Optimizing the Protocol for Pulmonary Cryoablation: A Comparison of a Dual- and Triple-Freeze Protocol

被引:81
作者
Hinshaw, J. Louis [1 ]
Littrup, Peter J. [2 ]
Durick, Nathan [1 ]
Leung, Winnie [1 ]
Lee, Fred T., Jr. [1 ]
Sampson, Lisa [1 ]
Brace, Christopher L. [1 ]
机构
[1] Univ Wisconsin, Dept Radiol, Madison, WI 53792 USA
[2] Karmanos Canc Inst, Dept Radiol, Detroit, MI 48201 USA
关键词
Interventional radiology; Radiofrequency ablation; Pulmonary; PERCUTANEOUS CRYOTHERAPY; RADIOFREQUENCY ABLATION; SAFETY; MASSES;
D O I
10.1007/s00270-010-9868-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to compare a double freeze thaw protocol to a triple freeze thaw protocol for pulmonary cryoablation utilizing an in vivo porcine lung model A total of 18 cryoablations were performed in normal porcine lung utilizing percutaneous technique with 9 each in a double- (10-5-10) and triple-freeze (3-3-7-7-5) protocol Serial noncontrast CT images were obtained during the ablation CT imaging findings and pathology were reviewed No imaging changes were identified during the initial freeze cycle with either protocol However, during the first thaw cycle, a region of ground glass opacity developed around the probe with both protocols Because the initial freeze was shorter with the triple freeze thaw protocol, the imaging findings were apparent sooner with this protocol (6 vs 13 min) Also, despite a shorter total freeze time (15 vs 20 min) the ablation zone identified with the triple freeze thaw protocol was not significantly different from the double freeze thaw protocol (mean diameter 1 67 +/- 0 41 cm vs 1 66 +/- 0 21 cm P = 0 77, area 2 1 +/- 0 48 cm(2) vs 1 99 +/- 062 cm(2), P = 07, and circularity 0 95 +/- 0 04 vs 0 96 +/- 0 03, P = 0 62 respectively) This study suggests that there may be several advantages of a triple freeze thaw protocol for pulmonary cryoablation including earlier identification of the imaging findings associated with the ablation the promise of a shorter procedure time or larger zones of ablation, and theoretically, more effective cytotoxicity related to the additional freeze-thaw cycle.
引用
收藏
页码:1180 / 1185
页数:6
相关论文
共 11 条
[1]   Percutaneous cryotherapy of the thorax: Safety considerations for complex cases [J].
Ahmed, A ;
Littrup, P .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) :1703-1706
[2]   Mechanisms of tissue injury in cryosurgery [J].
Gage, AA ;
Baust, J .
CRYOBIOLOGY, 1998, 37 (03) :171-186
[3]  
Hinshaw JL, 2009, J INTERV ONCOL, V2, P113
[4]   CT-guided percutaneous cryoterapy of renal masses [J].
Littrup, Peter J. ;
Ahmed, Abraham ;
Aoun, Hussein D. ;
Noujaim, Daniel L. ;
Harb, Ted ;
Nakat, Sam ;
Abdallah, Khaled ;
Adam, Barbara A. ;
Venkatramanamoorthy, Raghu ;
Sakr, Wael ;
Pontes, J. Edson ;
Heilbrun, Lance K. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 18 (03) :383-392
[5]   Cryotherapy for breast fibroadenomas [J].
Littrup, PJ ;
Freeman-Gibb, L ;
Andea, A ;
White, M ;
Amerikia, KC ;
Bouwman, D ;
Harb, T ;
Sakr, W .
RADIOLOGY, 2005, 234 (01) :63-72
[6]   Characterization of the RF ablation-induced 'oven effect': The importance of background tissue thermal conductivity on tissue heating [J].
Liu, Zhengjun ;
Ahmed, Muneeb ;
Weinstein, Yehuda ;
Yi, Ming ;
Mahajan, Roop L. ;
Goldberg, S. Nahum .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2006, 22 (04) :327-342
[7]   Differences in ablation size in porcine kidney, liver, and lung after cryoablation using the same ablation protocol [J].
Permpongkosol, Sompol ;
Nicol, Theresa L. ;
Link, Richard E. ;
Varkarakis, Ioannis ;
Khurana, Hema ;
Zhai, Qihui Jim ;
Kavoussi, Louis R. ;
Solomon, Stephen B. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (04) :1028-1032
[8]   A model for the time dependent three-dimensional thermal distribution within iceballs surrounding multiple cryoprobes [J].
Rewcastle, JC ;
Sandison, GA ;
Muldrew, K ;
Saliken, JC ;
Donnelly, BJ .
MEDICAL PHYSICS, 2001, 28 (06) :1125-1137
[9]   Pulmonary radiofrequency ablation: Long-term safety and efficacy in 153 patients [J].
Simon, Caroline J. ;
Dupuy, Damian E. ;
DiPetrillo, Thomas A. ;
Safran, Howard P. ;
Grieco, C. Alexander ;
Ng, Thomas ;
Mayo-Smith, William W. .
RADIOLOGY, 2007, 243 (01) :268-275
[10]   Unresectable pulmonary malignancies: CT-guided percutaneous radiofrequency ablation - Preliminary results [J].
Suh, RD ;
Wallace, AB ;
Sheehan, RE ;
Heinze, SB ;
Goldin, JG .
RADIOLOGY, 2003, 229 (03) :821-829