterns were consistent with clinical roentgen-stereophotogrammetric recordings with TKR. Additional indicators were proposed that could quantify the tendency to loosen/stabilize. The type and amount of damage found in the cement, as well as the migration patterns, were consistent with clinical experience with the specific TKR investigated. The proposed pre-clinical test yielded repeatable results, which were consistent with the clinical literature. Therefore, its relevance and reliability was proved.
引用
收藏
页码:319 / 331
页数:13
相关论文
共 37 条
[1]   A different fixation of the femoral component in total knee arthroplasty may lead to preservation of femoral bone stock [J].
Barink, M ;
Verdonschot, N ;
Malefijt, MD .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2003, 217 (H5) :325-332
[2]  
BARINK M, 2006, THESIS RADBOUD U NIJ
[3]  
BERNASEK TL, 1988, CLIN ORTHOP RELAT R, P52
[4]   Preclinical testing of femoral hip components: An experimental investigation with four prostheses [J].
Britton, JR ;
Prendergast, PJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (05) :872-880
[5]   MICROMOTION OF THE TOTAL KNEE [J].
BROSTROM, LA ;
GOLDIE, I ;
SELVIK, G .
ACTA ORTHOPAEDICA SCANDINAVICA, 1989, 60 (04) :443-445
[6]   Preclinical assessment of the long-term endurance of cemented hip stems. Part 1: effect of daily activities a comparison of two load histories [J].
Cristofolini, L. ;
Teutonico, A. Saponara ;
Savigni, P. ;
Erani, P. ;
Viceconti, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H6) :569-584
[7]   Preclinical assessment of the long-term endurance of cemented hip stems.: Part 2:: in-vitro and ex-vivo fatigue damage of the cement mantle [J].
Cristofolini, L. ;
Erani, P. ;
Savigni, P. ;
Bordini, B. ;
Viceconti, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H6) :585-599
[8]   Mechanical validation of whole bone composite femur models [J].
Cristofolini, L ;
Viceconti, M ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 1996, 29 (04) :525-535
[9]   Comparative in vitro study on the long term performance of cemented hip stems: validation of a protocol to discriminate between "good" and "bad" designs [J].
Cristofolini, L ;
Teutonico, AS ;
Monti, L ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 2003, 36 (11) :1603-1615
[10]  
CRISTOFOLINI L, 2002, 13 C EUR SOC BIOM CL, P48