共 9 条
Two-photon fluorescence microscopy with differential aberration imaging
被引:1
作者:
Chu, Kengyeh K.
[1
]
Leray, Aymeric
[1
]
Bifano, Thomas G.
[2
]
Mertz, Jerome
[1
]
机构:
[1] Boston U, Dept Biomed Engn, 44 Cummington St, Boston, MA 02215 USA
[2] Boston U, Dept Engn Mech, Boston, MA 02215 USA
来源:
MEMS ADAPTIVE OPTICS III
|
2009年
/
7209卷
关键词:
Two-photon fluorescence;
deformable mirror;
THICK TISSUE;
REJECTION;
D O I:
10.1117/12.812230
中图分类号:
TB3 [工程材料学];
学科分类号:
0805 ;
080502 ;
摘要:
We report our progress in the development of Differential Aberration Imaging (DAI), a technique that enhances two-photon fluorescence (TPEF) microscopy by improving rejection of out-of-focus background by means of a deformable mirror (DM). The DM is used to intentionally add aberrations to the imaging system, which causes dramatic losses to in-focus signal while preserving the bulk of the out-of-focus background. By taking the difference between TPEF images with and without added aberrations, the out-of-focus portion of the signal is further rejected. We now introduce an implementation of DAI using a new type of DM that can be produced at much lower cost.
引用
收藏
页数:5
相关论文