A disposable on-line microsystem for continuous sampling and monitoring of glucose

被引:22
作者
Suzuki, H
Tokuda, T
Miyagishi, T
Yoshida, H
Honda, N
机构
[1] Univ Tsukuba, Inst Mat Sci, Tsukuba, Ibaraki 3058573, Japan
[2] Yamatake Corp, Fujisawa, Kanagawa 2518522, Japan
基金
日本学术振兴会;
关键词
gel; phase transition; sampling mechanism; mosquito; microneedle; glucose sensor;
D O I
10.1016/j.snb.2003.07.014
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A copolymerized gel composed of poly(N-isopropylacrylamide) and acrylic acid was used to construct a mechanism that withdrew sample solution slowly. A gel of 11 x 11m x 4 mm was used and a volume change was caused by changing the temperature, changing the temperature and pH, and using a pH change caused by the enzymatic reaction of glucose oxidase. With the enzyme-loaded gel, the gel shrank most slowly spanning nearly 10 h. The volume change of the gel could be converted into a pressure change in a How channel with the help of a silicone rubber diaphragm. As a result, the mechanism could withdraw an external solution into the flow channel. The sampling mechanism was incorporated in a system that imitates a mosquito. An external glucose standard solution could be introduced into the system autonomously and the glucose concentration could be monitored continuously using an on-chip micro glucose sensor. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:90 / 97
页数:8
相关论文
共 9 条
[1]   VOLUME-PHASE TRANSITIONS OF IONIZED N-ISOPROPYLACRYLAMIDE GELS [J].
HIROTSU, S ;
HIROKAWA, Y ;
TANAKA, T .
JOURNAL OF CHEMICAL PHYSICS, 1987, 87 (02) :1392-1395
[2]   A new monolithic micro biosensor for blood analysis [J].
Kim, JH ;
Kim, BG ;
Yoon, E ;
Han, CH .
14TH IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 2001, :443-446
[3]   A sampling mechanism employing the phase transition of a gel and its application to a micro analysis system imitating a mosquito [J].
Kobayashi, K ;
Suzuki, H .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 80 (01) :1-8
[4]   Influence of charge density on the swelling of colloidal poly(N-isopropylacrylamide-co-acrylic acid) microgels [J].
Kratz, K ;
Hellweg, T ;
Eimer, W .
COLLOIDS AND SURFACES A-PHYSICOCHEMICAL AND ENGINEERING ASPECTS, 2000, 170 (2-3) :137-149
[5]   KINETICS OF DISCONTINUOUS VOLUME PHASE-TRANSITION OF GELS [J].
MATSUO, ES ;
TANAKA, T .
JOURNAL OF CHEMICAL PHYSICS, 1988, 89 (03) :1695-1703
[6]  
Oka K, 2001, TRANSDUCERS '01: EUROSENSORS XV, DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, P412
[7]   Microfabricated liquid junction Ag/AgCl reference electrode and its application to a one-chip potentiometric sensor [J].
Suzuki, H ;
Shiroishi, H ;
Sasaki, S ;
Karube, I .
ANALYTICAL CHEMISTRY, 1999, 71 (22) :5069-5075
[8]   A disposable "intelligent mosquito" with a reversible sampling mechanism using the volume-phase transition of a gel [J].
Suzuki, H ;
Tokuda, T ;
Kobayashi, K .
SENSORS AND ACTUATORS B-CHEMICAL, 2002, 83 (1-3) :53-59
[9]   GELS [J].
TANAKA, T .
SCIENTIFIC AMERICAN, 1981, 244 (01) :124-&