Agreement of intraocular lens power calculation between 2 SS-OCT-based biometers

被引:10
|
作者
Pfaeffli, Oliver A. [1 ]
Weber, Adrian [1 ]
Hoffer, Kenneth J. [2 ,3 ]
Savini, Giacomo [4 ]
Baenninger, Philipp B. [1 ]
Thiel, Michael A. [1 ]
Taroni, Leonardo [5 ]
Mueller, Luzia [1 ]
机构
[1] Cantonal Hosp Lucerne, Dept Ophthalmol, Luzern, Switzerland
[2] Univ Calif Los Angeles, Stein Eye Inst, Los Angeles, CA USA
[3] St Marys Eye Ctr, Santa Monica, CA USA
[4] Fondazione G B Bietti IRCCS, Rome, Italy
[5] Univ Bologna, S Orsola Malpighi Univ Hosp, Eye Clin, Bologna, Italy
关键词
OPTICAL COHERENCE TOMOGRAPHY; INTERFEROMETRY; REPEATABILITY; CONCORDANCE;
D O I
10.1097/j.jcrs.0000000000000788
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess agreement of measurements by 2 swept-source optical coherence tomography biometers and to evaluate the prediction error (PE) in intraocular lens power calculation with 7 formulas. Setting: Tertiary public eye hospital. Design: Consecutive observational. Methods: Axial length (AL), keratometry (K), anterior chamber depth (ACD), lens thickness (LT), and corneal diameter (CD) were measured with the IOLMaster 700 (Biometer A) and Anterion (Biometer B). Agreement was quantified by the limits of agreement and concordance correlation coefficient (CCC). The PE, the median absolute error, and the mean absolute error of the Barrett Universal II, EVO 2.0, Haigis, Hoffer Q, Holladay 1, Kane, and SRK/T formulas were investigated after constant optimization. Results: In 78 eyes from 78 patients, excellent agreement was obtained for AL (CCC >0.99), very good agreement for K, ACD, and LT (CCC >0.95), and strong agreement for CD (CCC >0.72). An additive offset of 0.07 mm was measured for ACD and LT whose mean values were higher with Biometer B (P <.001). No statistically significant difference was found between the PEs and their absolute values when comparing the results of each formula between the 2 biometers. Conclusions: Agreement of biometric measurements by the 2 biometers was high, although Biometer B provided higher mean values of ACD and LT by 0.07 mm. In cataract patients with normal eye length, measurements by the 2 biometers did not lead to different refractive outcomes with the 7 formulas investigated.
引用
收藏
页码:535 / 541
页数:7
相关论文
共 49 条
  • [41] Use of Scheimpflug corneal anterior-posterior imaging in ray-tracing intraocular lens power calculation
    Miyata, Kazunori
    Otani, Shinichiro
    Honbou, Naoto
    Minami, Keiichiro
    ACTA OPHTHALMOLOGICA, 2013, 91 (07) : E546 - E549
  • [42] Predictive Accuracy of Intraocular Lens Power Calculation: Comparison of Optical Low-Coherence Reflectometry and Immersion Ultrasound Biometry
    Turhan, Semra Akkaya
    Toker, Ebru
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2015, 41 (04): : 245 - 251
  • [43] Comparison of Hoffer Q and Haigis Formulae for Intraocular Lens Power Calculation According to the Anterior Chamber Depth in Short Eyes
    Eom, Youngsub
    Kang, Su-Yeon
    Song, Jong Suk
    Kim, Yong Yeon
    Kim, Hyo Myung
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2014, 157 (04) : 818 - 824
  • [44] Accuracy of intraocular lens power calculation formulae after laser refractive surgery in myopic eyes: a meta-analysis
    Li, Hongyu
    Nan, Li
    Li, Jun
    Song, Hui
    EYE AND VISION, 2020, 7 (01)
  • [45] Accuracy of corneal power measurements by a new Scheimpflug camera combined with Placido-disk corneal topography for intraocular lens power calculation in unoperated eyes
    Savini, Giacomo
    Barboni, Piero
    Carbonelli, Michele
    Hoffer, Kenneth J.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (05) : 787 - 792
  • [46] Use of the Posterior/Anterior Corneal Curvature Radii Ratio to Improve the Accuracy of Intraocular Lens Power Calculation: Eom's Adjustment Method
    Kim, Mingue
    Eom, Youngsub
    Lee, Hwa
    Suh, Young-Woo
    Song, Jong Suk
    Kim, Hyo Myung
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (02) : 1016 - 1024
  • [47] Agreement between the newly developed OCT glaucoma staging system and the standardized visual field glaucoma staging system 2
    Mossa, Elshimaa A. Mateen
    Khallaf, Heba
    Sayed, Khulood Muhammad
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2022, 32 (02) : 1009 - 1015
  • [48] Comparison of axial length, anterior chamber depth and intraocular lens power between IOLMaster and ultrasound in normal, long and short eyes
    Dong, Jing
    Zhang, Yaqin
    Zhang, Haining
    Jia, Zhijie
    Zhang, Suhua
    Wang, Xiaogang
    PLOS ONE, 2018, 13 (03):
  • [49] Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients
    Asawaworarit, Rosepon
    Satitpitakul, Vannarut
    Taweekitikul, Parichart
    Pongpirui, Krit
    PLOS ONE, 2022, 17 (05):