Agreement of intraocular lens power calculation between 2 SS-OCT-based biometers

被引:10
|
作者
Pfaeffli, Oliver A. [1 ]
Weber, Adrian [1 ]
Hoffer, Kenneth J. [2 ,3 ]
Savini, Giacomo [4 ]
Baenninger, Philipp B. [1 ]
Thiel, Michael A. [1 ]
Taroni, Leonardo [5 ]
Mueller, Luzia [1 ]
机构
[1] Cantonal Hosp Lucerne, Dept Ophthalmol, Luzern, Switzerland
[2] Univ Calif Los Angeles, Stein Eye Inst, Los Angeles, CA USA
[3] St Marys Eye Ctr, Santa Monica, CA USA
[4] Fondazione G B Bietti IRCCS, Rome, Italy
[5] Univ Bologna, S Orsola Malpighi Univ Hosp, Eye Clin, Bologna, Italy
关键词
OPTICAL COHERENCE TOMOGRAPHY; INTERFEROMETRY; REPEATABILITY; CONCORDANCE;
D O I
10.1097/j.jcrs.0000000000000788
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess agreement of measurements by 2 swept-source optical coherence tomography biometers and to evaluate the prediction error (PE) in intraocular lens power calculation with 7 formulas. Setting: Tertiary public eye hospital. Design: Consecutive observational. Methods: Axial length (AL), keratometry (K), anterior chamber depth (ACD), lens thickness (LT), and corneal diameter (CD) were measured with the IOLMaster 700 (Biometer A) and Anterion (Biometer B). Agreement was quantified by the limits of agreement and concordance correlation coefficient (CCC). The PE, the median absolute error, and the mean absolute error of the Barrett Universal II, EVO 2.0, Haigis, Hoffer Q, Holladay 1, Kane, and SRK/T formulas were investigated after constant optimization. Results: In 78 eyes from 78 patients, excellent agreement was obtained for AL (CCC >0.99), very good agreement for K, ACD, and LT (CCC >0.95), and strong agreement for CD (CCC >0.72). An additive offset of 0.07 mm was measured for ACD and LT whose mean values were higher with Biometer B (P <.001). No statistically significant difference was found between the PEs and their absolute values when comparing the results of each formula between the 2 biometers. Conclusions: Agreement of biometric measurements by the 2 biometers was high, although Biometer B provided higher mean values of ACD and LT by 0.07 mm. In cataract patients with normal eye length, measurements by the 2 biometers did not lead to different refractive outcomes with the 7 formulas investigated.
引用
收藏
页码:535 / 541
页数:7
相关论文
共 49 条
  • [31] Accuracy of toric intraocular lens power calculation depending on different keratometry values using a novel network based software platform
    Ramsauer, Michaela
    Luft, Nikolaus
    Vounotrypidis, Efstathios
    Priglinger, Siegfried G.
    Mayer, Wolfgang J.
    FRONTIERS IN MEDICINE, 2024, 11
  • [32] Accuracy comparison of tomography devices for ray tracing-based intraocular lens calculation
    Wagner, Felix M.
    Hoffmann, Peter
    Preussner, Paul-Rolf
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2024, 50 (02) : 110 - 115
  • [33] Accuracy of a dual Scheimpflug analyzer and a corneal topography system for intraocular lens power calculation in unoperated eyes
    Savini, Giacomo
    Barboni, Piero
    Carbonelli, Michele
    Hoffer, Kenneth J.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (01) : 72 - 76
  • [34] Comparison of biometry and intraocular lens power calculation performed by a new optical biometry device and a reference biometer
    Ventura, Bruna V.
    Ventura, Marcelo C.
    Wang, Li
    Koch, Douglas D.
    Weikert, Mitchell P.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (01) : 74 - 79
  • [35] Influence on intraocular lens power calculation of corneal radii measurement using an image-guided system
    Thomas, Bettina C.
    Mueller, Andreas
    Auffarth, Gerd U.
    Holzer, Mike P.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (11) : 1588 - 1594
  • [36] Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation
    Roncevic, Mirjana Bjelos
    Busic, Mladen
    Cima, Ivan
    Elabjer, Biljana Kuzmanovic
    Bosnar, Damir
    Miletic, Daliborka
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2011, 249 (01) : 69 - 75
  • [37] Agreement of predicted intraocular lens power using swept-source optical coherence tomography and partial coherence interferometry
    Tana-Sanz, Pedro
    Rodriguez-Carrillo, Maria Dolores
    Ruiz-Santos, Maria
    Montes-Mico, Robert
    Ruiz-Mesa, Ramon
    Tana-Rivero, Pedro
    EXPERT REVIEW OF MEDICAL DEVICES, 2021, 18 (12) : 1219 - 1234
  • [38] Comparison of Methodologies Using Estimated or Measured Values of Total Corneal Astigmatism for Toric Intraocular Lens Power Calculation
    Ferreira, Tiago B.
    Ribeiro, Paulo
    Ribeiro, Filomena J.
    O'Neill, Joao G.
    JOURNAL OF REFRACTIVE SURGERY, 2017, 33 (12) : 794 - 801
  • [39] Intraocular lens power calculation after myopic and hyperopic laser vision correction using optical coherence tomography
    Tang, Maolong
    Wang, Li
    Koch, Douglas D.
    Li, Yan
    Huang, David
    SAUDI JOURNAL OF OPHTHALMOLOGY, 2012, 26 (01) : 19 - 24
  • [40] Accuracy of swept-source optical coherence tomography based biometry for intraocular lens power calculation: a retrospective cross-sectional study
    An, Youngju
    Kang, Eun-Kyoung
    Kim, Hyojin
    Kang, Min-Ji
    Byun, Yong-Soo
    Joo, Choun-Ki
    BMC OPHTHALMOLOGY, 2019, 19 (1)