Total variation norm for three-dimensional iterative reconstruction in limited view angle tomography

被引:142
作者
Persson, M [1 ]
Bone, D [1 ]
Elmqvist, H [1 ]
机构
[1] Karolinska Inst, Dept Med Engn, S-14186 Huddinge, Sweden
关键词
D O I
10.1088/0031-9155/46/3/318
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An iterative Bayesian reconstruction algorithm for limited view angle tomography, or ectomography, based on the three-dimensional total variation (TV) norm has been developed. The TV norm has been described ill the literature as a method for reducing noise in two-dimensional images while preserving edges, without introducing ringing or edge artefacts. It has also been proposed as a 2D regularization function in Bayesian reconstruction, implemented in an expectation maximization algorithm (TV-EM). The TV-EM was developed for 2D single photon emission computed tomography imaging, and the algorithm is capable of smoothing noise while maintaining edges without introducing artefacts. The TV norm was extended from 2D to 3D and incorporated into an ordered subsets expectation maximization algorithm for limited view angle geometry. The algorithm, called TV3D-EM, was evaluated using a modelled point spread function and digital phantoms. Reconstructed images were compared with those reconstructed with the 2D filtered backprojection algorithm currently used in ectomography. Results show a substantial reduction in artefacts related to the limited view angle geometry, and noise levels were also improved. Perhaps most important, depth resolution was improved by at least 45%. In conclusion, the proposed algorithm has been shown to improve the perceived image quality.
引用
收藏
页码:853 / 866
页数:14
相关论文
共 11 条
[1]  
DALE S, 1990, J NUCL MED, V31, P1675
[2]   ECTOMOGRAPHY - A TOMOGRAPHIC METHOD FOR GAMMA-CAMERA IMAGING [J].
DALE, S ;
EDHOLM, PE ;
HELLSTROM, LG ;
LARSSON, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (11) :1237-1249
[3]  
Dale SM, 1997, J NUCL MED, V38, P754
[4]   ECTOMOGRAPHY - A NEW RADIOGRAPHIC METHOD FOR REPRODUCING A SELECTED SLICE OF VARYING THICKNESS [J].
EDHOLM, P ;
GRANLUND, G ;
KNUTSSON, H ;
PETERSSON, C .
ACTA RADIOLOGICA-DIAGNOSIS, 1980, 21 (04) :433-442
[5]  
Ganan S, 1985, Amer Statist Assoc, P12
[6]   BAYESIAN RECONSTRUCTIONS FROM EMISSION TOMOGRAPHY DATA USING A MODIFIED EM ALGORITHM [J].
GREEN, PJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1990, 9 (01) :84-93
[7]  
HEBERT T, 1990, IEEE T MED IMAGING, V9, P194
[8]   ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA [J].
HUDSON, HM ;
LARKIN, RS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :601-609
[9]   CONVERGENCE OF EM IMAGE-RECONSTRUCTION ALGORITHMS WITH GIBBS SMOOTHING [J].
LANGE, K .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1990, 9 (04) :439-446
[10]   Total variation regulated EM algorithm [J].
Panin, VY ;
Zeng, GL ;
Gullberg, GT .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (06) :2202-2210