Ultra-short pulse reconstruction software in high power laser system

被引:5
作者
Galletti, M. [1 ,2 ]
Galimberti, M. [3 ]
Giulietti, D. [1 ,2 ]
机构
[1] Univ Pisa, Dept Phys, I-56100 Pisa, Italy
[2] Ist Nazl Fis Nucl, Pisa, Italy
[3] Rutherford Appleton Lab, Cent Laser Facil, Didcot OX11 0QX, Oxon, England
关键词
Ultra-short pulse; High power laser; Ultra-short pulsed laser diagnostics;
D O I
10.1016/j.nimb.2015.02.029
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An ultra-short pulse reconstruction software, validated through a set of experimental measurements on the front-end of the Vulcan laser at the RAL Central Laser Facility is presented. The measurements were acquired in Target Area Petawatt of the Vulcan laser, both using a conventional autocorrelation technique and the GRENOUILLE technique in order to compare the results. The FWHM of the laser pulses considered came out to be comparable for the two techniques. In this experimental campaign for the first time a technique different from the autocorrelation one has been used for a PW class laser as Vulcan. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:232 / 236
页数:5
相关论文
共 4 条
[1]   Spectral phase interferometry for direct electric-field reconstruction of ultrashort optical pulses [J].
Iaconis, C ;
Walmsley, IA .
OPTICS LETTERS, 1998, 23 (10) :792-794
[2]   Highly simplified device for ultrashort-pulse measurement [J].
O'Shea, P ;
Kimmel, M ;
Gu, X ;
Trebino, R .
OPTICS LETTERS, 2001, 26 (12) :932-934
[3]  
Steinmeyer G., 2003, J OPT A, V5
[4]  
Trebino R., 2012, FREQUENCY RESOLVED O