Parallel cascades: New ways to describe noise transfer in medical imaging systems

被引:73
作者
Yao, J
Cunningham, IA
机构
[1] John P Robarts Res Inst, Imaging Res Labs, London, ON N6A 5K8, Canada
[2] London Hlth Sci Ctr, Dept Diagnost Radiol, London, ON, Canada
[3] Univ Western Ontario, London, ON, Canada
关键词
random point processes; amplified point processes; cross covariance; cross spectral density; noise power spectrum; transfer theory;
D O I
10.1118/1.1405842
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A generalized approach to describing transfer of the noise power spectrum through medical imaging systems has been developed over the past several years in which image-forming processes are represented in terms of a cascade of amplified point processes. Until recently, this approach has been restricted to serial cascades only. Here we develop a generalized expression for the cross covariance of amplified point processes and an expression for the cross spectral density for wide-sense stationary conditions. These results extend the generalized transfer-theory approach to include the description of more complex image-forming processes involving parallel cascades of quantum amplification processes. This parallel-cascade approach is used to develop a theoretical expression for noise-power transfer in a simple radiographic screen that includes the effect of characteristic x-ray reabsorption. The result confirms earlier work by Metz and Vyborny, who showed that reabsorption increases image noise and decreases the detective quantum efficiency at low spatial frequencies. Use of the transfer-theory approach facilitates a straightforward generalization to many new digital imaging systems including conventional angiographic and active-matrix flat-panel systems. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:2020 / 2038
页数:19
相关论文
共 39 条
[1]   Correlated point processes in radiological imaging [J].
Barrett, HH ;
Wagner, RF ;
Myers, KJ .
PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 :110-124
[2]  
Barrett HH., 1981, RADIOLOGICAL IMAGING
[3]  
Barrett HH., 2001, IMAGE SCI MATH STAT
[4]   A quantum accounting and detective quantum efficiency analysis for video-based portal imaging [J].
Bissonnette, JP ;
Cunningham, IA ;
Jaffray, DA ;
Fenster, A ;
Munro, P .
MEDICAL PHYSICS, 1997, 24 (06) :815-826
[5]  
Bracewell R. N., 1986, FOURIER TRANSFORM IT, V31999
[6]   ANALYSIS OF THE DETECTIVE QUANTUM EFFICIENCY OF A RADIOGRAPHIC SCREEN FILM COMBINATION [J].
BUNCH, PC ;
HUFF, KE ;
VANMETTER, R .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05) :902-909
[7]   Detective quantum efficiency dependence on x-ray energy weighting in mammography [J].
Cahn, RN ;
Cederström, B ;
Danielsson, M ;
Hall, A ;
Lundqvist, M ;
Nygren, D .
MEDICAL PHYSICS, 1999, 26 (12) :2680-2683
[8]   Linear-systems modeling of parallel cascaded stochastic processes: The NPS of radiographic screens with reabsorption of characteristic x radiation [J].
Cunningham, IA .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :220-230
[9]   Signal-to-noise optimization of medical imaging systems [J].
Cunningham, IA ;
Shaw, R .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1999, 16 (03) :621-632
[10]  
CUNNINGHAM IA, 2000, PHYS PSYCHOPHYSICS, V1, P79